ÄKTA start System Cue Card, 29-0240-42 AC 

8

Evaluating a run

Save Result as BMP image

The system generates a BMP file of the result which can be 
viewed under any OS that has an image viewer supporting 
BMP. The chromatogram is plotted as UV abs (mAU) vs 
Volume (ml), along with Fractionation marks. 

Note: BMP file is generated only for the last 4 hours of run. 

Evaluating a run using UNICORN start

Evaluation

 module of UNICORN start is used to manage 

and evaluate the results from chromatography runs. For 

instructions to evaluate results that are generated using 
ÄKTA start, refer to the 

UNICORN start 1.0 User Manual

.

Save Result to USB memory stick

Import Result to UNICORN start

The result file stored on the USB stick can be imported to 
UNICORN start for analysis, printing and generating a report.

Quick start methods

Quick start

 contains “ready to run” methods like Affinity, 

Ion exchange, Gel filtration and Desalting. The run 
parameters are pre-set in the method. Details of 

sample 

volume

 need to be provided before starting the run.

If required, the run parameters can be changed (

Edit 

run

 in Run view

 screen) when the run is in progress.

Use appropriate columns as indicated in the template 
names. E.g., use 1 ml HiTrap™ column in case of 

selecting the Quick start 

AC/IEX step 1 ml HiTrap

, or 5 ml 

column when selecting 

AC/IEX step 5 ml HiTrap

.

Quick start 

workflow

Connect USB stick to 

ÄKTA start

Select 

chromatography 

method (for method run) 

Manual run / Method run

In the 

Run parameters

 or 

Manual run

 screens, tick the 

checkbox 

Save Result to USB

Import and view 

Result file

 on 

UNICORN start or BMP image 

on any other OS.

UNICORN start

File : Import

Import 

Result 

from USB stick:

• Locate USB drive

• Locate and select the result

   file. Select multiple files using

   

Shift 

or 

Ctrl 

keys.

Open

Select destination folder 

for result

Evaluation

Import

Pre defined

UNICORN start

System control

Method run

Quick start

AC / IEX / DS / GF

Note: The result files will be saved in the GE folder which 
is automatically created by the instrument once the USB 
memory stick is plugged in. Only 10 results can be stored 
in the GE folder. To save further results, transfer the result 
files to another folder, PC or rename the GE folder.

Method run

Select

 the desired 

technique 

Quick start

Set 

Sample volume

Tick the checkbox 

Save Result to USB

Edit run

 to alter 

run parameters

ÄKTA start

End

Run

Exit

Содержание akta start

Страница 1: ...raphy system intended for preparative purification of proteins at laboratory scale The system can be used for a variety of research purposes to fulfill the needs of the users in the academia and in th...

Страница 2: ...w shows the locations of the modules placed on the wet side of the instrument The modules have the following functions To deliver the liquid in the system flow path and divert the flow as required To...

Страница 3: ...chromatography runs Allows creation of methods from predefined chromatography templates like Affinity Gel Filtration Ion Exchange and Desalting Gives flexibility to create a customized method by drag...

Страница 4: ...n Cond flow cell calibration 1 3 2 System preparation Preparation workflow Instruction Place the buffer bottles A B in the buffer tray and immerse the buffer inlet tubing A B in the respective buffer...

Страница 5: ...ple application The sample can be applied to the column in two ways via Pump The sample can be applied directly from a sample container through the Sample valve as shown in the image below It allows a...

Страница 6: ...Edit run Method run Prepare system Column preparation Set Run parameters to suit your requirement KTA start Pre defined System preparation Column preparation UNICORN start System control Method run En...

Страница 7: ...ument Display or from UNICORN start Operation from the Instrument Display The following options are available from Run view screen View real time values UV absorbance Conductivity Flow rate Pressure T...

Страница 8: ...it run in Run view screen when the run is in progress Use appropriate columns as indicated in the template names E g use 1 ml HiTrap column in case of selecting the Quick start AC IEX step 1 ml HiTrap...

Страница 9: ...100 B CV NA NA NA NA NA NA 5 0 5 0 NA Re Equilibration Volume CV 5 0 5 0 NA NA 5 0 5 0 5 0 5 0 NA Fractionation Volume ml 1 0 3 0 0 5 3 0 1 0 3 0 1 0 5 0 4 0 the results will be saved only if the USB...

Страница 10: ...ration Re equilibration Wash unbound Elution Pump wash Equilibration 100 50 5 CV 5 CV 5 CV 5 CV Sample B Fractionation collection 20 CV G r a d i e n t 0 t o 1 0 0 B GF 16 60 HiPrep Gel Filtration Chr...

Страница 11: ...formation visit www gelifesciences com AKTA Selecting protocols columns and buffer Column supported Volume ml Method HisTrap FF 1 or 5 AC HisTrap FF crude 1 or 5 HisTrap HP 1 or 5 GSTrap FF 1 or 5 GST...

Страница 12: ...alth Safety Declaration Form for Product Return or Servicing depending on whether the instrument is going to be serviced on site or returned for service respectively Copy the form you need from KTA st...

Отзывы: