Respiratory Modules, E-sCAiOVX, E-sCAiOV, E-sCAiO, E-sCOVX, E-sCOV, E-sCO
8 - 5
2062973-004
PEEP
i
=intrinsic PEEP, PEEP
tot
-PEEP
e
Static pressure measurement maneuvers are automatically identified based on an increased
zero flow period at the end of the inspiration or expiration.
Static Compliance is calculated, if Static PEEP and Static P
plat
measurements were made within
a 2 minute period.
The following airway flow parameters are displayed:
−
Real time flow waveform (V')
−
Compliance (Compl)
−
Airway resistance (Raw)
−
Pressure volume loop
−
Flow volume loop
The measurement is based on measuring the kinetic gas pressure and is performed using the
Pitot effect. A pressure transducer is used to measure the Pitot pressure. The pressure signal
obtained is linearized and corrected according to the density of the gas. Speed of flow is
calculated from these pressure values and the TV value is then integrated. The MV value is
calculated and averaged using TV and RR (respiratory rate) values.
D-lite
Patient Spirometry uses specific sensors called D-lite+/D-lite and Pedi-lite+/Pedi-lite flow
sensors. Different types of sensors are available: adult sensor for measuring adults and
pediatric sensor for children. Both are available as reusable and disposable versions.
D-lite and Pedi-lite adapters are designed to measure kinetic pressure by a two-sided Pitot
tube. Velocity is calculated from pressure difference according to Bernoulli's equation. Flow is
then determined using the calculated velocity.
(from Bernoulli's equation)
Formula 1
where:
V’ = flow (l/min), v = velocity (m/s), A = cross area (m
2
), dP = pressure difference (cmH
2
O),
ρ
= density (kg/m
3
)
Finally, the volume information is obtained by integrating the flow signal.
Compliance and airway resistance
Compliance is calculated for each breath from the equation
Formula 2
Compliance describes how large a pressure difference is needed to deliver a certain amount of
gas to the patient.
The airway resistance, Raw, is calculated using an equation that describes the kinetics of the
gas flow between the lungs and the D-lite. The equation states that the pressure at the D-lite
can at any moment of the breath be approximated using the equation
Formula 3
Содержание E-PRESTN
Страница 2: ......
Страница 4: ...Module Frames and Modules 2062973 004 For your notes ...
Страница 6: ...Module Frames and Modules 1 ii 2062973 004 ...
Страница 10: ...Module Frames and Modules 1 4 2062973 004 For your notes ...
Страница 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual ...
Страница 12: ......
Страница 92: ...Module Frames and Modules 2 78 2062973 004 For your notes ...
Страница 94: ...Module Frames and Modules 2 80 2062973 004 For your notes ...
Страница 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual ...
Страница 96: ......
Страница 124: ...Module Frames and Modules 3 26 2062973 004 For your notes ...
Страница 125: ...4 Cardiac Output and SvO2 Module E COPSv Rev 01 Cardiac Output Module E COP Rev 01 Technical Manual ...
Страница 126: ......
Страница 138: ...Module Frames and Modules 4 10 2062973 004 2 Configuration There is no configuration for the E COP or E COPSv modules ...
Страница 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004 ...
Страница 160: ...Module Frames and Modules 4 32 2062973 004 For your notes ...
Страница 162: ...Module Frames and Modules 4 34 2062973 004 For your notes ...
Страница 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual ...
Страница 164: ......
Страница 198: ...Module Frames and Modules 5 32 2062973 004 For your notes ...
Страница 199: ...6 Masimo Module E MASIMO Technical Manual ...
Страница 200: ......
Страница 202: ...Module Frames and Modules 6 ii 2062973 004 ...
Страница 207: ...Masimo Module E MASIMO 6 5 2062973 004 2 Configuration There is no configuration for the E MASIMO module ...
Страница 218: ...Module Frames and Modules 6 16 2062973 004 For your notes ...
Страница 220: ...Module Frames and Modules 6 18 2062973 004 For your notes ...
Страница 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual ...
Страница 222: ......
Страница 224: ...Module Frames and Modules 7 ii 2062973 004 ...
Страница 236: ...Module Frames and Modules 7 12 2062973 004 ...
Страница 244: ...Module Frames and Modules 7 20 2062973 004 For your notes ...
Страница 246: ...Module Frames and Modules 7 22 2062973 004 For your notes ...
Страница 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual ...
Страница 248: ......
Страница 284: ...Module Frames and Modules 8 34 2062973 004 ...
Страница 316: ...Module Frames and Modules 8 66 2062973 004 For your notes ...
Страница 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual ...
Страница 320: ......
Страница 336: ...Module Frames and Modules 9 14 2062973 004 For your notes ...
Страница 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004 ...
Страница 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004 ...
Страница 380: ...Module Frames and Modules 9 58 2062973 004 For your notes ...
Страница 381: ...10 Single width Airway Module E miniC Technical Manual ...
Страница 382: ......
Страница 391: ...Single width Airway Module E miniC 10 9 2062973 004 2 Configuration There is no configuration for the E miniC module ...
Страница 397: ...Single width Airway Module E miniC 10 15 2062973 004 Complete on the Maintenance check form on page 10 36 ...
Страница 410: ...Module Frames and Modules 10 28 2062973 004 ...
Страница 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual ...
Страница 420: ......
Страница 422: ...Module Frames and Modules 11 ii 2062973 004 7 2 Entropy Module E ENTROPY 01 11 17 Maintenance check form 11 19 ...
Страница 427: ...Entropy Module E ENTROPY Rev 01 11 5 2062973 004 2 Configuration There is no configuration for the E ENTROPY module ...
Страница 442: ...Module Frames and Modules 11 20 2062973 004 For your notes ...
Страница 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual ...
Страница 444: ......
Страница 446: ...Module Frames and Modules 12 ii 2062973 004 7 3 EEG Headbox N EEG 12 20 Maintenance check form 12 23 ...
Страница 452: ...Module Frames and Modules 12 6 2062973 004 2 Configuration There is no configuration for the E EEG module ...
Страница 468: ...Module Frames and Modules 12 22 2062973 004 For your notes ...
Страница 470: ...Module Frames and Modules 12 24 2062973 004 For your notes ...
Страница 471: ...13 BIS Module E BIS Rev 01 Technical Manual ...
Страница 472: ......
Страница 480: ...Module Frames and Modules 13 6 2062973 004 2 Configuration There is no configuration for E BIS module ...
Страница 494: ...Module Frames and Modules 13 20 2062973 004 For your notes ...
Страница 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual ...
Страница 496: ......
Страница 504: ...Module Frames and Modules 14 6 2062973 004 2 Configuration There is no configuration for the E NMT module ...
Страница 518: ...Module Frames and Modules 14 20 2062973 004 For your notes ...
Страница 520: ...Module Frames and Modules 14 22 2062973 004 For your notes ...
Страница 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual ...
Страница 522: ......
Страница 532: ...Module Frames and Modules 15 8 2062973 004 2 Configuration There is no configuration for the F5 and F7 module frames ...
Страница 558: ...Module Frames and Modules 15 34 2062973 004 For your notes ...
Страница 560: ...Module Frames and Modules 15 36 2062973 004 For your notes ...
Страница 561: ...16 Patient Data Module v2 Technical Manual ...
Страница 562: ......
Страница 620: ...Module Frames and Modules 16 56 2062973 004 ...
Страница 621: ......