GE Healthcare

URGENT MEDICAL DEVICE CORRECTION

3000 N. Grandview Blvd. - W440

Waukesha, WI 53188, USA

<Date of Letter Deployment>

GEHC Ref# 32047

To:   Nurse Managers, Labor & Delivery/NICU

Bio-Medical Engineering Department Managers

Risk Management Directors

RE:   Infant Warmer System (IWS) Hot “Heater Head” Screw could Fall onto the  Bed

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the “Heater Head” of certain

Infant Warmer System (IWS) devices.

Please ensure that all potential users, as well as those servicing these units, in your 

facility are made aware of this safety notification and the recommended actions.

Safety 

Issue

Hot screws from the “Heater Head” of the IWS could fall onto the bed if the “Heater Head” assembly has 

been  improperly  serviced.  This  situation can  be clinically hazardous  because  thermal injury to  a  patient

could result. Two injuries have been reported as a result of this issue

.

Safety 

Instructions 

The potential for this issue to occur is only present if the “Heater Head” assembly has been improperly 

serviced. 

As a result, if “Heater Head” service has been performed in the past,  check to ensure that the screws are 

tightened as soon as the unit becomes available.

If the “Heater Head” assembly has not been serviced, check to ensure that the screws are tightened as part 

of your next regularly scheduled annual Preventive Maintenance check. 

The  attached  Service  Manual  Addendum  provides  instructions  for  checking  and  tightening  the  screws.  

During each annual Preventive Maintenance check, continue to ensure that the screws are tight.

Affected

Product 

Details

Refer to Product Images below for images of Affected and Not affected products.

Affected model numbers:  Model 2001 IWS (International), 3000 IWS, 3050 IWS, 3100 IWS, 3150 IWS, 3300 

IWS, 3500 IWS, 4000 IWS, 4300 IWS, 4400 IWS, 5000 IWS.

Affected Products

Not Affected Products

Rendered PDF File  Page 2 of 4

DOC1881503, Rev:1

Released

Содержание 2001 IWS

Страница 1: ...tient could result Two injuries have been reported as a result of this issue Safety Instructions The potential for this issue to occur is only present if the Heater Head assembly has been improperly serviced As a result if Heater Head service has been performed in the past check to ensure that the screws are tightened as soon as the unit becomes available If the Heater Head assembly has not been s...

Страница 2: ...illing out and returning the attached Customer Response form Contact Information If you have any questions or concerns regarding this notification please contact GE Healthcare Service at 1 800 437 1171 or your local Service Representative Please be assured that maintaining a high level of safety and quality is our highest priority If you have any questions please contact us immediately per the con...

Страница 3: ... the Medical Device Correction Notice and no longer have a system affected by this Medical Device Correction Notice Please check appropriate disposition If multiple systems or further information attachment can be used Sold Returned Scrapped Other _______________ Device System Serial Number s ___________________________________________________ New Owner if known ___________________________________...

Страница 4: ...Addendum Infant Warmer System Heater Reflector Screws Maintenance Check For future reference add this addendum to your Infant Warmer System Service Manual 2098189 001 A 2016 General Electric Company ...

Страница 5: ...за работа е налично само на английски език Ако доставчикът на услугата на клиента изиска друг език задължение на клиента е да осигури превод Не използвайте оборудването преди да сте се консултирали и разбрали упътването за работа Неспазването на това предупреждение може да доведе до нараняване на доставчика на услугата оператора или пациентa в резултат на токов удар механична или друга опасност ZH...

Страница 6: ...brug for et andet sprog end engelsk er det kundens ansvar at sørge for oversættelse Forsøg ikke at servicere udstyret uden at læse og forstå denne servicemanual Manglende overholdelse af denne advarsel kan medføre skade på grund af elektrisk stød mekanisk eller anden fare for teknikeren operatøren eller patienten NL WAARSCHUWING Deze onderhoudshandleiding is enkel in het Engels verkrijgbaar Als he...

Страница 7: ...îner chez le technicien l opérateur ou le patient des blessures dues à des dangers électriques mécaniques ou autres DE WARNUNG Diese Serviceanleitung existiert nur in englischer Sprache Falls ein fremder Kundendienst eine andere Sprache benötigt ist es Aufgabe des Kunden für eine entsprechende Übersetzung zu sorgen Versuchen Sie nicht diese Anlage zu warten ohne diese Serviceanleitung gelesen und ...

Страница 8: ...þjónustuveitanda stjórnanda eða sjúklings frá raflosti vélrænu eða öðrum áhættum IT AVVERTENZA Il presente manuale di manutenzione è disponibile soltanto in lingua inglese Se un addetto alla manutenzione richiede il manuale in una lingua diversa il cliente è tenuto a provvedere direttamente alla traduzione Procedere alla manutenzione dell apparecchiatura solo dopo aver consultato il presente manua...

Страница 9: ... ansvar å sørge for oversettelse Ikke forsøk å reparere utstyret uten at denne servicehåndboken er lest og forstått Manglende hensyn til denne advarselen kan føre til at serviceleverandøren operatøren eller pasienten skades på grunn av elektrisk støt mekaniske eller andre farer PL OSTRZEŻENIE Niniejszy podręcznik serwisowy dostępny jest jedynie w języku angielskim Jeśli serwisant klienta wymaga ję...

Страница 10: ...исному персоналу клиента необходимо руководство не на английском а на каком то другом языке клиенту следует самостоятельно обеспечить перевод Перед техническим обслуживанием оборудования обязательно обратитесь к данному руководству и поймите изложенные в нем сведения Несоблюдение требований данного предупреждения может привести к тому что специалист по техобслуживанию оператор или пациент получит ...

Страница 11: ...t och förstår den här servicehandboken Om du inte tar hänsyn till den här varningen kan det resultera i skador på serviceteknikern operatören eller patienten till följd av elektriska stötar mekaniska faror eller andra faror SL OPOZORILO Ta servisni priročnik je na voljo samo v angleškem jeziku Če ponudnik storitve stranke potrebuje priročnik v drugem jeziku mora stranka zagotoviti prevod Ne poskuš...

Страница 12: ...enance Check Procedure 1 Turn off the Infant Warmer System and unplug its power cord from the outlet Allow the heat tube to cool down before proceeding with the next steps WARNING Disconnect the power to the warmer and allow the heat tube to cool to avoid the possibility of a burn 2 Locate the warmer unit serial number on the back of controller 3 For units that have serial numbers starting with th...

Страница 13: ...screw driver to remove the screw and alarm light lens from the bottom of the heater head Now the cushioned heater head can be pulled off the front of the unit c Use a wrench to remove the four screws from the support plate end cap and remove the plate d Proceed to step 5 Figure 2 Heater Head Assembly Units with HCC Serial Numbers 5 Pull out the ventilation plate on the top of the heater housing as...

Страница 14: ...n angle to engage the front reflector screw and then use a wrench or nut driver to loosen the nut about one turn Figure 6 Use care not to remove from screw You might need a small step stool to reach to the top of the heater assembly Figure 6 Engaging the Front Reflector Screw 8 While keeping the screw engaged with the screw driver use your fingers to tighten the nut until it is snug 9 Use a 5 16 w...

Страница 15: ...rew engaged with the screw driver use your fingers to tighten the nut until it is snug 12 Use a 5 16 wrench or nut driver to tighten the nut about an additional one third of a turn 13 Re install the ventilation plate into the system Push the side tabs down on the ventilation plate if necessary for proper installation Figure 9 Figure 9 Re installing the Ventilation Plate 14 Reverse step 3 or step 2...

Страница 16: ...IWS Service Manual Addendum 2098189 001 Rev A 13 ...

Страница 17: ... tech Park No1 Huatuo Road Zhangjiang Hi tech Park Pudong Shanghai P R China 201203 上海GE中国科技园 地址 中国上海市浦东张江高科技园华佗路1号 201203 Tel 8621 38777888 Fax 8621 38777402 Brazil Only GE Healthcare Clinical Systems Equipamentos Médicos Ltda Av Paulista 37 13º andar CEP 01311 902 Cerqueira César São Paulo SP Brasil Tel 55 11 3053 2500 Fax 55 11 3053 2573 Australia 1300 722 229 China 800 810 8188 India 1 800 425...