Action
Step
Set Interval time by pressing the value.
5
Result: This opens the Set interval time view.
Enter the value and select Set to to save the value and to return to the pre-
vious view.
6
The Reset button resets the value to zero.
Select the Start button and images are captured repeatedly for the set time
period (or until saturate). During repeated image capture, the contrast of
the images can be adjusted by pressing Auto. This adjustment can be reset
with the Reset button.
7
Result: The number of images matching the selected repetitions are displayed
as they are exposed. A progress bar indicates the time remaining of the
imaging process.
Amersham Imager 600 Operating Instructions 29-0645-17 AD
107
5 Operation
5.5 Capture an image
5.5.2 Chemiluminescence capture