8838 User Guide (BB1857-H)

Endovaginal, Anorectal, and Transrectal Imaging with Type 8838

11

 

Figure 5. Rotation screen keys on the flex Focus 1202. On the 2202 UltraView system you have 

to assign rotation keys to the user-definable keys on the the keyboard. For more 

information on customizing the 2202 UIltraView keyboard, see the ‘Setting Up and 
Customizing Your System’ chapter in the Pro Focus 2202 Extended User Guide 

(BB1279).

Alternatively, you can rotate the imaging plane using the remote control. See the 
applicable system user guide for further information about the remote control and its 
functions.

Each key press/click rotates the imaging plane by 3°. Hold down the key to rotate the 
plane quicker. The imaging plane can be rotated 180° in either direction from the 
default position, see Figure 2.

When you switch to 3D, the imaging plane moves back to its default position. You 
should always check the position of the imaging plane to make sure that it is in the 
desired postition.

3D Imaging

NOTE:

 See the advanced system user guide for more information about 3D imaging.

Typically, you use the 2D imaging to locate the area where you want a 3D image. 
When the transducer is in the desired location, begin the 3D imaging by pressing the 
3D button on the system. From the system, you can also customize the transducer 
control button.

When you begin the 3D imaging, you define the area for your image on a scale of 10°

 

to 360° in the 

Extent

 parameter on the system screen. When you choose an imaging 

angle, the image is centered on the default position such that each half of the image 
sector is on either side.

During an imaging, the images are captured at the angle and frame rate defined by 
the user on the system. When the desired imaging is completed, the array moves back 
to the default position.

Note that 

t

he transducer and the patient must remain still during 3D imaging to obtain 

accurate images.

Содержание BK MEDICAL 8838 Series

Страница 1: ...For Professional Users Only English BB1857 H April 2023 User Guide Type 8838 3DART Transducer...

Страница 2: ...products and services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for...

Страница 3: ...ing and Disinfection 7 Starting Imaging 7 Connecting the Transducer 7 Changing Frequency 8 Using a Transducer Cover 8 Using the Transducer Control Button 9 Changing Orientation 9 Endovaginal Anorectal...

Страница 4: ...4...

Страница 5: ...suitable for contrast imaging1 Patient Population The patient population is adults Figure 1 3DART Transducer Type 8838 Imaging Plane The transducer contains a linear array of ultrasound elements that...

Страница 6: ...with the patient Turn off the system Unplug the system from the wall and make sure it cannot be used until it has been checked Do not try to repair the system yourself Contact your BK service represen...

Страница 7: ...n and responsibility to provide the highest possible degree of infection control to patients co workers and themselves The instructions in this book are meant as a guide To avoid cross contamination f...

Страница 8: ...oluble agent inside and outside the cover to create good acoustic contact To put on the cover 1 Apply gel to the tip of the transducer so that it covers the entire imaging surface most of the black pa...

Страница 9: ...e the orientation of the image on the monitor refer to the applicable system user guide for instructions To change the angle of the 2D imaging plane see Rotating the 2D Imaging Plane on page 10 Endova...

Страница 10: ...the position of the transducer array Note that the default position of the imaging plane is opposite the transducer button see Figure 2 Figure 4 Rotation symbol on the flex Focus 1202 Not available o...

Страница 11: ...to its default position You should always check the position of the imaging plane to make sure that it is in the desired postition 3D Imaging NOTE See the advanced system user guide for more informat...

Страница 12: ...the stitch angle is opposite the array s default position that is towards the transducer button see Figure 6 Figure 6 The stitch angle s default position 2 Select the position you want by choosing the...

Страница 13: ...change this parameter on the system Untracked 3D acquisition available with some transducers is not an option for this transducer Select the Mover parameter on the screen There are two options IP8838...

Страница 14: ...lowed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling See the applicable system user guide for further details WARNING D w1 For disp...

Страница 15: ......

Страница 16: ......

Отзывы: