8838 User Guide (BB1857-H)

Endovaginal, Anorectal, and Transrectal Imaging with Type 8838

9

 

Make sure that there is enough imaging gel to cover the entire front end of the 

transducer (the black part of the transducer). This prevents image artifacts 

caused by air bubbles.

2

Pull the transducer cover over the transducer.

3

Before imaging, apply a small amount of gel to the outside of the transducer 
cover to create good acoustic contact between the patient and the transducer.

4

Re-apply the gel frequently to ensure good screen images.

Using the Transducer Control Button

The control button on the transducer controls the imaging.

Press the button to 

Start

 or 

Stop

 imaging (freeze frame).

The transducer makes a “beep” sound each time you press the button.

The button function can be customized. For more information, see the user guide for 
the system and “3D Imaging” on page 11.

Changing Orientation

To change the orientation of the image on the monitor, refer to the applicable system 
user guide for instructions.

To change the angle of the 2D imaging plane, see “Rotating the 2D Imaging Plane” 
on page 10.

Endovaginal, Anorectal, and Transrectal Imaging with Type 8838

Both 2D and 3D imaging in Type 8838 is on the sagittal plane as shown in Figure 2 
and Figure 3.

NOTE:

 The default position for the imaging plane is opposite the button on the 

handle. All linear images start with the linear array in this default position where it 
can display live, 2D images.

Caution 

T-c3

Use only water-based gel (sterile if you are using a sterile transducer cover). Products 
containing parabens, petroleum, or mineral oils may harm the transducer or transducer 
cover.

WARNING

Colo-w1

Do not use excessive force during insertion. Do not make excessive lateral movements 
during or after insertion. Risk of injury or tissue damage to the patient could occur under 
certain circumstances. A digital palpation of the rectum may need to be carried out by a 
clinician prior to insertion or use of the probe as a precautionary measure.

Содержание BK MEDICAL 8838 Series

Страница 1: ...For Professional Users Only English BB1857 H April 2023 User Guide Type 8838 3DART Transducer...

Страница 2: ...products and services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for...

Страница 3: ...ing and Disinfection 7 Starting Imaging 7 Connecting the Transducer 7 Changing Frequency 8 Using a Transducer Cover 8 Using the Transducer Control Button 9 Changing Orientation 9 Endovaginal Anorectal...

Страница 4: ...4...

Страница 5: ...suitable for contrast imaging1 Patient Population The patient population is adults Figure 1 3DART Transducer Type 8838 Imaging Plane The transducer contains a linear array of ultrasound elements that...

Страница 6: ...with the patient Turn off the system Unplug the system from the wall and make sure it cannot be used until it has been checked Do not try to repair the system yourself Contact your BK service represen...

Страница 7: ...n and responsibility to provide the highest possible degree of infection control to patients co workers and themselves The instructions in this book are meant as a guide To avoid cross contamination f...

Страница 8: ...oluble agent inside and outside the cover to create good acoustic contact To put on the cover 1 Apply gel to the tip of the transducer so that it covers the entire imaging surface most of the black pa...

Страница 9: ...e the orientation of the image on the monitor refer to the applicable system user guide for instructions To change the angle of the 2D imaging plane see Rotating the 2D Imaging Plane on page 10 Endova...

Страница 10: ...the position of the transducer array Note that the default position of the imaging plane is opposite the transducer button see Figure 2 Figure 4 Rotation symbol on the flex Focus 1202 Not available o...

Страница 11: ...to its default position You should always check the position of the imaging plane to make sure that it is in the desired postition 3D Imaging NOTE See the advanced system user guide for more informat...

Страница 12: ...the stitch angle is opposite the array s default position that is towards the transducer button see Figure 6 Figure 6 The stitch angle s default position 2 Select the position you want by choosing the...

Страница 13: ...change this parameter on the system Untracked 3D acquisition available with some transducers is not an option for this transducer Select the Mover parameter on the screen There are two options IP8838...

Страница 14: ...lowed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling See the applicable system user guide for further details WARNING D w1 For disp...

Страница 15: ......

Страница 16: ......

Отзывы: