Revision B
MAC 1200
12-3
2012250-022
Troubleshooting: Troubleshooting Chart
Troubleshooting Chart
Regular AC Line Interference
Symptom
Cause
Remedy
Periodic superimposition of AC line
interference (50 Hz) (see figure below)
Interference from the power line
Ground bed, verify position of the leadwires, switch
on AC line filter
Superimposition of irregular AC line
interference (see figure next page)
Muscle artifact caused by patient
movements, hiccup, coughing
The patient should be warm enough and resting
comfortably (place cushions under arms and knees).
Comfort or distract patient, enable muscle filter
(20 Hz / 40 Hz), if necessary.
The printed date and time are incorrect
Built-in lithium battery is depleted.
The battery has a life of
approximately 5 years
Notify service to check and/or replace battery
The green indicator does not light up,
although the recorder is connected to the
power line
Defective AC power adapter or
fuse
Notify service to check and/or replace fuse
The recorder does not write over the
entire paper width
Paper compartment not properly
closed
Paper door must lock into place on both sides
In Automatic mode, the recorder does not
stop and continues to feed paper. This
does not happen in Manual mode.
The paper pad was inserted the
wrong way round so the recorder
cannot detect the cue mark
Insert the paper pad as instructed
Recorder does not start after activation of
the
key, or the recording is
aborted.
Unit operated on battery power:
battery discharged
Connect recorder to the power line. After a few
minutes, the recorder is able to resume operation.
Always connect recorder to the power line when
indicator lights up. The battery capacity depends on
age, temperature and charge level. (See Chapter 3,
“Operating and Performance Tests”
No recording in Automatic mode
Failure of at least one electrode
Check all electrodes or enable Override function.
(See Chapter 9,
for details).
Paper jam
Open paper compartment and removed jammed
sheet, place beginning of paper between the marks,
close paper compartment and press
.
054A
Содержание MAC 1200
Страница 1: ...MAC 1200 Operator s Manual Software Version 6 2012250 022 Revision B ...
Страница 11: ...Revision B MAC 1200 vii 2012250 022 Entering Special Characters A 1 Special Characters A 3 Index Index 1 ...
Страница 12: ...viii MAC 1200 Revision B 2012250 022 For your notes ...
Страница 13: ...Revision B MAC 1200 1 1 2012250 022 1 The Basics ...
Страница 14: ...1 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 27: ...Revision B MAC 1200 2 1 2012250 022 2 Controls and Indicators ...
Страница 28: ...2 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 33: ...Revision B MAC 1200 3 1 2012250 022 3 Operating and Performance Tests ...
Страница 34: ...3 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 44: ...3 12 MAC 1200 Revision B 2012250 022 Operating and Performance Tests Connecting Peripheral Equipment ...
Страница 45: ...Revision B MAC 1200 4 1 2012250 022 4 Preparing for ECG Recording ...
Страница 46: ...4 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 60: ...4 16 MAC 1200 Revision B 2012250 022 Preparing for ECG Recording Entering Patient Data For your notes ...
Страница 61: ...Revision B MAC 1200 5 1 2012250 022 5 Recording in 12 Lead Mode ...
Страница 62: ...5 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 87: ...Revision B MAC 1200 6 1 2012250 022 6 Recording in 6 Lead Mode ...
Страница 88: ...6 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 93: ...Revision B MAC 1200 7 1 2012250 022 7 Arrhythmia Mode Recording ...
Страница 94: ...7 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 102: ...7 10 MAC 1200 Revision B 2012250 022 Arrhythmia Mode Recording Recording For your notes ...
Страница 103: ...Revision B MAC 1200 8 1 2012250 022 8 Pacemaker Patients Recording During Defibrillation ...
Страница 104: ...8 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 107: ...Revision B MAC 1200 9 1 2012250 022 9 System Setup ...
Страница 108: ...9 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 138: ...9 32 MAC 1200 Revision B 2012250 022 System Setup Direct ECG Transmission For your notes ...
Страница 139: ...Revision B MAC 1200 10 1 2012250 022 10 Loading Chart Paper ...
Страница 140: ...10 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 145: ...Revision B MAC 1200 11 1 2012250 022 11 Cleaning Disinfection and Maintenance ...
Страница 146: ...11 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 149: ...Revision B MAC 1200 12 1 2012250 022 12 Troubleshooting ...
Страница 150: ...12 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 153: ...Revision B MAC 1200 13 1 2012250 022 13 Technical Specifications ...
Страница 154: ...13 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 165: ...Revision B MAC 1200 14 1 2012250 022 14 Order Information ...
Страница 166: ...14 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 171: ...Revision B MAC 1200 A 1 2012250 022 A Entering Special Characters ...
Страница 172: ...A 2 MAC 1200 Revision B 2012250 022 For your notes ...
Страница 174: ...A 4 MAC 1200 Revision B 2012250 022 Entering Special Characters Special Characters For your notes ...
Страница 175: ...Revision B MAC 1200 Index 1 2012250 022 Index ...
Страница 176: ...Index 2 MAC 1200 Revision B 2012250 022 ...
Страница 179: ......