18
M1110150 006
© 2007 by General Electric Company. All rights reserved.
Chapter 5: Replacement/Checkout Procedures
Center-Row Irradiance*
(bare fiberoptic light pad)
µW•cm
-2
•nm
-1
Average Irradiance *
(bare fiberoptic light pad)
(+/- 25%) µW•cm
-2
•nm
-1
Large Pad
(D + E + F)/3 ≥ 48.5
(A + B + C + D + E + F + G + H + I)/9 =
49
Small Pad
(C + D)/2 ≥ 70.0
(A + B + C + D + E + F) /6 =
70
* Using an Ohmeda Medical BiliBlanket Meter II
NOTE:
When the BiliSoft fiberoptic light pad is inserted into a BiliSoft Pad Cover or BiliSoft Nest, the nominal
spectral irradiance is:
• 35 µW•cm
-2
•nm
-1
(large pad)
• 50 µW•cm
-2
•nm
-1
(small pad)
Once the check is complete, turn the unit off. The unit is ready for use.
It is recommended that the BiliSoft LED Phototherapy System be checked with the Ohmeda Medical BiliBlanket
®
Meter II for desired therapeutic intensity before use with each patient.