Weighting with tare: place the weighing box and press the [TARE] button. After standstill control the zero 
display appears. The weight of the container is now internally saved. Weigh the material, the net weight 
will be indicated. The weight of the weighing container will be displayed as a minus number after removing 
the weighing container. The tare weight is saved until it is deleted. Remove the load from the balance and 
press the [TARE] button. The tare procedure can be repeated as many times as necessary, for example 
with initial weighing of several components for a mix (add-on weighing). The limit is reached when the 
total weighing range capacity is full. 

9.3

 

LIGHT - [Display backlight] 

Press the [LIGHT] key to change the display’s backlight mode. 
[

b ON

] - Background illumination on. Display rich of contrast which can also be read in the darkness. 

[

b OFF

] - Backlight switched off to save battery. 

[

b AUTO

] - This display will appear backlit for weight value is not stable. When the weight value is stable, 

the display will be extinguished 3 seconds later. 

9.4

 

COUNT - [Pieces counting function] 

Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by 
putting on a certain number of the parts to be counted. The balance determines the total weight and 
divides it by the number of parts, the so-called reference quantity (1, 10, 20, 50, or 100). Counting is then 
carried out on the basis of the calculated average piece weight. 
As a rule: The higher the reference quantity the higher the counting exactness. 
Follow these steps: 
a)

 

Press [COUNT]-key to change the balance into counting mode. 

b)

 

Place as many pieces to add-up as required by the reference quantity. Press the [UNIT] button to enter 
the setting menu. The display now shows the number "1" and the unit "Pcs". 

c)

 

Using the [UNIT] button select the reference quantity (1, 10, 20, 50, or 100). 

d)

 

After successful reference determination the current quantity is displayed and the unit “Pcs”. 

e)

 

Remove reference weight. The balance is now in parts counting mode counting all units on the 
weighing plate. 

9.5

 

UNIT- [Weighing unit selection] 

Use the [UNIT]-key to choose weighing units: g (gram), ct (carat), ozt (troy ounce), oz (ounce), GN (gran), 
dwt (pennyweight) and lb (pound). 
 
 

 

       

 

 

  1 g   = 5 ct  

= 0.032150747 ozt 
= 0.035273962 oz 
= 15.43235835 GN 
= 0.643014931 dwt 
= 0.0022046226 lb 

9.6

 

CAL - [Calibration with external weight] 

The balance has an external calibration function.  
The adjustment procedures se

chap. "External calibration”

. 

Содержание E100YB-C

Страница 1: ...6 0 7 Version 2 0 04 2015 EN Precision balances EY E150Y E300Y E600Y E1200Y E3000Y E6000Y E150Y C E300Y C E600Y C E1200Y C E100YB C E200YB C E300YA E600YA E1200YA E3000YA E6000YA E2000YB E3000YB To e...

Страница 2: ...acking Installation 9 6 3 General power supply 10 6 3 1 Battery power supply 10 6 3 2 Mains connection during rechargeable battery operation 10 6 3 3 Battery exchange 10 6 4 Initial Commissioning 10 6...

Страница 3: ...12 1 Technical Data 17 12 2 RS232 Pinout 9 pin D Sub 18 12 3 Description of the data transfer 18 12 4 Remote control instructions 18 13 Maintenance and Servicing 19 13 1 Cleaning 19 13 2 Maintenance...

Страница 4: ...1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight l...

Страница 5: ...ming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight lb Pfund LCD...

Страница 6: ...ght 1 d 1 d 1 d 1 d 1 d 1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz...

Страница 7: ...or battery accumulator 2 1 Proper Use The balance you purchased is intended to determine the weighing value of material to be weighed It is intended to be used as a non automatic balance i e the mater...

Страница 8: ...afety Precautions COUTION Carefully read this operation manual before setup and commissioning Environmental conditions only use indoor max Altitude 2000m Operating temperature 20 C 20 C humidity of ai...

Страница 9: ...ou select the right location for your balance Therefore observe the following for the installation site Major display deviations incorrect weighing results may be experienced should electro magnetic f...

Страница 10: ...dapter Otherwise there is a risk that the Balance be impaired If the balance is not used for a long time remove the battery accumulator and store them separately Leaking battery liquid could damage th...

Страница 11: ...technical measuring Info about adjustment weights can be found on the Internet at www gandg de Operations sequence Observe stable environmental conditions A warming up time 30 min is required for stab...

Страница 12: ...Note The PVC wind protection is supplied for model E150Y E300Y E600Y E1200Y E300YA E600YA und E1200YA see chap 1 Technical data Note The Glass wind protection is supplied for model E150Y C E300Y C E6...

Страница 13: ...G G GmbH Operating elements 13 8 Operating elements Pic 1 E150Y E300Y E600Y E1200Y E300YA E600YA and E1200YA Pic 2 E3000Y E6000Y E3000YA E6000YA E2000YB and E3000YB...

Страница 14: ...lance before the first commissioning Level the balance with the adjustable feet horizontal until the air bubble in the level is in the center of the Level Indicator Give the balance after changing the...

Страница 15: ...parts it must know the average part weight i e reference Proceed by putting on a certain number of the parts to be counted The balance determines the total weight and divides it by the number of part...

Страница 16: ...re changes must be made only at special conditions of use Overview In the off state press the CAL button and keep it pressed press the ON OFF button turn on the balance Now the balance is accessed to...

Страница 17: ...enable Default setting is 1 h Save settings After all points have been configured press CAL button returns the balance automatically into weighing mode i Restore the factory settings In the off state...

Страница 18: ...if the value is positive a space is transmitted Data 7 bit Value for weight decimal point Unit 3 bit Weight unit CR 1 bit Carriage Return LF 1 bit Line Feed Example The output value 12 345 kg Output...

Страница 19: ...ual vacuum cleaner Remove spillages immediately 13 2 Maintenance and repair The appliance may only be opened by trained service technicians who are authorized by G G Before opening disconnect from pow...

Страница 20: ...not at zero Adjustment is no longer correct Great temperature fluctuations Warm up time was ignored Electromagnetic fields static charging choose different location switch off interfering device if p...

Отзывы: