3.2.12 Jitter Correction
Another approach to care about the timing error problem is to use for exact timing the
1, -2, 1 trigger filter and to interpolate the zero crossing of the second derivate. This
should increase the timing accuracy to something like 10ns...20ns. The amplitude is then
correspondingly interpolated. Jitter correction may give a resolution improvement in some
cases, but it is still experimental. Default is off, this feature is not available with MCA527L.
40
Figure 23: Measurement same as Figure 19, PZC=0 but with jitter correction.