MPCe/LPCe Users Manual
35
Hex
Command
Description
Data Field
Response
Data/Response Interpretation
61
IS HIGH VOLTAGE ON
Indicates if the high voltage for
supply 1 or 2 is enabled
“1” or “2”
YES or NO
1 or 2 indicates the supply
62
SET SERIAL ADDRESS
Sets the controllers serial address
-
nnn
nnn = new serial address (000-255)
63
IHIGH OFFSET
Sets the offset value for high
current metering
s, nnn
-
s= configured supply (1 or 2)
nnn = -999 to 999 mA
68
HVE RESTART
Specifies the mode of the interlock
input pins to either restart or
remain off after an open circuit is
detected.
c
-
c = “y” for yes or “n” for no
Yes indicates that shorted pins will
restart the supply after an open
circuit is detected.
No indicates that the shorted pins will
not restart the supply after an open
circuit is detected.
69
GET HVE RESTART
Displays the mode of the interlock
input pins to either restart or
remain off after an open circuit is
detected.
-
YES or NO
Yes indicates that shorted pins will
restart the supply after an open
circuit is detected.
No indicates that the shorted pins will
not restart the supply after an open
circuit is detected.
CB
READ EVENT LOG
Reads event data from the event
log
0-50
String
0 = latest event
1-50 = specific event
String = pump#, message type,
time/date, voltage, current, pump
size, supply size, # of arcs, # cool
down cycles, 0 0 0 0 (reserved)
CC
READ LAST EVENT
Reads event data from the last
entry in the event log
-
String
String = pump#, error code,
time/date, voltage, current, pump
size, supply size, # of arcs, # of cool
down cycles, 0 0 0 0
CD
CLEAR EVENT LOG
Removes all entries from the event
log
-
-
-
D3
SET COMM MODE
Sets the communication mode to
local, full, or remote
N
-
N is the mode
0 = Local
1 = Remote
2 = Full
D4
GET COMM MODE
Returns the current
communication mode.
-
“N”
N is the mode
0 = Local
1 = Remote
2 = Full
RS232/422/485 DIGITEL MPCe/LPCe Supply Specific Commands
Hex
Command
Description
Data Field
Response
Data/Response Interpretation
0A
READ CURRENT,
Reads pump current from supply
one or two
“1” or “2”
X.XE-X AMPS
1 or 2 indicates the supply.
0.1E-9 = HV OFF STATUS
0B
READ PRESSURE,
Reads pump pressure from supply
one or two
“1” or “2”
X.XE-XX UUU
1 or 2 indicates the supply
UUU is pressure units (Torr, MBR, or
PA)
0.1E-10 = HV OFF STATUS
0C
READ VOLTAGE,
Reads pump voltage from supply
one or two
“1” or “2”
XXXX
1 or 2 indicates the supply
In volts