70
Figure 65: Herriott Cell Secured to L-bracket Mount
5.
Make note of the orientation of the Herriott cell cover in relation to the rest of the Herriott
cell. This is important to ensure proper reassembly after maintenance is complete.
6.
Remove all eight screws holding the Herriott cell cover in place. Figure 74 shows one of the
eight screws which has been partially removed.
Figure 66: Screws Holding Herriott Cell Cover in Place
7.
Gently slide the cover off and set it aside. The Herriott cell mirrors should now be exposed.
See Figure 75. Use extra caution from this point forward to avoid inadvertently scratching
the mirrors.