CHECKING BACKGROUND
Check these two parameters :
1
). Max value _ Pass if reading > 40,000
2
). Y-intercept _ Pass if reading > 10,000
Contact Gasmet or your representative if
there is a Fail for either above parameters.
3
). The CO2 band should not reach the x-axis.
(refer line 3.above)
If this fails _ Repeat Background but first check following :
a. Is
(N
2
) gas
turned on & contain pressure ?
b. Is
(N
2
) gas
connected to GT5000 zero gas connector ?
This message is displayed
if background successful