![GCE druva LAB 3100 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/gce-druva/lab-3100/lab-3100_instructions-for-use-manual_3876153028.webp)
28/36
EN
Installation procedure:
RWA with cover rosette
Drilling pattern RWA
Fitting the Rear wall adapter (RWA) Insert a seal into the wall adapter. Screw the screw adapter
tightly into the wall adapter using the hexagon socket head wrench.
WARNING!
•
Screw the side without centralizing pin into the wall adapter!
Installing the pressure reducer:
Centralize the seal using the pin on the male connection. Place the seal on the pin of the wall
adapter. Screw the EMD 3100 onto the threaded connection by hand.
Keep the pressure reducer at an angle of 45° in front of the desired position. Tighten the
lockable quick fitting coupling with the open-end wrench. Only use the open-end wrench
SW 36 (order no. H 11006002) which is available as an accessory. As the tightening torque
increases, the pressure reducer is rotated as well. Tighten the quick coupling until the pressure
reducer reaches its desired position. A gas-tight connection is achieved.
Screw in
Release
Mounting EMD directly on RWA
WARNING!
•
With C
2
H
2
use only compression fittings and pipelines made from stainless steel!
Содержание LAB 3100