background image

 

8

the instrument is located in fresh air.  Oxygen 
sensor models should show 20.9 in fresh air. 

   

or

   

 

2.1.1  Calibration due lockout 

When the ToxiPro’s calibration due lockout 
function is enabled and calibration is due, the 
calibration due warning will be shown at 
instrument start up and can not be bypassed.  
The instrument must be calibrated 
immediately before it will show any gas 
readings.  The calibration may be performed 
manually, or by placing it in an IQ Express 
Dock.  Upon successful calibration, the 
ToxiPro will proceed to the current gas 
readings screen. 
The calibration due lockout function can be 
enabled or disabled with BioTrak software. 

2.2 Backlight 

The ToxiPro includes a backlight that is 
automatically turned on during an alarm 
condition.  The backlight can also be 
manually activated while the current gas 
reading is displayed by pressing the MODE 
button once.  When manually activated, the 
backlight will automatically turn itself off in 
about twenty seconds.  If activated by an 
alarm condition, the backlight will remain on 
until the instrument is no longer in alarm. 

2.3 Functions 

From the current gas 
reading screen, press the 
MODE button once to 
activate the backlight.  
Press MODE again to 
scroll through the screen options.   
Once the backlight has been lit, press MODE 
once more to view the MAX gas values 
screen.  The MAX figure represents the 
highest gas value reading that has been 
recorded by the instrument during the current 
operating session.  ToxiPro O

2

 models will 

sequentially display both the highest and the 
lowest readings that have been recorded by 
the instrument in the current operating 
session. 
From the MAX screen, press the MODE 
button to view the time screen.  A dot below 
the colon between the hours and the minutes 
digits indicates afternoon/evening hours (pm). 

         

 

10:45 am                    10:45 pm 

If the STEL (Short Term Exposure Limit) 
alarm is enabled, press the MODE button 
once to view the STEL reading.  The STEL 
value displayed represents 
the average value of the 
instrument readings for the 
target gas for the most 
recently completed 15 
minutes of operation.

 

If the TWA (Time Weighted Average) alarm is 
enabled, press the MODE button once more 
to view the current TWA value.  TWA values 
are calculated by taking the sum of the 
instrument readings for the 
target gas for the current 
operating session in terms 
of parts-per-million-hours 
and dividing by an eight-
hour period. 

Note:  Due to the nature of the TWA 
calculation, the TWA value can 
accumulate over time and may cause the 
instrument to go into alarm.  Sperian 
Instrumentation recommends resetting the 
TWA value at the beginning of any work 
shift by turning off the instrument and 
then turning it back on again. 

Press the MODE button again to return to the 
current gas readings screen. 

2.4 

Calibration and bump test due 
notices 

When the calibration due notice is enabled 
and the ToxiPro is due for fresh air calibration 
“0-CAL will be displayed at the bottom of the 
current gas readings 
screen along with the 
triangular warning symbol.   
When the calibration due 
notice is enabled and the 
ToxiPro is due for span 
calibration the calibration 
bottle icon will be displayed 
at the bottom of the current 
gas readings screen along 
with the warning symbol. 
The bump test due notice 
is designed for instruments 
that are processed in an IQ 
Express Dock.  When the 
bump test due notice is 
enabled and the ToxiPro is 
due for a bump test, the calibration bottle 
icon will be displayed at the bottom of the 

Содержание ToxiPro

Страница 1: ...ToxiPro S ing le G as Detec tor R eferenc e M anual GasTech Australia Pty Ltd 24 Baretta Rd Wangara Western Australia 6065 Tel 1800 999 902 Fax 1800 999 903 http www gastech com au ...

Страница 2: ...E INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro Reference Manual Sperian Instrumentation Part Number 13 264 Version 2 01 Copyright 2008 by Sperian Protection Instrumentation LLC Middletown Connecticut 06457 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Sp...

Страница 3: ...kout 8 2 2 Backlight 8 2 3 Functions 8 2 4 Calibration and bump test due notices 8 2 5 Low battery alarms 9 2 6 Turning the ToxiPro off 9 2 7 Always On Mode 9 2 7 1 Turning Off in Always On Mode 9 2 8 Sampling 9 2 8 1 Sample draw kit usage 9 3 Calibration 10 3 1 Verification of accuracy 10 3 1 1 Verifying accuracy and response ToxiPro O2 10 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor 10...

Страница 4: ...erian Instrumentation Warranty Gas Detection Products 18 Operating Temperature Range The safe operating temperature range of the gas detector is printed on the back label of the instrument Use of Sperian Gas Detectors outside of the instrument s specified operating temperature range may result in inaccurate and potentially dangerous readings Certifications The ToxiPro carries the following certifi...

Страница 5: ...accurate and potentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air 5 A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrume...

Страница 6: ... B 1 2 2 ToxiPro O2 sensor ranges The oxygen sensor used in the ToxiPro O2 has a range of 0 30 by volume 1 2 3 ToxiPro O2 warm up phase The ToxiPro O2 requires a one time 15 minute warm up phase prior to initial activation See section 2 1 for further details 1 3 Alarm and warning logic 1 3 1 Gas alarms ToxiPro gas alarms are user adjustable and may be set anywhere within the range of the specific ...

Страница 7: ...n A built in manually activated backlight allows the display to be read even in low light conditions Visual alarm LED IrDA port A bright red LED Light Emitting Diode alarm light provides a visual indication of the alarm state The LED also functions as the IrDA port MODE button The large push button on the front of the instrument is called the MODE button The MODE button is used to turn the ToxiPro...

Страница 8: ...hat the 6 digit serial number can not fit on a single screen so it is shown on two screens In this case the instrument serial number is 481098 Datalogging versions will then show the dL screen Once the sensor is recognized the instrument will display the sensor type The warning alarm level will then be displayed followed by the danger alarm level During the display of the warning alarm level the L...

Страница 9: ...the MAX screen press the MODE button to view the time screen A dot below the colon between the hours and the minutes digits indicates afternoon evening hours pm 10 45 am 10 45 pm If the STEL Short Term Exposure Limit alarm is enabled press the MODE button once to view the STEL reading The STEL value displayed represents the average value of the instrument readings for the target gas for the most r...

Страница 10: ...re Note The ToxiPro must be turned off to replace the battery See section 2 7 1 2 7 1 Turning Off in Always On Mode BioTrak software must be used to turn the ToxiPro off when it is in Always on Mode BioTrak is available at http www biodownloads com 1 Open BioTrak and select ToxiPro 2 Select Configure in the software 3 Hold the MODE button down for about 10 seconds until IrDA is shown and establish...

Страница 11: ...ds then slowly exhale directly onto the face of the sensor in the same way you would to fog up a piece of glass If the descending oxygen alarm is set to 19 5 the instrument should go into alarm after a few seconds 3 1 2 Verifying accuracy ToxiPro with toxic gas sensor Verification of accuracy is a two step procedure for ToxiPro instruments equipped with a toxic gas sensor Step one is to take the T...

Страница 12: ... between the following two screens 3 For instruments equipped with a toxic gas sensor the fresh air zero calibration is complete when the instrument begins a second 5 second countdown for the span calibration If span calibration is not required allow the countdown to reach 0 without pressing the MODE button For further instructions concerning the span calibration of toxic sensor equipped models pr...

Страница 13: ...ibration is fully automatic from this point on Once the instrument successfully completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach ...

Страница 14: ...an calibration failure in which calibration gas is not detected the no and GAS screens will be alternately displayed as shown below with the calibration bottle icon highlighted When the instrument returns to the current gas readings screen the warning symbol and the calibration bottle icon will be shown which signifies that the last attempt to span calibrate the instrument has failed Note If the C...

Страница 15: ...utomatically restart with the hours digits blinking The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once the appropriate hour setting is shown wait 5 seconds and the minutes sett...

Страница 16: ... instrument functionality 4 4 Storage ToxiPro detectors may be stored for long periods in a fresh air environment at temperatures between 10 C 50 F and 30 C 86 F 5 PC Instrument Communications 5 1 Event logging Each ToxiPro includes a built in event logger that stores instrument readings during alarm conditions calibration values and other data BioTrak software and an IrDA port are necessary to do...

Страница 17: ... 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d n d no data undetermined posi...

Страница 18: ...mmends the following procedure to establish a safe and prudent accuracy check schedule for your Sperian instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that the sensors are exposed...

Страница 19: ... is in service ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Toxi3Ltd 3 years after activation or 3 years after the Must Be Activated By date whichever comes first Mighty Tox 2 Prorated credit is given towards repair or purchase of a new unit of the same type 0 6 months of use 100 credit 6 12 mon...

Страница 20: ...xiPro c UL Version ATEX CQST Version IECEx Version Type O2 H2S SO2 CO CO H CO NH3 Cl2 ClO2 NO2 HCN PH3 SO2 Hi Software Version _____ Vibrating Motor Datalogging Belt Clip Serial Number ____________ MFG Date ...

Отзывы: