Tips if Calcmet does not IDENTIFY any gases - 2
B. Calcmet Identification Tool Settings
All Identification settings must be accessed in Calcmet
EXPERT software (refer slide#16 to move from
Calcmet EASY
→
EXPERT)
Suggestions if no gases identified :
1. Change “
Limit HQI <
” value to 60.00 and re
-run the
IDENTIFY function.
2. Uncheck “
do not search in NIST Library
” and re
-run
the IDENTIFY function.
3. Measure another sample or Open previously taken
sample and re-run the IDENTIFY function.
If no success with identifying the Unknown it is possible
that this component(s) is not in the reference library.
Suggest contact Gasmet or local representative.