Command name
CMD_SET_REPEAT
Description
The command sets the number(rep: 0 ... 65535) of sweeps for repetitive measurement.
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
rep
0
end flag
Byte String (HEX)
A5
5A
4A
00
rep
l
rep
h
00
00
00
00
B9
9B
Used setup value
None
Affected setup value
Repeat
MCA state
STATE_READY
Command name
CMD_SET_MCS_CHANNEL
Description
The command set the number of channels(ch: 128 ... 4096) for MCS mode
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
ch
0
end flag
Byte String (HEX)
A5
5A
63
00
ch
l
ch
h
00
00
00
00
B9
9B
Used setup value
None
Affected setup value
mcs_channels
MCA state
STATE_READY
Command name
CMD_SET_TIME_PER_CHANNEL
Description
The command sets the dwell time(tpc: 1 ... 65535) per channel for the MCS mode
40
.
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
tpc
0
end flag
Byte String (HEX)
A5
5A
4B
00
tpc
l
tpc
h
00
00
00
00
B9
9B
Used setup value
None
Affected setup value
time_per_channel
MCA state
STATE_READY
40 It is the real time per channel multiplied with 10
61