15
1. Follow the directions in section 4.1 steps
1-4 to remove the battery.
2. The sensor is located to the left of the
LED alarm light on the main board.
Gently remove the old sensor and install a
new sensor of the same type.
3. Reinstall the battery and follow the
remaining steps in section 4.1 to reset the
instrument’s time and date settings.
4. New sensors must be allowed to stabilize
prior to use according to the following
schedule. The detector must be powered
off and functional batteries must be
installed for the sensors to stabilize.
Instrument Stabilization
Period
ToxiPro O
2
1
hour
ToxiPro with
toxic gas sensor
15 minutes
5. Once the sensor has stabilized, calibrate
the detector. For ToxiPro O
2
units,
perform the Fresh Air/Zero Calibration as
described in section 3.3. For ToxiPro
units with toxic gas sensors, perform both
the Fresh Air/Zero calibration (section 3.3)
and the Span calibration (section 3.5).
Note: The ToxiPro must be calibrated
after any sensor change. See section 3.3
and 3.5 above for details.
4.3 Proper
Cleaning
The exterior surfaces of the ToxiPro may be
cleaned using a damp cloth only. Do not use
cleaning agents of any kind. The introduction
of cleaning agents to the detector may affect
instrument functionality.
4.4 Storage
ToxiPro detectors may be stored for long
periods in a fresh air environment at
temperatures between 10
°
C/50
°
F and
30
°
C/86
°
F.
5. PC-Instrument
Communications
5.1 Event
logging
Each ToxiPro includes a built-in event logger
that stores instrument readings during alarm
conditions, calibration values and other data.
BioTrak software and an IrDA port are
necessary to download the collected data to a
PC. See section 5.3 for instructions on
initiating communications.
5.2
Instrument firmware Upgrades
The ToxiPro’s instrument firmware may be
upgraded at any time with a PC through the
instrument’s IrDA port. See section 5.3 for
instructions on initiating communications.
For the latest version of the instrument
firmware see the Sperian Instrumentation
download website at
http://www.biodownloads.com
5.3 Initiating
communications
To initiate communications,
hold the MODE button as if
you are turning the instrument off. Once
“OFF” or “on” is shown, continue to hold
MODE until the “IrdA” screen is shown and
the LED/IrDA port turns on.
Once “IrdA” is shown release the MODE
button. Then align the infrared port at the top
front of the ToxiPro with the PC’s infrared port
to proceed with the download. The LED/IrDA
port will blink during communication with the
PC. The ToxiPro should be located 1-2” (2.5-
5cm) from the IrDA Port.