Quick Start 17
LOGIQ
500 Quick Start Guide
2276613–100 Rev. 0
Color Flow Mode (CFM) Controls
There are four Soft-Menu pages for Color Flow Mode imaging adjustments.
CFM Map—Allows for the selection of how Doppler velocities are
mapped as color over the gray scale.
Slant Scan—Used to control the position of the CFM window for linear
probes only.
Diag Mode—Allows for the selection of the best color flow display
format.
MTI Filter—Filters out low flow velocity color.
Frame Average—Averages color information from previous frames with
the current frame.
Penet.—Penetration can be increased by lowering the operating
frequency of the active probe.
High Resoltn—Provides a quick way to maximize resolution for the
CFM display.
Display Thrshld—Gray scale level at which the overlay of color
information stops.
Capture—Displays the highest mean velocity (average velocity)
detected over a specific time interval.