GANSHORN Medizin Electronic GmbH
Service Manual Ref.No. 03 140 0 046
PowerCube-Ergo-CASE_ENG_rev.02, 23.11.2012
chapter 3, Disassembly, page 3of 10
_____________________________________________________________________
Physical Overview
The PowerCube housing consists of two metal shells.
The base assembly contains all electronic parts of the PowerCube.
The cover is connected to the base shell with a cable for potential equalization.
For removing some components it is necessary to remove other components first.
This is mentioned in the description of the respective procedure
General
To prevent devices from scratching, abrasion or other damage, always place it on a clean, soft
and antiskid ground for any maintenance procedures
When removing the cover, take care not to place any strain on the equalization cable. Ensure
that it is not bended or twisted. If necessary, unplug it for repair (Faston-plug). Take care to plug
it again and not to squeeze it when putting the cover on the base assembly
Proceed carefully when removing and replacing connectors. Never use force. Never strain cable
assemblies
Always perform the processing steps in the given order
After repair allow for warm-up time before adjustments and tests
This chapter shows how to remove and replace modules that are considered as spare parts.
The instructions show:
The requirements for removing / replacing the module
Necessary tools for removing / replacing and functional testing
Removal and replacement procedures
It is possible to remove /exchange the following modules:
Electronic unit complete (Analog Module and Controller Module)
Controller Module
Analog Module
Mounting plate
Electronic board holder
CO2-Analyzer
O2-Analyzer (SSO)
Suction pump
Calibration valve
Compensation vessel
Tools and Accessories for PowerCube Maintenance
Philips screwdriver PH1
Torx TX10
Fork wrench (metric size 5 and 5.5)
Side cutter
Flat bladed screwdriver 1.2 x 60
Cable tie
Additional equipment for adjustment procedures:
Manometer 0-10 mbar (accuracy: 0.5% of the maximal readings)
Flowmeter 0-1.5 l/min (accuracy: 5% of the maximal readings)
Multimeter (DC: range 1 mV to 30V, accuracy 0.15% of maximal readings)
COM98IO / PowerTool Software (automatically installed with LF8 installation)