GANSHORN Medizin Electronic GmbH
Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012 Chapter 6, Troubleshooting, Checks & Adjustments, page 9 of 16
Checks and Adjustments
Service program COM98IO
Every LF8-installation performs automatically an installation of the COM98IO into the LF8-installation
directory (main directory). COM98IO allows checking separately the different channels of the
PowerCube.
To work with this service program,
open the file “com98io.ini” in the LF8-installation directory. Use the program “Windows notepad”
or “Editor”
Enter the correct COM-
port for the check: “Port=X”
“X “means the used PowerCube-COM port
Pict.: COM98IO.ini-window
Close the file and store the modifications.
Start the COM98IO program by clicking on “com98io.exe” in the LF8-installation directory and
select PowerCube as hardware
Open the drop-
down menu “Control” and click on “Start”
= The control measurement starts
= The window “com98io-info” gives actual numeric measurement values for every channel
= Use the measurement values for check or adjustment purpose
Pict.: CO98IO, Pull-down-
menu “Control”