8838 User Guide (BB1857-H)

Cleaning after Use

13

 

When you start the imaging, the control panel is locked to avoid accidental 
interruption of the imaging process. A progress bar is displayed at the top of the 
image as a graphic image and as the number of seconds remaining to show how much 
time is left to complete the image.

The Stitch Line

The 3D imaging must be completed before you can begin to analyze the resulting 
data in any direction. The stitch line is only shown in the final 3D image result. 

Hide or show

stitch line

To hide the stitch line – on Flex Focus 1202:

1

On the 3D General tab, click 

Stitch line

 to hide or show the stitch line.

See the 

Flex Focus Advanced User Guide

 for more information about how to 

assign screen keys.

To hide the stitch line – on Pro Focus 2202 UltraView:

1

On the right side, click 

Layout

.

2

In the drop-down menu, click 

Stitch line

 to hide or show the stitch line.

3D Mover

The 3D mover is automatically selected, so you cannot change this parameter on the 
system. Untracked 3D acquisition, available with some transducers, is not an option 
for this transducer. 

Select the 

Mover

 parameter on the screen. There are two options:

IP8838

 – Moves the internal array while you hold the transducer still.

Untr. Fan

 (Untracked Fan) – Lets you freely move the transducer to acquire

images. See the system user guide for the untracked freehand warning

information, as well as the extended system user guide for image issues

concerning untracked freehand acquisition.

Cleaning after Use

Use a suitable brush to make sure that biological material and gel are removed from 
all channels and grooves. See 

Care and Cleaning

 for cleaning instructions.

WARNING

Reproc-w3

Immediately after use, you must pre-clean the device until visually clean (including 
device lumens if existing). Conduct the thorough cleaning process as soon as 
possible after use in order to prevent bioburden drying on the surface. Dried 
bioburden can lead to inefficient cleaning, disinfection and sterilization, causing a 
risk of cross-contamination.

If pre- and thorough cleaning cannot be done immediately, keep the device moist until 
cleaning.

Содержание BK MEDICAL 8838 Series

Страница 1: ...For Professional Users Only English BB1857 H April 2023 User Guide Type 8838 3DART Transducer...

Страница 2: ...products and services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for...

Страница 3: ...ing and Disinfection 7 Starting Imaging 7 Connecting the Transducer 7 Changing Frequency 8 Using a Transducer Cover 8 Using the Transducer Control Button 9 Changing Orientation 9 Endovaginal Anorectal...

Страница 4: ...4...

Страница 5: ...suitable for contrast imaging1 Patient Population The patient population is adults Figure 1 3DART Transducer Type 8838 Imaging Plane The transducer contains a linear array of ultrasound elements that...

Страница 6: ...with the patient Turn off the system Unplug the system from the wall and make sure it cannot be used until it has been checked Do not try to repair the system yourself Contact your BK service represen...

Страница 7: ...n and responsibility to provide the highest possible degree of infection control to patients co workers and themselves The instructions in this book are meant as a guide To avoid cross contamination f...

Страница 8: ...oluble agent inside and outside the cover to create good acoustic contact To put on the cover 1 Apply gel to the tip of the transducer so that it covers the entire imaging surface most of the black pa...

Страница 9: ...e the orientation of the image on the monitor refer to the applicable system user guide for instructions To change the angle of the 2D imaging plane see Rotating the 2D Imaging Plane on page 10 Endova...

Страница 10: ...the position of the transducer array Note that the default position of the imaging plane is opposite the transducer button see Figure 2 Figure 4 Rotation symbol on the flex Focus 1202 Not available o...

Страница 11: ...to its default position You should always check the position of the imaging plane to make sure that it is in the desired postition 3D Imaging NOTE See the advanced system user guide for more informat...

Страница 12: ...the stitch angle is opposite the array s default position that is towards the transducer button see Figure 6 Figure 6 The stitch angle s default position 2 Select the position you want by choosing the...

Страница 13: ...change this parameter on the system Untracked 3D acquisition available with some transducers is not an option for this transducer Select the Mover parameter on the screen There are two options IP8838...

Страница 14: ...lowed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling See the applicable system user guide for further details WARNING D w1 For disp...

Страница 15: ......

Страница 16: ......

Отзывы: