1.4.2.1 Direct voltage
(automatic range selection)
range
resolution
accuracy
400mV
0,1mV
+/-(0,8% rdg + 5dgts)
4V
1mV
+/-(0,8% rdg + 3dgts)
40V
10mV
400V
100mV
1000V
1V
+/-(1,0% rdg + 5dgts)
Input impedance:
10MΩ
Overload protection: 1000V- or 750~
200mV range: 250V -/~
Max. input voltage:
1000V-
1.4.2.2 Alternating voltage
(automatic range selection)
range
resolution
accuracy
4V
1mV
+/-(1,0% rdg + 8dgts)
40V
10mV
400V
100mV
750V
1V
+/-(1,2% rdg + 8dgts)
impedance:
10MΩ
Frequency range:
40Hz – 400 Hz
Overload protection: 1000V- oder 750~ rms
Response:
True-RMS
Max. input voltage:
750V~
1.4.2.3
Temperature
range
resolution
accuracy
-40 to 1370 °C
1°C
-40 to 150 °C ±(2,5% +
4)about
150°C to 1370°C ±(2,5% + 4)
-40°F to 2000°F
1°F
-40 to 302 °F ±(2,5% +
4)about
302 to 2000°F ±(2,5% + 4)
NOTE:
A K-type thermocouple is included in the delivery. If
other types are used, the measurement results may
differ.