4
ApexPro and CARESCAPE Telemetry
Cleaning and disinfection overview
The following cleaning and disinfection information applies to devices, device
components, supplies, and accessories manufactured by GE.
The information provided in this chapter does not supersede any instructions for use
provided by the manufacturer or provided with a device, device component, supply,
or accessory.
For cleaning, disinfection, and care information for devices, device components,
supplies, and accessories made by manufacturers other than GE, see the applicable
instructions for use provided by the manufacturer.
Cleaning
Cleaning warnings
WARNING
SHOCK HAZARD. Disconnect AC-powered equipment from the
power line before cleaning or disinfecting its surface. Turn off
the power to battery powered equipment before cleaning or
disinfecting its surface.
WARNING
CONTAMINATED LEADWIRES. Contaminated leadwires may
cause infection. Always follow skin preparation guidelines and
leadwire cleaning instructions.
WARNING
IMPROPER TELEMETRY MONITORING DEVICE/LEADWIRE
APPLICATION. Applying a telemetry monitoring device and/or
leadwire that is not thoroughly dry to a patient can result
in an electrically conductive path being established and a
LEADS FAIL
alarm not being provided if leadwires come off
the patient.
WARNING
Avoid using chemicals other than the ones described in this
document, as they may damage device surfaces or labels,
or cause equipment failures.
WARNING
Do not pour or spray any liquid directly on the device, cables,
or leadwires or permit fluid to seep into connections or
openings.
2110145-001
Cleaning and disinfecting
23
Содержание CARESCAPE B105
Страница 8: ...8 Cleaning and disinfecting 2110145 001 ...
Страница 30: ...ApexPro and CARESCAPE Telemetry 30 Cleaning and disinfecting 2110145 001 ...
Страница 48: ...CARESCAPE B450 Monitor 48 Cleaning and disinfecting 2110145 001 ...
Страница 54: ...CARESCAPE B650 Monitor 54 Cleaning and disinfecting 2110145 001 ...
Страница 71: ...content ...