Loading Chart Paper: Procedure

Aging Stability

The standard ECG writer paper CONTRAST® is designed to guarantee 
full contrast for a period between 1 and 5 years if it is handled as 
described below before and after recording.

„

Avoid continued, direct exposure of the paper to sunlight and other 
UV radiation and fluorescent light.

„

Store the paper in suitable rooms at a temperature between 18°C 
and 24°C and a relative humidity between 40% and 60%.

„

Avoid direct contact of the paper with carbon and carbonless forms, 
chart papers and adhesives containing tributyl phosphate, dibutyl 
phthalate, or any other organic solvents, and document protectors, 
envelopes, and sheet separators containing plasticizers. Solvents or 
solvent-based products containing alcohols, ketones, esters, or other 
substances from this chemical group.

&$87,21

The above components may also be found in recycled 
papers.

„

Archiving ECG recordings on our ECG filing cards only (P/N 217 043 
03 is recommended.

„

If longer storage periods are required, use our ARCHIVIST chart 
paper (image legibility up to 15 years) or other image storage 
technologies.

Содержание MAC 1200

Страница 1: ...MAC 1200 Operator s Manual Software Version 6 2012250 022 Revision B ...

Страница 2: ... MULTI LINK MULTISCRIPTOR MUSE MUSE CV Neo Trak NEUROSCRIPT OnlineABG OXYMONITOR Pres R Cuff PRESSURE SCRIBE QMI QS Quantitative Medicine Quantitative Sentinel RAC RAMS RSVP SAM SEER SILVERTRACE SOLAR SOLARVIEW Spectra 400 Spectra Overview Spectra Tel ST GUARD TRAM TRAM NET TRAM RAC TRAMSCOPE TRIM KNOB Trimline UNION STATION UNITY logo UNITY NETWORK Vari X Vari X Cardiomatic VariCath VARIDEX VAS a...

Страница 3: ...irements of standard EN 60601 1 2 Electromagnetic Compatibility Medical Electrical Equipment The safety and effectiveness of this device has been verified against previously distributed devices Although all standards applicable to presently marketed devices may not be appropriate for prior devices i e electromagnetic compatibility standards this device will not impair the safe and effective use of...

Страница 4: ...eneral Information The device is designed to comply with IEC 60601 requirements It is a protection class I device The CE mark covers only the accessories listed in the chapter Order Information The information contained in this manual describes software version 6 ...

Страница 5: ... 7 Illustrations and Names 1 7 Safety Information 1 8 Definitions 1 9 Underwriters Laboratories Inc 1 12 Biocompatibility 1 13 Literature 1 13 Service Information 1 14 Requirements 1 14 2 Controls and Indicators 2 1 General Information 2 3 MAC 1200 Control Panels 2 3 MAC 1200 Keyboard 2 4 Symbols 2 6 3 Operating and Performance Tests 3 1 Power Supply 3 3 Installation and Mains Connection 3 4 Perfo...

Страница 6: ... Patient 4 12 Last Name First Name 4 12 Date of Birth 4 12 Patient ID 4 12 Chest Pain 4 13 Pacemaker 4 13 Gender Race 4 13 Height Weight 4 13 Systolic BP Diastolic BP 4 13 Ordering Physician Referring Physician Technician 4 14 Phone Number 4 14 Medication 4 14 Comments 4 14 ID Required 4 14 Secondary ID 4 14 Secondary ID Required 4 15 Last First Name Required 4 15 Location Number 4 15 Room 4 15 Or...

Страница 7: ...ement Points 5 22 Local T Offset Measurement Point QT Dispersion 5 24 Brief Operating Instructions 12 Lead Mode 5 26 6 Recording in 6 Lead Mode 6 1 General Information 6 3 Recording 6 4 7 Arrhythmia Mode Recording 7 1 General Information 7 3 Recording 7 5 During the Recording 7 6 Final Report 7 8 8 Pacemaker Patients Recording During Defibrillation 8 1 Recording ECGs of Pacemaker Patients 8 3 ECG ...

Страница 8: ...er Feed 9 11 Arrhythmia Mode 9 12 Report Sequence 9 12 Gain 9 12 Muscle Filter 9 12 AC Filter 9 12 Muscle Filter Frequency 9 13 Trend Recording 9 13 Arrhythmia Data 9 13 Episodes 9 13 Pharma 9 14 Patient Data Customization 9 14 Project Code 9 15 Trial ID 9 15 Investigator ID 9 15 Visit Number 9 16 Visit Type 9 16 Dose Type 9 16 Extra Questions 9 16 High Security 9 17 Device Password 9 17 Don t All...

Страница 9: ...est Data 9 23 Restore Defaults 9 23 Print Setup Lists 9 23 Transmission Log 9 24 Check Record Retxn 9 24 Communication 9 25 Baud Rate PC 9 25 Protocol 9 25 Modem 9 25 PIN Dialing 9 26 Patient Data Menu Setup 9 27 Required Data Fields 9 28 Extra Questions 1 to 4 9 28 Option Code 9 29 ECG Transmission via Modem 9 30 Selecting the Communication Protocol 9 30 Direct ECG Transmission 9 31 Selecting the...

Страница 10: ... Specifications 13 1 Recording 13 3 Printer Paper 13 3 Paper Transport 13 3 Membrane Keypad 13 4 Display 13 4 Indicators LEDs 13 4 Lead Selection 13 4 Automatic Functions 13 4 Detection of Pacer Pulses 13 5 Heart Rate Indication 13 5 Signal Inputs 13 5 Data Interface 13 6 Transfer of ECGs 13 6 Receiving Data with the CSI Communication Protocol from the Following Units 13 6 Sending ECGs to the Foll...

Страница 11: ...Revision B MAC 1200 vii 2012250 022 Entering Special Characters A 1 Special Characters A 3 Index Index 1 ...

Страница 12: ...viii MAC 1200 Revision B 2012250 022 For your notes ...

Страница 13: ...Revision B MAC 1200 1 1 2012250 022 1 The Basics ...

Страница 14: ...1 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 15: ...e safe and effective operation of the MAC 1200 unit 127 This document describes the functionality of the U S interface for the MAC 1200 unit Information in this manual differs from operating information for MAC 1200 units developed for use internationally Please refer to PN 2012250 021 for information on using the international interface Revision Date Comments A 31 March 2003 Initial release of ma...

Страница 16: ...he MAC 1200 prior to shipping Software Package Functionality Option Code MEAS Measurement measurement of the 10 second resting ECG _ _ _ _ _ _ _ _ _ _ _ _ DIAG Interpretation interpretation of the 10 second resting ECG _ _ _ _ _ _ _ _ _ _ _ _ MEMO Memory storage of a maximum of 40 10 second resting ECGs _ _ _ _ _ _ _ _ _ _ _ _ C100 Activates the three options MEAS DIAG MEMO for a maximum of 100 EC...

Страница 17: ...red with the appropriate options It is intended to be used by trained operators under direct physician supervision when ECG records are required It is not suitable for intracardiac application It is designed for continuous operation It is not intended for home use The MAC 1200 is designed as a portable device and can easily be moved from one patient to another or to different locations It is not i...

Страница 18: ... a setup menu to customize the system parameters Patient and user data can be entered for reliable and safe archiving of patient records The patient name is annotated on each printed report page All other data is printed on request The MAC 1200 units are designed to comply with IEC 60601 EN 60601 requirements They are protection class I devices devices with an internal power source They are classi...

Страница 19: ...appearing in examples and illustrations are fictitious The use of any real person s name is purely coincidental Black text Indicates keys on the keyboard text to be entered or hardware items such as buttons or switches on the equipment Italicized text Indicates software terms that identify menu items buttons or options in various windows Ctrl Esc Indicates a keyboard operation A sign between the n...

Страница 20: ...alid at the time of printing All rights are reserved for devices circuits techniques software programs and names appearing in this manual On request GE will provide a service manual The GE quality management system complies with the standards DIN EN ISO 9001 and EN 46001 To ensure patient safety the specified measuring accuracy and interference free operation we recommend to use only original GE c...

Страница 21: ...a potential hazard or unsafe practice which if not avoided could result in death or serious injury 87 21 indicates a potential hazard or unsafe practice which if not avoided could result in minor personal injury or product property damage 127 provides application tips or other useful information to assure that you get the most from your equipment The safety information given in this manual is clas...

Страница 22: ...ected to an appropriate power supply with a non fused grounded to earth wire If these requirements cannot be met operate the device on battery power Do not use multiple portable socket outlets MPSO to connect the device to the power line Devices may be connected to other devices or to parts of systems only when it has been made certain that there is no danger to the patient the operators or the en...

Страница 23: ...these devices and verify the recorder performance before use SUFFOCATION HAZARD Dispose of the packaging material observing the applicable waste control regulations Keep the packaging material out of children s reach 87 216 EQUIPMENT DAMAGE Devices intended for emergency application must not be exposed to low temperatures during storage and transport to avoid moisture condensation at the applicati...

Страница 24: ... protection against electrical shock Type CF defibrillation proof applied part Degree of protection against harmful ingress of water Ordinary Equipment enclosed equipment without protection against ingress of water Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Equipment not suitable for use in the presence of a flammable ...

Страница 25: ...le standards If you have questions in this matter please contact GE Medical Systems Information Technologies or its representatives Literature Medical Device Directive of August 2 1994 EN 60601 1 1990 A 1 1993 A 2 1995 Medical electrical equipment General requirements for safety EN 60601 1 1 9 1994 A1 12 1995 General requirements for safety Requirements for the safety of medical electrical systems...

Страница 26: ...e user s responsibility to report the need for service to GE Medical Systems Information Technologies or to one of their authorized agents Every GE Medical Systems Information Technologies device has a unique serial number for identification The serial number appears on the device label Table 1 Equipment Identification Item Description A name of device B manufacturer C location code D serial numbe...

Страница 27: ...Revision B MAC 1200 2 1 2012250 022 2 Controls and Indicators ...

Страница 28: ...2 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 29: ...alt Q 1 W 2 E 3 R 4 T 5 Y 6 U 7 I 8 O 9 P 0 sta rt sto p arr hy pat info sto re retr ieve lea d cop y set up gai n forma t spe ed mu scl e filte r 12 6 stan dby batt ery low on off MAC 1200 1 001A 1 2 3 4 5 Description 1 Power input 2 Paper door windows allows you to check the paper supply 3 Patient cable connector 4 Connection for electrode application system KISS option 5 Serial interface See Ch...

Страница 30: ...he configuration menu 6 Enables disables the muscle filter elimination of muscle artifact 7 Selects the writer speed 25 50 or 5 mm s in 6 Lead Mode and the report formats in 12 Lead Mode 8 Selects the gain 5 10 20 40 mm mV 9 Press to print the report or additional copies of the ECG or to send receive ECGs 10 Selects the ECG leads displayed and recorded in 6 Lead Mode and displayed 12 Lead Mode 11 ...

Страница 31: ...Information 17 Enables entry of patient data 18 Indicator lights up when battery needs to be recharged 19 Indicator is illuminated when unit is connected to the power line 20 Cursor control keys 21 Space bar 22 Shift key 23 Press to access special characters Description ...

Страница 32: ...2 6 MAC 1200 Revision B 2012250 022 Controls and Indicators Symbols Symbols Consult accompanying documents Signal input Type CF signal input highly insulated defibrillation proof Start Stop ...

Страница 33: ...Revision B MAC 1200 3 1 2012250 022 3 Operating and Performance Tests ...

Страница 34: ...3 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 35: ...hen the battery needs to be charged The unit can also be set up to emit an additional audio signal when the battery requires charging With a full battery about 50 ECGs 1 page each can be recorded in the 12 Lead mode When its capacity drops to about 25 recordings the battery is used up and must be replaced by a service specialist 127 To prolong the battery life fully discharge the battery at least ...

Страница 36: ...cable and the power cord do not run parallel Arranging the Electrocardiograph and the Examination Couch Use the power cord to connect the device to the power line see figure below Use only the original power cord or an equivalent cable AC Power Input Indicator 2 will illuminate Check the paper supply the window in the paper door allows you to look inside the paper compartment Refer to Chapter 10 L...

Страница 37: ...tified the display will show an error message Error In this situation notify service to check and repair the device The self test can be skipped with the button In this case the device immediately activates the default mode confugured in the Setup menu 127 The backlighting of the display switches off automatically when no key is activated for 20 minutes adjustable The illumination is turned on aga...

Страница 38: ...3 6 MAC 1200 Revision B 2012250 022 Operating and Performance Tests Performance Check Contrast Adjustment Simultaneously press and the appropriate cursor key more contrast less contrast Alt ...

Страница 39: ...xt box text box for 40 characters Cart 1 1 9999 Site 1 1 1 255 Location 1 600 Ease of Clock Setting No Yes Date dd mm yyyy current date Time hh mm current time Lead fail beep No Yes High HR beep No Yes Lead Labels IEC AAMI Pace Enhancement Yes No Baseline roll filter 0 08 Hz 0 04 Hz 0 16 Hz Date dd mm yyyy mm dd yyyy Time 24 12 Units cm kg in lb Mains 50 Hz 60 Hz LCD light off after 20 min 1 99 mi...

Страница 40: ...3 8 MAC 1200 Revision B 2012250 022 Operating and Performance Tests System Setup 1 only for CSI communication protocol ...

Страница 41: ...onnected to the same circuit must be electrically isolated use isolated RS232 interface as per IEC 60601 1 A PC connected to the electrocardiograph should meet the requirements of EN 60601 If it doesn t it must be set up outside the patient environment If the PC fulfills the requirements of EN 60950 it must be set up within the medically used area but outside the patient environment Do NOT connect...

Страница 42: ... ECG Transmission section for details The table below shows the factory defaults and all possible adjustments For instructions on changing the default setup see Communication on page 9 25 for details Table 2 Modem Configuration Menu Parameter Factory Default Options Choices for Modem Other no other MultiTech 19 2 MultiTech 56k Elsa 14 4 Elsa 28 8 Elsa 33 6 Elsa 56k Choices for Modem other Phone Nu...

Страница 43: ...Revision B MAC 1200 3 11 2012250 022 Operating and Performance Tests Connecting Peripheral Equipment For your notes ...

Страница 44: ...3 12 MAC 1200 Revision B 2012250 022 Operating and Performance Tests Connecting Peripheral Equipment ...

Страница 45: ...Revision B MAC 1200 4 1 2012250 022 4 Preparing for ECG Recording ...

Страница 46: ...4 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 47: ...ctive parts such as the patient connectors electrodes transducers that are connected to the isolated patient signal input do not come into contact with other grounded conductive parts This would bridge the patient s isolation and cancel the protection provided by the isolated input The neutral electrode in particular must not come into contact with ground If your electrocardiograph is equipped wit...

Страница 48: ...ENT HAZARD Delayed ECG Display Use only silver silver chloride electrodes when recording the ECG of a patient who may have to be defibrillated Refer to Chapter 8 Pacemaker Patients Recording During Defibrillation for details Careful application of electrodes and skin preparation are the key to an interference free ECG ...

Страница 49: ...ad Electrodes 87 21 Use only silver silver chloride electrodes if the patient may have to be defibrillated See ECG Recording During Defibrillation on page 8 4 for details Applying Suction Electrodes Chest Shave application points if necessary Moisten the electrode paper with tap water and place it between skin and electrode Electrode cream or gel can be used instead of paper On hairy chests the cr...

Страница 50: ...leads Description V1 4th intercostal space at the right border of the sternum V2 4th intercostal space at the left border of the sternum V3 Midway between locations V2 and V4 V4 At the mid clavicular line in the 5th intercostal space V5 At the anterior auxiliary line on the same horizontal level as V4 and V6 V6 At the mid auxiliary line on the same horizontal level as V4 V7 At the left posterior a...

Страница 51: ... Connecting the Patient Cable 10 Lead Cable Standard ECG Leads Arrange the leadwires and patient cable as shown below Arranging the Patient Cable V1 V1 V2 V3 V4 V5 V6 V2 V3 V4 V5 V6 009A right leg right arm left arm left leg RL green RA white V1 red V2 yellow V3 green V4 blue V5 orange V6 purple LA black LL red 010A Correct Incorrect ...

Страница 52: ...red see Sample Recording figure For the Manual Mode the anti drift system cubic spline can be enabled and disabled from the setup menu in the 12 Lead and Arrhythmia Modes it is always enabled When electrodes are not properly applied these measures may not fully compensate for artifact High polarization voltages induced by electrodes applied without conductive gel may cause the amplifier to overran...

Страница 53: ...e cream spray etc Wait approximately 10 seconds before initiating a recording After the 10 second period the automatic functions are enabled and the polarization voltages have stabilized provided the electrodes are properly applied In case of improper electrode application an error message will appear on the display RL LL LA LL V1 to V6 If required the ADS cubic spline and the filters 20 40 Hz 50 ...

Страница 54: ...t appear in the patient data menu in the default configuration are marked as Yes in the Menu item displayed column the other menu items are marked as No To skip a menu item press or the cursor key or It is not possible to write capital and small letters do not press the Shift key For entry of numbers e g date of birth it is not necessary to press the Shift key For fields that allow alphanumeric en...

Страница 55: ...est pain1 Unknown Yes Secondary Complaint Chief Complaint Pacemaker No Yes Yes Gender Yes female male Height Yes Weight Yes Race unknown Yes other Systolic BP 0 mmHg Yes Diastolic BP 0 mmHg Yes Ordering Physician Yes selection from a list of 10 names Referring Physician Yes selection from a list of 10 names Technician Yes selection from a list of 10 names Phone No Yes Medication Yes 1 unknown Yes ...

Страница 56: ...mbol appears in the display to indicate the Shift Lock status Patient ID This field accepts 3 to 16 characters The exact length is determined in the patient data setup menu 127 When entering a patient ID which consists of numerals only the blanks preceding the number are replaced with 0 Example If a 6 digit text box is configured and you enter the patient ID 123 the final ID number will read 00012...

Страница 57: ...n the program will determine a percentage value indicating the probability of acute ischemia This value along with a reason for this conclusion will appear in the interpretation Pacemaker Influences the identification of pacer pulses in Arrhythmia Mode Enable the function Yes when recording the ECG of a pacemaker patient The recording will then be annotated with the message Pacemaker Patient Gende...

Страница 58: ...n press to quit the list The Referring Physician is only relevant if you send ECGs to the MUSE CV system This name will not be annotated on the ECG recording Phone Number Enter the patient s telephone number Medication Enter the patient s medications and confirm entries with Comments 4 lines of 30 characters each ID Required Default value is Yes Secondary ID This field accepts 3 to 16 characters T...

Страница 59: ... Required Yes patient last or first name required No patient last or first name not required Location Number ID number for the location 3 place When you select Yes for New patient the default value from the System Setup will automatically be adopted but you can overwrite this number Room 5 digits Order Number 9 digits for entry of the ECG order number Prompts Answer the prompts entered in the pati...

Страница 60: ...4 16 MAC 1200 Revision B 2012250 022 Preparing for ECG Recording Entering Patient Data For your notes ...

Страница 61: ...Revision B MAC 1200 5 1 2012250 022 5 Recording in 12 Lead Mode ...

Страница 62: ...5 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 63: ...interpreted analyzed and printed out with the interpretative statements MAC 1200 1200 ST with option DIAG Units equipped with the optional memory function MEMO MAC 1200 only can save up to 40 resting ECG These ECGs can be printed or sent to CardioSys CardioSoft A5 protocol or to the MUSE CV system CSI protocol see The Storage Program on page 5 8 The unit offers different report formats for printou...

Страница 64: ...RA Rhythm Leads II V1 V5 I III aVR aVL aVF V2 V3 V4 V6 Gain 10 mm mV auto 5 20 40 mm mV Report Format 4x2 5R1 1x10R12 2x5R1 2x5_50 4x2 5R1 1x10R3 4x2 5R3 Detailed Results No Yes Muscle Filter No Yes Muscle Filter Frequency 40 Hz 20 Hz AC Filter Yes No Manual copy to EKG HOST Number of copies 1 0 to 9 Delete ECG after Transmission No Yes Autosave ECG No Yes Use Screening criteria No Yes Suppress NO...

Страница 65: ...tion voltages see Chapter 4 Artifact Due to Poor Electrode Application If you initiate a recording with immediately after selecting the 12 Lead mode the message Collecting data displays and you will need to wait for 10 to 12 seconds Before initiating a recording check the display for error messages Check all electrodes if the message persists there must be a break in the patient cable Replace the ...

Страница 66: ... filters if necessary With the system defaults unchanged the unit will activate the following functions and settings after power up system defaults which can be changed are indicated with configurable 12 Lead mode configurable Standard report sequence lead to channel correspondence I II III aVR aVL aVF V1 V2 V3 V4 V5 V6 Rhythm leads II V1 and V5 configurable Gain of 10 mm mV configurable calibrati...

Страница 67: ...G measurement interpretation The locations of all relevant device settings are shown below The display shows three leads at a time With you can display all leads of the report sequence in groups of three The recording can be stopped with For a description of the different reports see Report Formats on page 5 10 Operating mode Muscle filter1 1 if enabled AC line filter1 Anti drift system1 Patient N...

Страница 68: ... at the top of the menu Print prints the selected ECG Send see ECG Transmission on page 5 12 Delete deletes the selected ECG Change enables modification of the patient data When you view a saved ECG the Print function is selected The list on the next page shows all patients for whom ECG are stored To perform an action with one or more ECGs you must first select the ECG s Follow these steps 1 Move ...

Страница 69: ... Initiate it with Other Considerations With the Print directory command you can print a list of all stored ECGs The printout includes all columns except the Comments column If you try to save an ECG when the memory is full a message informs you of the memory status When you delete an ECG from memory the new ECG will automatically be saved The unit may be set up to automatically save ECGs without p...

Страница 70: ...x3 10 s 3 25 mm s Yes Yes 1 4x2 5R3 2x5R1 2x5 s 2x6 10 s 1 25 mm s Yes Yes 1 2x5R1 2x5_50 2x5 s 2x6 No 50 mm s Yes Yes 2 2x5_50 1x10R12 10 s 1x12 No 25 mm s No No 1 1x10R12 1x10R3 10 s 1x3 10 s 3 25 mm s Yes Yes 1 1x10R3 4x2 5R1 default format 4x2 5s 4x3 10 s 1 25 mm s Yes Yes 1 4x2 5R1 default format 4x2 5R3 4x2 5 s 4x3 10 s 3 25 mm s Yes Yes 1 4x2 5R3 2x5R1 2x5 s 2x6 10 s 1 25 mm s Yes Yes 1 2x5...

Страница 71: ...e Detailed Results page When selected this page will be appended to the report It contains patient data measurement results interpretative statements reasons are only annotated if the unit is equipped with the HEART interpretation program medians and the tabular measurement values 127 To obtain a printout of the full patient data select the 6 Lead mode and press copy ...

Страница 72: ...ng 1 of 1 or Transmitting 2 of 6 With the transmission can be stopped ECGs that were successfully transmitted are identified with the letter S for Sent After the selected ECG records have been successfully transmitted a confirmation message displays e g 6 of 6 ECG records successfully transmitted All ECG records are time stamped hours minutes and seconds To view a record s time stamp highlight it ...

Страница 73: ...elect multiple ECGs select them with while holding down Selected ECGs are marked with a black bar in the margin of the display Deselect ECGs with the same key combination 3 When you have selected the ECG s to transfer press You will be returned to the function key menu 4 Using the cursor select the Send command Confirm your choice with The transmission menu will be displayed showing the phone numb...

Страница 74: ...ng modem setup modem off etc the unit will display an error message such as Transmission Error A5 In this situation you have the following choices Repeat the transmission with Change the settings with Stop the transmission with You must select one of these choices to clear the transmission error message from the display 127 Unsuccessful transmission does not remove delete the ECG record from the l...

Страница 75: ...n is provided Then it sends batch of ECG records 3 After the transmission a message on the display indicates the number of successfully transmitted ECGs As soon as you acknowledge the message with the 12 Lead mode acquisition screen appears Transmission Log The MAC 1200 unit can be set up to automatically print a log of the recently completed transmission See Transmission Log on page 9 24 for more...

Страница 76: ...ins a unique serial number for tracking purposes It is made up of the MAC 1200 serial number date time 127 The ECG record serial number is printed on the transmission log only if enabled It does not appear on the ECG report printed at the MAC 1200 unit See Transmission Log on page 9 24 for more information Direct Transmission The unit must be connected to the PC or to the MUSE CV system by means o...

Страница 77: ...gin of the display Deselect the ECGs with the same key combination 3 When you have selected the ECG s to transfer press and you are back in the function key menu 4 Using the cursor select the Send command Confirm your choice with The message Initializing transmission is displayed and the option to cancel the transmission is provided Then it starts The message ECG Transmission A5 is displayed 5 Aft...

Страница 78: ...ted protocol 127 Pacemaker information telephone number and comments entered in the patient data are not transmitted to the MUSE CV system In this situation you have the following choices Repeat the transmission with Change the settings with Stop the transmission with You must select one of these choices to clear the transmission error message from the display 127 Unsuccessful transmission does no...

Страница 79: ...e 12 Lead REC indicates that the unit is ready to receive data 4 A message displays on the screen when the unit is receiving data The reception of data can be aborted with 5 The ECG which has just been received is processed for the printout The report is printed in the selected format Multiple ECGs are received and printed one after the other After printout of the last ECG the receive data mode is...

Страница 80: ...ease refer to the modem user instructions Three command sequences have to be entered each of which defines a specific modem operating state 1 The modem is initialized init string 2 A communication link is established dial string 3 The communication is terminated hangup string These three strings are entered in the modem setup menu See Chapter 3 Connecting Peripheral Equipment The example below sho...

Страница 81: ...s terminated in two steps First the MAC 1200 sends an escape command to return from the on line state to the command state Then the hangup command follows Symbol Description AT Prefix that precedes every command line DT Touch tone dial mode xxx After DT enter the characters for access to the public telephone network e g 0 W W placed after a number tells the modem in a PBX system to wait for the di...

Страница 82: ...quired Global Measurement Points After acquisition of an ECG in the 12 Lead mode the global measurement points for P onset P offset QRS onset QRS offset and T offset can be adjusted manually After acquisition of the ECG press to display the screen for verification of the global measurement point markers Verification of the Global Measurement Points Description 1 Amplitude in mV 2 Selected lead 3 A...

Страница 83: ...ground The active measurement point marker is large the four inactive markers are small Use these keys to adjust the markers moves the active marker right or left selects the next or previous marker activates the next lead changes the gain terminates the adjustment saving the changes terminates the adjustment without saving the changes space bar for the P onset and P offset values the space bar to...

Страница 84: ...lobal measurement points the screen for verification of the T offset measurement point appears automatically Verification of the T Offset Measurement Point Therefore this screen always shows only one lead at a time and the T offset point Changing the local T offset point also affects the QT dispersion value Description 1 Amplitude in mV 2 Selected lead 3 Time in s 4 Active marker 1 mV 1 s 029A 1 2...

Страница 85: ...ng the changes When you exit the screen the acquisition screen for the 12 Lead mode reappears The corrected ECG can be printed with the button If the unit is equipped with the MEMO memory option the corrected data can be saved or it will be saved automatically when the corresponding function is enabled If the original ECG had already been saved the corrected data will overwrite this ECG 127 Changi...

Страница 86: ...t to end Apply electrodes to patient Enter the following patient data Check device settings Report sequence Report format AC filter Override function 12SL interpretation configuration Modify device settings if required Wait for patient to lie motionless and for the unit to collect 10 seconds of ECG data Check that no lead failure message is displayed Start recording with pat info setup ...

Страница 87: ...Revision B MAC 1200 6 1 2012250 022 6 Recording in 6 Lead Mode ...

Страница 88: ...6 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 89: ...ith configurable The following information refers to a unit with the system defaults For instructions on changing the default setup refer to 6 Lead Mode on page 9 9 for details Table 1 Setup Menu for the 6 Lead Mode Parameter Factory Default Options Report Sequence STANDARD CABRERA SEQ NO 4 Gain 10 mm mV auto 5 20 40 mm mV Speed 25 mm s 5 50 mm s Muscle Filter No Yes Muscle Filter Frequency 40 Hz ...

Страница 90: ...table below Check all electrodes if the message persists there must be a break in the patient cable Replace the cable with a new one The recording is started and stopped with Table 2 Messages Indicating Disconnected Electrodes RL Right leg electrode disconnected RA Right arm electrode disconnected LA Left arm electrode disconnected LL Left leg electrode disconnected V1 Chest electrode V1 disconnec...

Страница 91: ...5 10 20 and 40 mm mV 127 Please note that filters may suppress diagnostically relevant portions of the signal because they limit the transmission range Filters should therefore only be enabled if necessary selects the other 3 leads of the selected report sequence display the other 3 leads of the selected group of 6 The writer prints at a speed of 25 mm s the speed can be changed with The muscle fi...

Страница 92: ... is enabled there will be a short delay of 2 2 seconds before the recording starts The heart rate limit is automatically calculated from the date of birth WHO 100 220 age When the date of birth is not entered the unit will set the limit at 180 bpm This value can be changed with and in steps of 5 bpm Operating mode Muscle filter1 1 if enabled AC line filter1 Anti drift system1 Patient Name Report f...

Страница 93: ...Revision B MAC 1200 7 1 2012250 022 7 Arrhythmia Mode Recording ...

Страница 94: ...7 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 95: ...In the setup menu Chapter 9 System Setup you determine the conditions for a recording The recorder starts each time it detects a single beat event The recorder starts each time it detects an event different from the previous event The recorder does not start at all Event Recording 127 After starting the program press to select a continuous recording at 5 mm s c If the unit identifies an arrhythmia...

Страница 96: ... default setup refer to Arrhythmia Mode on page 9 12 Table 1 Setup Menu for the Arrhythmia Mode Parameter Factory Default Options Report sequence STD_C chest leads V1 through V6 STD_RED I II III V2 V4 V6 STD_LI I II III aVR aVL aVF CABR_LI aVL I aVR II aVF III HIGH_C V1 through V6 Gain 10 mm mV auto 5 20 40 mm mV Muscle Filter No Yes Muscle Filter Frequency 40 Hz 20 Hz AC Filter Yes No Trend Rec N...

Страница 97: ... all electrodes if the message persists there must be a break in the patient cable Replace the cable with a new one 3 Start and stop the recording with the key Table 2 Messages Indicating Disconnected Electrodes Symbol Definition RL right leg electrode disconnected RA right arm electrode disconnected LA left arm electrode disconnected LL left leg electrode disconnected V1 chest electrode V1 discon...

Страница 98: ...relevant portions of the signal because they limit the transmission range Only enable filters if necessary With the system defaults the MAC 1200 will activate the following functions and settings STD_C report sequence lead to channel correspondence V1 to V6 configurable Gain of 10 mm mV configurable calibration pulse at the beginning of the recording The unit can be set up to automatically adapt t...

Страница 99: ...it will set the limit at 180 bpm This value can be changed with and in steps of 5 bpm 127 The letter A on a recording indicates the presence of artifact which does not allow the algorithm to identify arrhythmias Causes include wandering baselines The anti drift system largely prevents these disturbances Check the electrodes and leadwires Operating mode Muscle filter1 1 if enabled AC line filter1 A...

Страница 100: ...d with Then the final report can be printed with The final report consists of the following Patient ID sheet with all patient data as well as with all analyzed QRS complexes type and number of detected events and the analysis duration in tabular form Episode report 3 sheets maximum with 2 episodes each Arrhythmia codes copy ...

Страница 101: ...beat PAU1 Early PVC EPVC Ventricular bigeminy VBIG New form e g intermittent bundle branch block NF Multiform PVCs MULT Supraventricular arrhythmia SVAR Paroxysmal supraventricular tachycardia PSVT Tachycardia TACH Bradycardia BRAD Pacemaker malfunction PERR Ventricular escape beat ESC Premature ventricular contraction PVC Premature supraventricular contraction PSVC Aberrant beat ABR Pacemaker cap...

Страница 102: ...7 10 MAC 1200 Revision B 2012250 022 Arrhythmia Mode Recording Recording For your notes ...

Страница 103: ...Revision B MAC 1200 8 1 2012250 022 8 Pacemaker Patients Recording During Defibrillation ...

Страница 104: ...8 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 105: ...asier to identify The MAC 1200 records the pulse with the correct polarity with a width of 5 ms and with the same amplitude in all leads depending on the polarity of the pacer pulse in leads I and II the pacer pulse in lead III may be suppressed The amplitude of the reverse current in the figure below shows an ECG recording with pacer pulses ECG Recording with Pacer Pulses 51 1 INCORRECT HR NO HR ...

Страница 106: ...0 to 6 Lead mode when you may have to defibrillate the patient while recording the ECG and disable the anti drift system which would cause a 2 second signal delay 6 Lead Mode on page 9 9 for details When using other electrodes disconnect the patient cable from the recorder while defibrillating the patient 51 1 EQUIPMENT DAMAGE For reasons of patient safety use only the original GEMS IT patient cab...

Страница 107: ...Revision B MAC 1200 9 1 2012250 022 9 System Setup ...

Страница 108: ...9 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 109: ...e appropriate mode before entering the configuration menu To access the menu options position the bar cursor on the option with the cursor keys and confirm the selection with The operating steps to select a setting are always the same Using the cursor keys and you select the setting and confirm the selection with The cursor will move to the next menu item Individual items can be skipped with or To...

Страница 110: ...9 4 MAC 1200 Revision B 2012250 022 System Setup General Information System Setup Keys N M start stop arrhy pat info store retrieve gain standby battery low 045A ...

Страница 111: ...L aVF V1 V2 V3 V4 V5 V6 CABRERA aVL I aVR II aVF III V1 V2 V3 V4 V5 V6 Rhythm Leads Any three of the available ECG leads can be selected as rhythm leads They are printed with report formats 4x2 5R3 and 1x10R3 Formats 2x5R1 and 4x2 5R1 will show only the first of the rhythm leads Gain 5 10 20 40 mm mV auto Report Format Refer to Chapter 5 Report Formats section Detailed Results The Detailed results...

Страница 112: ... will be appended to the report name Report formats 6_F1 3_F1 4x2 5R3 6_F2 H2 become 6_F1_R 3_F1_R 4x2 5R3_R 6_F2_R H2_R Muscle Filter AC Filter Elimination of muscle artifact and AC line interference Muscle filter no AC line filter yes 127 Filters may suppress diagnostically relevant portions of the signal because they limit the transmission range Therefore only enable filters if necessary Muscle...

Страница 113: ...at were successfully sent to a host system via the RS232 interface will be cleared from the recorder memory Default no If this menu item is set to Yes and ECGs have already been sent from the recorder memory these ECGs will be deleted after the next transmission of a stored ECG Autosave ECG MAC 1200 with Memory Option MEMO Only After report generation the ECG will or will not be automatically save...

Страница 114: ...recorder will print in 12 Lead mode even when not all electrodes are applied or do not supply a good signal When electrodes are disconnected a message informing the user of poor signal quality will be printed on the recording strip Measurement and interpretation results may be adversely affected by disconnected electrodes ...

Страница 115: ...DARD l II III aVR aVL aVF VI V2 V3 V4 V5 V6 CABRERA aVL l aVR II aVF III VI V2 V3 V4 V5 V6 SEQ NO 4 users can define a custom report sequence here Position the cursor on SEQ NO 4 Press The display shown below will appear Press Report sequence SEQ NO 4 Lead Label Chan 1 Chan 2 Chan 3 Chan 4 Chan 5 Chan 6 Chan 7 Chan 8 Chan 9 Chan 10 Chan 11 Chan 12 I II III AVR AVL AVF V1 V2 V3 V4 V5 V6 I II III AV...

Страница 116: ...bove steps for channel 2 etc You can write over SEQ NO 4 if you wish to enter another name for the report sequence Gain The unit will automatically determine the appropriate gain setting for the 6 simultaneous leads auto 5 10 20 40 mm mV with auto Speed Changes the paper speed Default 25 mm s Muscle Filter AC Filter Elimination of muscle artifact and AC line interference Muscle filter no AC line f...

Страница 117: ... System ADS In case of wandering baselines the anti drift system restores the baseline to its original position Signal delay with ADS approximately 2 seconds Default no Auto Paper Feed Before each recording the recorder feeds the paper to the beginning of a new page yes no ...

Страница 118: ...V6 STD_RED I II III V2 V4 V6 STD_LI I II III aVR aVL aVF CABR_LI aVL I aVR II aVF III HIGH_C V1 V2 V3 V4 V5 V6 C chest leads RED reduced number of leads LI limb leads Position the cursor on HIGH_C Press and create the report sequence as described in 6 Lead Mode section Gain The unit will automatically determine the gain setting auto 5 10 20 40 mm mV with auto Muscle Filter Elimination of muscle ar...

Страница 119: ...tomatically begins at program start no yes Arrhythmia Data The recorder will document arrhythmias in the following situations all unequal no All the recorder documents all arrhythmias that occur Unequal the recorder documents every arrhythmia that occurs that is different from the preceding event No arrhythmias are not documented Episodes Final report includes episode report with episodes listed b...

Страница 120: ...can be password protected Password protection is enabled in the system setup Please refer to the MAC 1200 Service manual for access information Before recording ECGs with the Pharma option enabled the unit must be set up accordingly When the Pharma option is enabled heart rate display text and printed text is labeled with Vent rate when the Pharma option is not enabled heart rate display text and ...

Страница 121: ...igned by pharma trial administrator at CRO Enter the 10 digit trial ID number Investigator ID Assigned by pharma trial administrator at CRO Enter the alphanumeric investigator ID number Table 1 Patient Data Entry Menu Pharma Parameter Factory Default Options Adjusted by Menu Item Displayed Project Code Pharma trial administrator project number Trial ID Pharma trial administrator trial ID Investiga...

Страница 122: ...trial administrator to enter up to 20 dose types which act as predefined dose types for user technicians The default value is Dose1 When Others is selected a drop down menu displays other preconfigured dose types Extra Questions Password protected adjusted by pharma trial administrator at CRO The pharma trial administrator can predefine up to five extra questions in addition to the existing four q...

Страница 123: ...he technician name field This requirement defaults to Yes on but it can be turned off in the Enable High Security Features menu Use the cursor keys to position the bar cursor on HIGH SECURITY and confirm the selection with Press again to enter the setup menu for the High Security option The setup menu for the High Security option will appear The High Security options available on the MAC 1200 unit...

Страница 124: ...0 unit after they have been transmitted to a MUSE CV system or CardioSoft CardioSys unit If a user attempts to delete an untransmitted record when this option is enabled the message Can t delete unsent data Refer to Operator s Manual is displayed The default value for Delete only after transmission is No Memory Full Messages 127 These messages display only when the High Security option is enabled ...

Страница 125: ...Name is Yes 419270 471680 bytes 32 39 Warning approaching ECG storage limit Transmit and delete ECGs to free memory To close this message Press key When ECGs 32 39 are acquired Every power up 471680 bytes 40 Memory is full When ECG 40 and all subsequent ECGs are acquired1 Every power up 1 The system will allow users to acquire print ECGs beyond the 40th ECG but these subsequent ECGs will not be st...

Страница 126: ...ted at power up 127 Physician and technician numbers must be unique the system will accept non unique IDs The Referring Physician is only relevant if ECGs are sent to the MUSE CV system This name will not be annotated on the ECG recording To exit the menu press Institution Name The name entered here will be printed on each report page Cart Number 4 place number default 1 range 1 to 9999 Site Numbe...

Страница 127: ...icate that the heart rate exceeds the limit value yes no only in 6 Lead and Arrhythmia modes The limit value 220 age can be changed manually 1 The value selected here is the default value appearing in the patient data 2 with CSI protocol only Lead Labels AAMI codes RA LA RL LL V1 to V6 or IEC codes R L F N C1 to C6 Pace Enhancement Pacer pulses are displayed enhanced and with a given amplitude yes...

Страница 128: ...in the selected period of time the display backlighting automatically switches off system default 20 min adjustment range l to 99 min Low Battery Beep Intermittent audible signal indicating that the battery requires charging The interval time for the tone can be selected between 5 and 60 seconds 0 s no audio signal Default mode This is the operating mode the unit defaults to after power up 12 Lead...

Страница 129: ...ter new password 3 Verify the new password Test Data Used for demonstration purposes yes It should be set to no default for proper clinical use Restore Defaults Selecting Yes will restore the factory default settings including the defaults of the three operating modes The electrocardiograph must be switched off and on again for the new settings to become effective Print Setup Lists Selecting Yes w...

Страница 130: ...o the prompt Enable the Print ECG Tx Log 2 Select Yes 3 Save this setting by pressing Check Record Retxn If a user attempts to transmit a record that has already been transmitted when the Check Record Retxn function is enabled a message displays informing the user that this record has already been transmitted The Check Record Retxn function defaults to No To enable the Check Record Retxn function ...

Страница 131: ...V system Modem Select the modem type Choose among the standard modems MultiTech MT 19 2 56k Elsa 14 4 28 8 33 6 56k and a user defined modem When using one of the standard modems enter the following The dial mode pulse or tone depending on your telephone network The telephone number 20 digits maximum The number to access the public telephone network e g 0 For a user defined modem enter the followi...

Страница 132: ...ction follow these steps 1 Within the Communication Setup window select the appropriate modem 2 Move the cursor to the prompt PIN Dialing 3 Select Yes 4 Enter information for the following fields Delay dd Service provider sssss PIN pppppppp Phone number xxxxx Outside line yyyyyy 5 Save this setting by pressing The Delay field hold two digits and its value ranges from 1 to 10 This has to be set by ...

Страница 133: ...d 1 Use the cursor keys to position the bar cursor on Patient Data Setup and confirm the selection with The patient data setup menu will appear 2 Select No for prompts that you want to remove from the dialog The following items cannot be removed Name First name Date of birth Patient ID 3 The following items are disabled They can be enabled from this menu ID required ID patient ID number Secondary ...

Страница 134: ... name ID required Secondary ID required 127 If Secondary ID is specified as a required data field make sure that it is not configured out of the patient data set up see Patient Data Menu Setup on page 9 27 for more information Last name required First name required Extra Questions 1 to 4 Enter any text here 10 characters maximum When you have entered the text you can select the format of the respo...

Страница 135: ...ond resting ECG MEMO storage of up to 40 resting ECGs C100 activates the three options MEAS DIAG MEMO for a maximum of 100 ECGs C500 activates the three options MEAS DIAG MEMO for a maximum of 500 ECGs EVAL activates the three options MEAS DIAG MEMO for a period of 4 weeks ERGO stress test MAC 1200 ST only Using the cursor keys position the bar cursor on the option that you want to activate Enter ...

Страница 136: ...ata to CardioSys CardioSoft or to MUSE Use the cursor keys to position the bar cursor on Protocol Select the protocol A5 if you will send data to CardioSys CardioSoft or select CSI to send data to MUSE Use the cursor keys to position the bar cursor on Modem other and confirm the selection with Choose the modem you use from the list and confirm the selection with If your modem is not included in th...

Страница 137: ...r on Communication and confirm the selection with Select the same baud rate as at the receiving modem 2400 4800 9600 19200 38400 57600 Selecting the Communication Protocol This adjustment is only necessary with MAC 1200 units because this device can send data to CardioSys CardioSoft or to the MUSE system 1 Using the cursor keys position the bar cursor on Protocol 2 Select the protocol A5 if you wi...

Страница 138: ...9 32 MAC 1200 Revision B 2012250 022 System Setup Direct ECG Transmission For your notes ...

Страница 139: ...Revision B MAC 1200 10 1 2012250 022 10 Loading Chart Paper ...

Страница 140: ...10 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 141: ...paper pad 4 Remove the cardboard from the top of the new pad and place the pad including the cardboard backing at the bottom and with the arrow pointing towards the unit into the paper compartment Inserting the New Z fold Pad Z X C V B N M L X K J H G F D S A alt Q 1 W 2 E 3 R 4 T 5 Y 6 U 7 I 8 O 9 P 0 setu p on off MAC 1200 049A Z X C V B N M L X K J H G F D S A alt Q 1 W 2 E 3 R 4 T 5 Y 6 U 7 I ...

Страница 142: ... it locks into place on both sides Closing the Paper Compartment Door 127 When inserting an already started Z fold pad the grid side must face up and the first fold must point toward the paper compartment Z X C V B N M L X K J H G F D S A alt Q 1 W 2 E 3 R 4 T 5 Y 6 U 7 I 8 O 9 P 0 sta rt sto p arrh y pat info stor e retri eve lead cop y setu p gain form at spe ed mus cle filte r 12 6 stan dby bat...

Страница 143: ...chnologies writer paper This paper has a special coating that prevents contamination and debris collecting on the printhead and electrostatic build up Furthermore the thermosensitive layer and the printhead characteristics are exactly matched Using other paper may result in recordings of poor quality Moreover the printhead may wear out prematurely Use of other paper may void the warranty A stripe ...

Страница 144: ...een 40 and 60 Avoid direct contact of the paper with carbon and carbonless forms chart papers and adhesives containing tributyl phosphate dibutyl phthalate or any other organic solvents and document protectors envelopes and sheet separators containing plasticizers Solvents or solvent based products containing alcohols ketones esters or other substances from this chemical group 87 21 The above comp...

Страница 145: ...Revision B MAC 1200 11 1 2012250 022 11 Cleaning Disinfection and Maintenance ...

Страница 146: ...11 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 147: ... it down with a cloth moistened with soap water Use a disinfectant for disinfection Do not immerse the cable in liquid Cleaning and Disinfecting the Electrodes In addition to the information given in this manual observe the instructions for use of the respective electrode types Discard disposable adhesive electrodes immediately after use to prevent that they are reused Clean reusable electrodes im...

Страница 148: ... by GE Medical Systems Information Technologies within the framework of a service contract The inspections include the following checks Visually inspect the device and the accessories for signs of mechanical damage that may impair the device functions Check that the device labeling relevant for safety is legible Run a performance test as described in the operator s manual Measure the resistance of...

Страница 149: ...Revision B MAC 1200 12 1 2012250 022 12 Troubleshooting ...

Страница 150: ...12 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 151: ...ify service to check and or replace fuse The recorder does not write over the entire paper width Paper compartment not properly closed Paper door must lock into place on both sides In Automatic mode the recorder does not stop and continues to feed paper This does not happen in Manual mode The paper pad was inserted the wrong way round so the recorder cannot detect the cue mark Insert the paper pad...

Страница 152: ...bleshooting Troubleshooting Chart 127 In the presence of very strong AC line interference in all leads the thermal printhead may interrupt the recording Activate the AC line filter 50 Hz 60 Hz in these situations Irregular AC Interference 054A ...

Страница 153: ...Revision B MAC 1200 13 1 2012250 022 13 Technical Specifications ...

Страница 154: ...13 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 155: ...h appropriate software documentation of analysis results in the respective operating mode Resolution of the recording Vertical 8 dots mm Horizontal 25 µm at 25 mm s Printer Paper CONTRAST ARCHIVIST Z fold pad 150 pages per pad equivalent to a chart length of approximately 45 m Paper width 210 mm or 8 5 inch 215 9 mm US format Sheet length 297 mm or 11 inch 279 4 mm US format To prevent damage to t...

Страница 156: ...and start stop function Lead Selection Manual selection of different lead combinations or automatic lead sequencing Lead programs configurable EINTHOVEN GOLDBERGER WILSON NEHB CABRERA lead sequence and user specific combinations in Manual and Arrhythmia modes Automatic Functions They assist and facilitate operation by the following Automatic amplifier blocking Automatic control of lead selection p...

Страница 157: ...eutral electrode 50 MW up to 60 Hz Dynamic range for differential signals between any two electrode connections for AC voltage 10 mV for superimposed DC voltage polarization voltage 600 mV Dynamic range for common mode signals referred to neutral electrode l V referred to chassis 263 V AC rms Quiescent input current via any electrode connection for l kW termination referred to neutral electrode 50...

Страница 158: ...tic discharge for 10 kV maximum Transfer of ECGs Transfer of ECGs with the CSI protocol between the MAC 1200 and the following units MUSE CVIS SW version 004A and later MAC 5000 SW version 001B and later MAC VU SW version 002A and later MAC 1200 SW version V5 01 and later Receiving Data with the CSI Communication Protocol from the Following Units CardioSmart SW version V4 21 and later CardioSmart ...

Страница 159: ...emote control connection depending on selected operating mode External make contact referred to chassis via circuit reference Source impedance Ri 300 Ω Contact dwell 100 ms Non destructive load 10 V ESD interface protection up to 10 kV Pin Assignment of Data Port 2 RDXE 3 TDXE 5 Circuit reference 6 Modem supply 8 RDXE 2 5 8 3 6 104A ...

Страница 160: ... Hz IEC Signal sampling rate 1000 per second Resolution referred to the input 5 µV Output rate to recorder 2000 per second For all leads gain adjustment in four steps 40 20 10 5 mm mV With active muscle filter low pass characteristic 3 dB drop of the amplitude frequency response at approximately 40 or 20 Hz With active AC line filter detection and compensation of periodic 50 Hz or 60 Hz frequency ...

Страница 161: ...tic adjustment range5 40 mm mV Amplitude range 6 channels 18 to 31 mm Baseline Automatic adjustment of the baseline to the optimal recording range in dependence of the signal amplitude Anti drift System ADS Automatic compensation of baseline fluctuations caused by polarization voltage fluctuations at the lead electrodes delay in recording 4 2 seconds ECG Storage In Automatic mode storage of approx...

Страница 162: ... PC configurable Test Automatic performance test upon power up including verification of the signal path starting at the signal input Stored test data for demonstration of the device functions Power Supply From the power line or from a built in rechargeable battery automatic switchover automatic battery charging during line power operation from integrated AC adapter module Mains Operation Instrume...

Страница 163: ...attery life approximately 5 years replacement by service only Battery Maintenance The battery in the MAC 1200 unit should be exercised regularly This means the battery should be allowed to drain between charging cycles If the ECG unit is left connected to wall power and or only used sporadically on battery power a discharge cycle should be performed on a monthly basis Turn the MAC 1200 unit on and...

Страница 164: ...ty between 25 and 95 Atmospheric pressure between 700 and 1060 hPa Transport and Storage Temperature between 30 and 60 C including battery Relative humidity between 25 and 95 Atmospheric pressure between 500 and 1060 hPa Recorder Dimensions Width 370 mm Height 95 mm Depth 320 mm include handle Weight Recorder with battery approximately 5 6 kg ...

Страница 165: ...Revision B MAC 1200 14 1 2012250 022 14 Order Information ...

Страница 166: ...14 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 167: ...unk cable 10 lead IEC MultiLink 223 418 09 Patient trunk cable 10 lead AHA MultiLink 384 018 16 Set of leadwires 4 mm connector 10 leads defibrillator proof IEC 384 018 17 Set of leadwires 4 mm connector 10 leads defibrillator proof AHA Electrodes 217 225 03 Adhesive electrodes for babies 13 mm diameter with 4 mm socket 60 cm lead fixation with adhesive rings 927 224 00 217 110 03 Adhesive electro...

Страница 168: ...on bulb with 4 mm socket 919 202 32 Clip electrode for limb leads package of 4 electrodes red yellow green black 303 442 96 Adapter for connection of electrodes with press stud to patient cable with 4 mm connector ECG Accessories NEHB 223 403 05 Patient trunk cable 12 leads NEHB IEC 384 017 65 Set of leadwires 10 leads defibrillation proof IEC with 4 mm plug 384 017 66 Set of leadwires 2 leads NEH...

Страница 169: ...Electrode paper 200 sheets for electrode 217 144 01 02 217 043 02 ECG filing cards 50 cards Electrode Application Systems 216 121 13 Electrode Application System KISS 10 10 lead system without pump 216 124 13 Electrode Application System KISS 10 10 lead system with pump 216 122 02 Electrode Application System KISS 12 12 lead system NEHB without pump 303 443 77 Swivel arm for KISS 384 015 84 Table ...

Страница 170: ...bal modem kit 223 362 03 Connection cable MAC 1200 to CardioSys 223 378 01 Connection cable MAC 1200 to modem 9 pin 223 378 02 Connection cable MAC 1200 to modem 25 pin 223 362 03 Connection cable EC 560 ECB 561 or MAC 1100 MAC 1200 to CardioSys 223 330 04 Connection cable M700 to MAC 1100 MAC 1200 223 366 04 Connection cable EC 1200 to MAC 1100 MAC 1200 2006795 001 Connection cable Variobike 500 ...

Страница 171: ...Revision B MAC 1200 A 1 2012250 022 A Entering Special Characters ...

Страница 172: ...A 2 MAC 1200 Revision B 2012250 022 For your notes ...

Страница 173: ...roke Combination Alt Q Alt W Alt E Alt R Alt Y Alt D Alt F Ç Alt G Å Alt K Ü Alt L î Alt X Ñ Alt C Alt V Ø Alt B Ö Alt N Ä Alt M Á Alt I then A É Alt I then E Í Alt I then I Ó Alt I then O Ú Alt I then U À Alt O then A enter È Ì Ò Ù in the same manner  Alt P then A enter Ê Î Ô Û in the same manner ÿ Alt U then Y enter Ï Ä Ö Ü in the same manner à Alt H then A enter Ñ Õ in the same manner Ð Alt A ...

Страница 174: ...A 4 MAC 1200 Revision B 2012250 022 Entering Special Characters Special Characters For your notes ...

Страница 175: ...Revision B MAC 1200 Index 1 2012250 022 Index ...

Страница 176: ...Index 2 MAC 1200 Revision B 2012250 022 ...

Страница 177: ...ion 4 4 Electrode application points 4 6 Electrode disconnected 5 5 EMC requirements 1 11 End of writer paper 10 5 Entering patient data 4 10 Explosion hazard 1 9 F Final report 7 8 9 13 Functional Description 1 5 G Global measurement points 5 22 H Heart rate limit 6 6 High Security Option 9 17 9 18 Host system 9 7 how to read label 1 14 I Indicators 2 3 Installation 3 4 Intended Use 1 5 L Languag...

Страница 178: ...Allowing 9 18 Recording problems 4 8 Referring physician 4 14 Report formats 5 10 Revision history 1 3 Rhythm leads 9 5 S Self test abort 3 5 serial number 1 14 service information 1 14 requirements 1 14 Software Version CE 2 Specifications 13 1 Standard leads 4 6 Suction electrodes 4 5 Symbols explanation 2 6 System defaults 9 23 System setup 9 20 System version CE 2 T T offset measurement point ...

Страница 179: ......

Страница 180: ...esentative GE Medical Systems Information Technologies GmbH Munzinger Straße 3 5 D 79111 Freiburg Germany Tel 49 761 45 43 0 Fax 49 761 45 43 233 Asia Headquarters GE Medical Systems Information Technologies Asia GE China Co Ltd 24th Floor Shanghai MAXDO Center 8 Xing Yi Road Hong Qiao Development Zone Shanghai 200336 P R China Tel 86 21 5257 4650 Fax 86 21 5208 2008 0459 gemedical com ...

Отзывы: