Technical Data
Parameter
MCA527
MCA527L
MCA527OEM
ADC
Sample Rate
10MSps
Resolution
14bit
Integral Non-Linearity
≤0.05%
Temperature Stability
TK50
Digital Signal Processing
Trigger Filter
single and double differential trigger filters are selectable
Pile-Up Suppression
yes
Pulse Pair Resolution
~400ns, depending on trigger filter
Trigger Threshold Adjustment automated (default) and manual
Shaping Time
0.1µs to 25µs
0.1µs steps
0.1µs to 2µs
0.1µs steps
Flat Top Time
0µs to 5µs, changeable in steps of 0.1µs
Fine Gain Adjustment
0.5 to 6.5, changeable in steps of 0.01%
Channel Splitting
128, 256, 512,
1024, 2048, 4096,
8192, 16384
128, 256, 512, 1024, 2048
Differential Non-Linearity
≤1% for 4096 channels and 1µs shaping time
Base Line Restorer
base line restorer with adjustable averaging
Pole Zero Adjustment
40µs to 1ms detector decay time constants can be
compensated, best spectroscopic performance is
achieved with 50µs decay time constant detectors
Peak Stabilization
standard mode,
LED mode,
temperature
mode
standard mode
Modes of Operation
PHA (Pulse Height Analysis)
yes
MCS (Multichannel Scaling)
yes
Sample Mode (Oscilloscope)
yes
Gated Mode
yes
not available
59