Multichannel Scaling (MCS)
4 Multichannel Scaling (MCS)
This mode is for semi-automated measurements of time distributions with any radiation
detector (HPGe, NaI, CdTe, CdZnTe and neutron counters). The software used for this is
WinMCS.
It allows to measure a time distribution of count rates, and in case of a spectroscopical
detector, measurement of an integral spectrum at the same time.
Input can be set to the following:
•
TTL input (no spectrum possible in this case)
•
Input rate (corresponds to the fast count rate of MCA measurements)
•
LLD/ULD (corresponds to the content of a partial region of a spectrum, defined by
the lower end LLD and the upper end ULD.
New compared to MCA166 is that the MCA527 principally supports dwell times down to
100µs, whereas the MCA166 was limited to 10ms minimum. Also the MCA527 allows up
to 16k time channels, where the MCA166 was limited to 4k.
However, 100µs dwell time and 16k channels will be only supported with WinMCS
versions >2.0.
49