Non-Alarm Troubleshooting
This section describes probable causes and corresponding action steps for non-alarm problems.
Agitate/Wash-In Operation Not Working (AGITATE WASH IN Light Illuminated)
The centrifuge does not agitate and/or a valve does not open.
Probable Cause
Action
No valve or multiple valves were programmed to open during an
Agitate/Wash-In operation, placing the COBE 2991 a Hold status.
Reprogram the COBE 2991 as needed:
•
For the LED program board, see “VALVE SELECT” on page 4-5.
•
For the pin-diode program board, see “VALVE” on page 5-4.
Circuit breaker CB3 has been tripped.
1
Reset circuit breaker CB3.
See “CIRCUIT BREAKERS” on page 3-10.
2
If the condition recurs, contact a qualified service
representative.
The CENTRIFUGE SPEED was set to zero during the Start/Spin or
Supernatant-Out operations. If the centrifuge is not spinning during
the Supernatant-Out operation, the Agitate/Wash-In circuitry will not
be properly activated.
1
Check the CENTRIFUGE SPEED setting.
2
If the CENTRIFUGE SPEED was set to zero, press STOP RESET to
reset the COBE 2991.
Internal malfunction.
Contact a qualified service representative.
Caution:
Before attempting to reset a circuit breaker, always unplug the
Warning:
Содержание COBE 2991
Страница 1: ......
Страница 18: ...This page intentionally left blank ...
Страница 19: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 25: ......
Страница 26: ...Hướng dẫn sử dụng Part No 770273 010 Reorder No 709016 000 2005 05 Máy xử lý tế bào ...
Страница 27: ...COBE 2991 Cell Processor Essentials Guide 1 1 Introduction 1 1 Giới thiệu ...
Страница 37: ...Operator Controls 3 COBE 2991 Cell Processor Essentials Guide 3 1 Các kiểm soát của người vận hành ...
Страница 48: ...LED Programming 4 COBE 2991 Cell Processor Essentials Guide 4 1 4 Lên chương trình LED ...
Страница 59: ...Pin Diode Programming 5 COBE 2991 Cell Processor Essentials Guide 5 1 5 Lên chương trình Pin Diode ...
Страница 105: ...Troubleshooting 7 COBE 2991 Cell Processor Essentials Guide 7 1 7 Xử lý sự cố ...
Страница 116: ...General Maintenance 8 COBE 2991 Cell Processor Essentials Guide 8 1 8 Bảo dưỡng ...
Страница 120: ...Essentials Guide 1 6 17 05 10 29 08 AM ...
Страница 121: ...Essentials Guide Part No 770273 010 Reorder No 709016 000 2005 05 Cell Processor ...