Maintenance
5-9
October 2007
“Balance” Remains on the
Status Line
Check that the sensor and sensor cable are connected. If necessary,
connect them, and the unit will complete the PACER cycle after a
short time (5 to 15 minutes).
The sensor bridge may be out of balance (refer to page 5-1).
No Analog Output
If there is no analog output, but the digital display indicates correctly,
check the analog output scaling.
No Serial Output
Check the serial port programming for proper settings.
Содержание 1111H
Страница 1: ...GE Sensing OptiSonde General Eastern Chilled Mirror Hygrometer User s Manual ...
Страница 9: ...Chapter 1 ...
Страница 10: ...Features and Capabilities Introduction 1 1 Electronics Enclosure 1 2 The System 1 3 Sensors 1 4 ...
Страница 15: ...Chapter 2 ...
Страница 33: ...Chapter 3 ...
Страница 43: ...Chapter 4 ...
Страница 44: ...Programming the OptiSonde Introduction 4 1 Programming Technique 4 2 Programmable Functions 4 3 ...
Страница 57: ...Chapter 5 ...
Страница 68: ...Appendix A ...
Страница 73: ...Appendix B ...
Страница 74: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 77: ...Humidity Equations and Conversion Chart B 3 October 2007 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 78: ...Appendix C ...
Страница 79: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 82: ...Appendix D ...
Страница 83: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing OptiSonde Sensors D 3 ...
Страница 87: ...Appendix E ...
Страница 88: ...Theory of Operation and Glossary Theory of Operation E 1 The PACER Cycle E 4 Glossary E 5 ...