EN
Screw in
Release
Release
Screw in
EMD mit MVA-G + RWA
EMD mit MVA-R + Platte
Installation procedure:
•
Centralize the seal using the pin on the m. connection.
•
Place the seal on the pin of the wall adapter.
•
Screw the shut-off valve onto the threaded connection by hand. Keep the pressure reducer at
an angle of 45° in front of the desired position.
•
Tighten the lockable quick-fitting coupling with the open-end wrench. To do this, use only
the open-end wrench SW 36 which is available as an accessory. As the tightening torque in-
creases, the pressure reducer is rotated as well. Tighten the quick coupling until the pressure
reducer reaches its desired position. This will give a gas-tight connection.
WARNING!
•
With C
2
H
2
use only compression fitting and pipe lines made from stainless steel
6.2. USING THE SHUT-OFF VALVE
Open and close the valve by turning the operating lever. Turn the lever by 90°. The lever latches
easily into an opened or closed position.
To close valve:
•
Turn the lever to the right.
•
The lever and the word “OFF” is now crosswise to the connections.
To open valve:
•
MVA 3100 G: Turn the lever to the left.
•
MVA 3100 W: Turn the lever to the left.
•
The lever and the word “ON” is now pointing in the direction of the connections.