Caring for your 403 LE

Phantom comes ready to scan. Do not remove surface material.

Store the 403 LE with cover closed securely.

Always attach the scanning surface cover and store the phantom out of direct sunlight
when it is not in use. Store in the plastic bag that is provided or in an air and water-tight
container.

Store the 403 LE at 35º –105ºF (2º–40ºC).

Freezing temperatures will damage the phantom and high temperatures will accelerate
desiccation.

Weigh the 403 LE to monitor desiccation.

Weigh the phantom when you first receive it and then every 6 months after that. Record
the values on the data sheet.

Do not drop or damage the phantom.

Return the phantom for inspection and/or repair if it has been dropped or damaged.
Physical damage to the case will cause premature desiccation. 

Gammex recommends annual servicing of your 403 LE to ensure proper operation. 

Our qualified service technicians will check for desiccation and provide any needed 

rejuvenation, scanning/certification to original specifications, and repairs.

Содержание 403 LE

Страница 1: ...403 LE Precision Multi Purpose Phantoms User s Guide...

Страница 2: ...e Phantom 7 Target Specifications 11 Cystic Targets 11 Pin Targets 11 Resolution Target Groups 11 Phantom Specifications 13 Physical Specifications 13 Tissue Mimicking Background Material 13 Low Scatt...

Страница 3: ...onic caliper measurement Pin targets have been included to allow for testing high frequency transducers that are designed to image the first few centimeters of tissue The phantom uses the latest techn...

Страница 4: ...temperatures will accelerate desiccation Weigh the 403 LE to monitor desiccation Weigh the phantom when you first receive it and then every 6 months after that Record the values on the data sheet Do n...

Страница 5: ...l or lanolin based gels as a coupling medium Poor transmission is a result of insufficient coupling Do not press the transducer into the scanning surface This damages the scanning surface and will sho...

Страница 6: ...l correspond very well to measured values seen in human tissue These qualities give the phantom its smooth appearance and allow the use of normal scanner control settings which mimic a realistic scann...

Страница 7: ...om You may notice that the tissue echoes near the bottom of the phantom fade into noise The depth at which usable echoes disappear is called the depth of penetration The depth markers on the phantom l...

Страница 8: ...in targets The vertical pins have 2 cm spacing while the horizontal pins have 3 cm spacing Use the pin targets to determine vertical distance accuracy and horizontal distance accuracy Note that the hi...

Страница 9: ...o vertical overlap is called the scanner s axial resolution Pin targets are resolved axially if an imaginary horizontal line can be drawn between the targets without touching either target The targets...

Страница 10: ...et groups as part of the cyst imaging test 9 Decrease image depth to the minimum and examine the dead zone target group The dead zone targets can also be used to measure lateral resolution in the extr...

Страница 11: ...cing 10 mm at 2 to 4 cm deep 20 mm at 4 to 16 cm deep Horizontal spacing 30 mm at 2 and 12 cm deep additional pin spaced at 10 mm in shallow set Resolution Target Group 3 8 and 14 cm deep All acoustic...

Страница 12: ...12...

Страница 13: ...eed of sound 1540 10 m s Temperature dependence on speed of sound 1 5 m s C Attenuation coefficient 0 7 0 05 dB cm MHz 0 5 0 05 dB cm MHz refer to phantom side label Nonlinearity parameter B A 6 6 0 3...

Страница 14: ...attenuation as there would be if the higher frequency were used to generate the pulses at the transducer So higher frequency resolution occurs for greater depths within the subject than if conven tion...

Страница 15: ...controlled environments the phantom weight should be checked every six months Phantoms used in high temperature low humidity environments or in mobile situations should be tested more frequently As th...

Страница 16: ...y Seller upon Seller s ex amination Buyer shall pay all transportation charges SELLER SHALL NOT BE OTHERWISE LIABLE FOR ANY DAMAGES INCLUDING BUT NOT LIMITED TO INCIDENTAL DAMAGES CONSEQUENTIAL DAMAGE...

Страница 17: ...ours are Monday through Friday 7 30 am to 5 00 pm Central Time 1 800 GAMMEX 1 426 6391 1 608 828 7000 1 608 828 7500 Fax e mail sales gammex com Service Department hours are Monday through Friday 7 30...

Страница 18: ...18 Notes...

Страница 19: ...19 Notes...

Страница 20: ...0 www gammex com e mail sales gammex com Gammex RMI LTD Broadway Business Centre 32A Stoney Street Nottingham NG1 1LL United Kingdom 44 0115 9247188 Fax 44 0115 9247189 e mail uksales gammex com Gamme...

Отзывы: