Troubleshooting 

Only manufacturing firm and authorized service workshops may work in the 

biovac 106 

Aspiration System 

and accessory components during the warranty period. 

Trouble 

Cause 

Remedy 

by: 

with: 

Vacuum pump 
does not start 

no power supply 

Qualified 
electrician 

Check electrical installation 

Motor defective 

Service 
 workshop 

Exchange 

Pump body defective 

Repair and/or exchange 

Vacuum pump 
does not  
generate a 
vacuum or 
only an 
inadequate 
one 

Connected apparatus 
and/or 
connecting elements 
leaking 

User or 
Service 
workshop

 

Identify and seal the leak, 
replace the seals and/or 
hoses if necessary. 

Vacuum pump leaking 

Check the hose  
connections between the 
pump heads, replace the 
hoses and/or fittings if 
necessary. 

Pump head leaking 

Service 
workshop 

Repair and/or exchange 

Diaphragm defective 

User or 
Service 
workshop 

Exchange of the diaphragm 

(see chapter 6.2)

 

Valve defective 

Exchange of the valve 

(see chapter 6.2)

 

Vacuum pump dirty 

General maintenance / 
leaning 

Valves dirty 

Cleaning condensates  
and foreign objects out of 
the valves. 

Running noise 

Vacuum pump dirty 

User or 
Service 
workshop 

General maintenance / 
cleaning 

Glass 
components 

defective 
and/or leaking 

User

 

Exchange of the glass parts 
or seals 

Cable(s) 

defective 
and/or brittle 

Qualified 
electrician

 

Exchange of the cable(s) 

Содержание biovac 106 with 2 L Container

Страница 1: ...Operation Manual EN Translation of the german original manual Aspiration Systems Models biovac 106 with 2 L Container biovac 106 with 4 L Container 112037 2017 04 10...

Страница 2: ...ritten permission of Co Gardner Denver Thomas GmbH All rights under the copyright laws are expressly reserved by Co Gardner Denver Thomas GmbH We reserve the right to make changes and amendments Gardn...

Страница 3: ...eration 14 5 4 Storage 15 5 5 Scrap Disposal 15 6 Maintenance and Servicing 16 6 1 General Requirements 16 6 2 Maintenance Performed by the User 16 6 2 1 Maintenance of the aspiration systems 16 6 2 1...

Страница 4: ...s must have the technical competence required to perform the work that has to be done The user must authorize the operating personnel to do the work that has to be done The personnel must have read an...

Страница 5: ...ring method grade 2 DIN EN 60204 1 2014 10 Safety of machinery Electrical equipment of machines Part 1 General requirements DIN EN 61000 6 2 2011 06 DIN EN 61000 6 4 2011 09 Electromagnetic compatibil...

Страница 6: ...se of portable devices The local regulations apply in foreign countries Please note the following when connecting to the electrical power supply system The electrical power supply system must have a p...

Страница 7: ...ent such sub stances from leaking out of the appliance or pump Treat these substances according to the applicable environmental protection regulations The diaphragm pump control valves and hose lines...

Страница 8: ...a Aspiration System biovac 106 with secretion glass container 2 liters A quick coupling 4 is used for the connection to the diaphragm pump 2 It is provided with a stop valve The used resistant diaphra...

Страница 9: ...hragm pump is used for the safe and precise aspiration of non flammable chemical and biological liquids The fluids can be removed easily and very precisely from slides Petri dishes cell culture contai...

Страница 10: ...Application The biovac 106 Aspiration Systems are intended for the safe and precise extraction and catching of liquid and non explosive excess fluids use in the chemical biological and medical areas g...

Страница 11: ...Technical Data 112037 11 4 Technical Data 4 1 Dimensions Fig 2 Dimension biovac 106 with secretion glass container 2 liters Fig 3 Dimension biovac 106 with secretion PP container 4 liters...

Страница 12: ...mode S 1 Type of protection DIN EN 60529 IP 40 Class of insulation DIN EN 600034 1 F 160 C Dimensions W D H without hose mm 342 363 197 395 363 197 Weight kg 8 0 Order number biovac 106 inclusive mai...

Страница 13: ...vice must be returned in packaging that is suitable for protecting it during transport 5 2 Installation and Connection Procedure Fig 5 Rear side connections biovac 106 with secretion glass container 2...

Страница 14: ...ing plug g Elbow plug with valve d Separator h Connecting coupling of the pump e Float valve b Coupling f PTFE Filter 1 m 2 Liters container i Diaphragm pump Inline filter 0 22 m 4 Liters container hy...

Страница 15: ...rve the operating instructions included when using the handvac 5 4 Storage The pumps are to be stored in a low dust interior room within the temperature range from 5 to 40 C and at a relative air humi...

Страница 16: ...only be performed by the manufacturer or a dealer authorized by him Beware of the pump parts being possibly contaminated by hazardous substances Wear protective clothing if there is contamination The...

Страница 17: ...r no 826801 4 Cross head screwdriver size 2 Order no 826801 2 Open spanner size SW 14 6 2 2 1 Disassembly 1 Disconnect the power supply and ensure that it cannot be switched on again 2 To open the cov...

Страница 18: ...ping ring screw fittings 9 Close the casing by shutting the cover 6 2 2 3 Test Connect the pump to the electrical supply Connect a vacuum measuring device to the suction connector and measure the ulti...

Страница 19: ...only be performed by the manufacturer or authorized workshops CAUTION The user shall be liable for the consequences of an incorrect damage report or a contaminated pump The statements in the damage r...

Страница 20: ...dentify and seal the leak replace the seals and or hoses if necessary Vacuum pump leaking Check the hose connections between the pump heads replace the hoses and or fittings if necessary Pump head lea...

Страница 21: ...ON We are not liable for any damage caused by the installation of any parts not supplied by the manufacturer 8 1 Service kit Diaphragm pump Designation Order no Service kit 402045 The service kit cons...

Страница 22: ...112573 Secretion glass container complete 2 Liters consisting of position 2 6 1 112523 1 2 Secretion glass container 2 Liters 1 828840 3 Cover complete for secretion glass container 1 828840 1 4 Sili...

Страница 23: ...P container complete 4 liters consisting of position 2 7 1 112523 7 2 Secretion PP container with cover 4 liters 1 828853 3 Float valve in the cover 1 112581 4 Silicone hose inside 6 mm x wall thickne...

Страница 24: ...Spare Parts Overview 24 112037 8 2 3 Spare parts view Basic unit complete Fig 10 Explosion view Basic unit complete...

Страница 25: ...PVDF 8 1 8 1 829919 829919 46 Quick release coupling PP NW 7 2 mm 6 4 mm 1 829996 2 829996 2 47 Angle with hose connector PP NW 7 2 mm 6 4 mm 1 829996 1 829996 1 48 Vacuum hose PTFE 8 6x1 mm 0 1 m 828...

Страница 26: ...ospheres part 1 Basic method and requirements Appareils non lectriques destin s tre utilis s en atmosph res explosibles partie 1 prescriptions et m thodologie DIN EN 13463 5 2011 10 Nicht elektrische...

Отзывы: