14
TROUBLESHOOTING GUIDE
Symptom /
Issue
Possible Cause / Resolution
Pump does not
ACTIVATE
If the user experiences issues with pump activation, press/hold the
MODE key to enter into settings screen. Click ENTER to access 2/2
and modify
“
Flow Rate Level
”
to
“
5
”
. If pump is still inactive, pump
needs to be replaced
–
contact
Device does not
power ON
Make sure the power supply is connected properly and that
there is power in the outlet the adapter is connected has the
appropriate voltage. If using batteries, make sure they are not
depleted.
Slow
response
Check the air flow channels to make sure they
are not obstructed.
Reading Error
Ensure that the Inlet/Outlet Luer Cap Fittings are
removed
to
enable an accurate means of sampling.
Red SD Card Icon
Appears
This indicates that the SD Card is either damaged,
missing, or full of data. Replace SD Card.
LCD Display Turns
Off During
Measurements
If the user experiences LCD display turning off during
calibration, measurements, or data logging
–
please
reference the
“
settings
”
section to adjust backlight / auto
power off from default settings.
LCD Screen
Troubleshooting
During Use
If the user experiences issues with the screen orientation,
this setting can be changed by referencing the
“
settings
”
section and adjusting the
“
screen auto rotate
”
to
“
NO
”
.
Rechargeable Batteries Index
‘Battery OK’:
Measurements
are possible
‘Battery Low’:
The battery needs to be
recharged, measurements are
still possible
‘
Battery
Exhausted’:
Measurements are
not possible
Battery installation:
This device is supplied with Li-ion 18650 3.7V rechargeable batteries (3x), which are pre-installed upon device shipment.
Battery charging:
During battery charging, the temperature of the device will rise by 5°C~10°C. At this time, the measurements of
temperature and humidity will be affected by temperature rise. This could cause an effect on the accuracy of
temperature when charging. Please use a fan to blow toward the Temperature & Humidity Sensor (B )
directly, in order to gain a compensated balance of temperature and humidity between that of the
temperature sensor and surrounding area. (5V/1A USB adapter charger).