GASMAX II Operation & Maintenance Manual, Revision 7.0
Page 43
Coefficient
Electrochemical Sensor Type
Default EUNITS
Negative
Negative
Negative
Negative, Bias
Negative, Bias
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Negative
Positive
Positive
Positive
Hydrogen Sulfide
Oxygen
Carbon Monoxide
Ammonia
Nitric Oxide
Ethylene Oxide
Hydrogen Chloride
Hydrazine
Arsine
Sulfur Dioxide
Hydrogen
Hydrogen Cyanide
Phosgene
Phosphine
Hydrogen Fluoride
Nitrogen Dioxide
Ozone
Chlorine
ppm H2S
% Oxygen
ppm CO
ppm NH3
ppm NO
ppm Eth O2
ppm HCL
ppm N2H4
ppm Arsine
ppm SO2
ppm H2
ppm HCN
ppm COCl2
ppm PH3
ppm HF
ppm NO2
ppm Ozone
ppm Cl2
ZERO CAL VALUE
should always be set to “ ”.
RAW MIN / MAX COUNTS
is factory preset and should not be changed.
MARKER ACTIVE
enables an input level monitor on the channel 2 input if the input is set to accept a 4-
20mA signal. If Marker Active is set the YES, the following actions may be programmed to occur when the
input signal matches the Marker Level setting.
MARKER TX LED
will cause the TX LED to illuminate if the 4-20mA input matches the Marker Level
setting.
MARK LEVEL
is the value at which the Marker is active. The input must be /- 1% of this value to
be active. Setting is in mA.
MARK AS MESSAGE
is a text message string that appears on the LCD when the input level matches the
Marker Level setting.
Channel Floating Point
bit plus sign ’s complement with
-5% over / under range applied.
Contact GDS Corp for more information.