Chapter 2. Instrument Setup
62
DMS Go Operator’s Manual
2.5.1 1-Point Calibration (PROBE & CAL - CALIBRATION)
To calibrate the
DMS Go
, follow the instructions in this section.
1.
In the
MAIN
menu, scroll to the
PROBE & CAL
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
CALIBRATION
. Then, press the function toggle or move the joystick (
) to
Start
.
3.
For 1-point calibration, verify the thickness of the calibration block/piece with the value displayed on the DMS Go when in calibration
mode (if applicable - after probe zeroing has been completed). Press F2 (
ENTER
) to manually change the calibration value. If the
calibration block/piece value is correct, couple the probe to the thickness calibration block. The DMS Go will automatically perform
the calibration and adjust the material velocity to match that of the calibration block/piece.