Menu Guidance
16
Hydromette BL A plus
In the default setting (type 91) the measuring instrument displays a
too low measured value for both rough sawn timber as well as thin
material thickness of ≥ 10 mm to < 20 mm. The selection of type 92
compensates for this in the case of rough sawn timber. Type 93 is
selected for compensation for thin material thicknesses.
The menu can be reached as follows:
Main menu
Settings
Type
3.1.4
Temperature Compensation (resistance-based
measurement)
This menu is only available if a resistance-
based measurement method is selected
(types 1 to 7).
The input of the material temperature is
performed in steps of 1 °C. The temperature
compensation of the measured value is
always performed using the temperature
displayed in the measuring mode.
The menu can be reached as follows:
Main menu
Settings
Comp.temp
Figure 3-4:
Temperature
Compensation Menu