C
ALIBRATION
PROCESS
K0553 Revision D
111
[EN] English
Calibration process
The calibration process is PIN protected. This chapter describes the Ps and Pt sensor calibration.
The
Tools
,
Calibration
Sensor
menu contains the calibration adjustment as part of the function
Calibration check
.
NOTE
:
Calibration checks must be performed within the calibration check menu because the main
Ps/Pt measure display screen Pt reading includes an “Auto Zero” correction for any residual offset
error between the Ps and Pt sensors.
Precision
Precision is a term which includes the ADTS measurement system errors of non-linearity, hyster-
esis and repeatability over the ADTS operating temperature range. Precision does not include
errors introduced by the equipment used for calibration of the ADTS or ADTS calibration stability
errors as stated on the ADTS data sheet.
Table 4-1:
Calibration requirements ADTS542F
ADTS542F
Accuracy specifications(k = 2,
95% uncertainty)Includes: NL, H
and R over the ADTS operating
temperature range and
calibration stability over 12
months and calibration
equipment uncertainty stated in
Table 4-3
Test method
Pneumatic pressure
Ps Range:
92 mbar to 1130 mbar absolute
Accuracy:
Refer to current product data sheet
Compared against calibration
standard
Pt Range:
92 mbar to 1997 mbar absolute
Accuracy:
Refer to current product data sheet
Содержание ADTS552F
Страница 19: ...Chapter 1 Introduction...
Страница 30: ...DUAL HAND TERMINAL EN English 19 K0553 Revision D Figure 1 8 ADTS Touch Slave...
Страница 31: ...Chapter 2 Installation...
Страница 34: ...PACKING FOR STORAGE OR TRANSPORTATION EN English 24 K0553 Revision D...
Страница 40: ...Chapter 3 Operation...
Страница 114: ...Chapter 4 Calibration...
Страница 123: ...Chapter 5 Maintenance...
Страница 132: ...Chapter 6 Testing and Fault Finding...
Страница 152: ...Chapter 7 Specification...