7
7
UV PREVENTATIVE MAINTENANCE CONTINUED
SAFETY CONCERNS WITH UVC
11. Check all wires for damage
Replace any wires showing damage
•
12. Lubricate each of the new UV lamp(s) before installing with a
small amount of G-510 or similar detergent
13. Re-connect the lamp connectors
Each lamp connector should be labeled from 1 to 4
•
Lamp #1 is at the Front of the ventilator when the UV
•
Module is installed
Lamp #4 is at the Back of the ventilator when the UV
•
Module is installed
14. Re-install the UV Module end cap and torque the bolts to
7-10 in-lbs.
15. Re-install the UV Module in the ventilator
17. Tighten all quarter-turn screws holding the UV Module
cover to the ventilator
16. Re-connect the UV Module power coNNECTOR
18. Check for proper operation of UV lamps & pressure
switches
Warning:
Do
NOT
defeat any interlocks during cleaning and maintenance!
CERTIFIED SERVICE AGENT MAINTENANCE WARNING!
FIGURE 12
As with many types of technology if it is not used properly and/or
proper precautions are not taken there is the potential for injury
or harm. This is especially true with UVC light due to the fact that
it does not physically hurt at the time of exposure. While UVC is
very effective at breaking down grease molecules, direct exposure
C
V
U
f
o
t
n
u
o
m
a
e
h
T
.
s
e
y
e
d
n
a
n
i
k
s
o
t
l
u
f
m
r
a
h
s
i
s
t
n
u
o
m
a
e
g
r
a
l
o
t
generated in these hoods is greater than that what results from
direct exposure to the sun. Under no circumstances is it acceptable
to view the lighted lamps without proper eye protection or expose
bare skin directly to the light. All interlocks and safety precautions
called for in this manual must be followed to avoid the potential for
harm to service personnel and/or operators. In addition, only trained
and authorized personnel may perform some maintenance See
previous page for details.
Personal Protective Equipment
Eye protection that prevents 100% of UVC being transmitted
1.
through the lens must be worn at all times when performing
service work on any ELX-UV that is energized and/or has
the potential to be energized and expose personnel to UVC
light.
Whenever service work is performed it is recommended that
2.
long sleeve pants and shirts be worn to minimize the potential
for inadvertent exposure of the skin to UVC.
Safety Interlocks
This product comes equipped with redundant pressure switches to
ensure the Exhaust Fan is running, and that every filter is in place.
All of the switches must be working and/or adjusted properly in
order for the system to operate properly.
Содержание ELX SERIES
Страница 12: ...9 9 GFBD INTERNAL WIRING SINGLE SECTION FIGURE 16 GFBD GEBD Internal Wiring Diagram ...
Страница 13: ...10 10 GFBD EXTERNAL WIRING MULTIPLE SECTION FIGURE 16B GFBD External Wiring Diagram ...
Страница 25: ...22 20142 20143 20163 20139 20140 19289 20144 76047 UV MODULE PARTS ...
Страница 26: ...23 UV HOOD WIRING BOARD SCHEMATIC ...
Страница 28: ...25 ELX GBD OR ND NO UV WIRING DIAGRAM ...
Страница 29: ...26 ELX GFBD NO UV WIRING DIAGRAM ...
Страница 30: ...27 ELX UV GBD ND WIRING DIAGRAM ...
Страница 31: ...28 ELX UV GFBD WIRING DIAGRAM ...
Страница 32: ...29 ELX UV GFBD WITH GPC 6000D WIRING DIAGRAM ...