Gas Detection (Australia) Pty Ltd
GTD 2000 Ex Operating Manual
Page 6 of 26
M-GTD2000Ex-01-01
EMC (electromagnetic compatibility)
Comply with EN50270
3.5 Electrical
Supply voltage
18VDC/31VDC
Supply voltage ripple & noise max
1Vpp
Current consumption (typical /max)
150mA/250mA@24VDC
Analog output current range
0-22.0 mA
Gas concentration output current range
4-20 mA
Analog output current ripple & noise max
20uApp
Analog output termination resistance
(including total cable resistance)
< 600Ω
3.6 Cross sensitivity
Combustible gases
Cross sensitivity (% relative to CH4)
Methane
CH
4
100
iso-Butane
C
4
H
10
80
Hydrogen
H
2
110
Acetylene
C
2
H
2
100
Ethylene
C
2
H
4
95
n-Pentan
C
5
H
12
70
Propane
C
3
H
8
75
n-Hexane
C
6
H
14
65
n-Heptane
C
7
H
16
55
IPA
C
3
H
7
OH
60
Methanol
CH
3
OH
90
Ethanol
C
2
H
5
OH
70
Acetone
(CH
3
)
2
CO
70
MEK
C
2
H
5
COCH
3
70
n-Octane
C
8
H
18
70
Ethyl acetate
CH
3
CO
2
C
2
H
5
60
Buthyl acetate
C
6
H
12
O
2
5
* Please inquire to GDA for other gases