5. Practical considerations  

 5.1 Handling and maintenance 

Physilog®5 should be handled carefully. In particular, it should not receive shocks, such as 
fall, crushing, being hit, etc. Do not press forcefully on the on/off button, this can damage 
the button. 

 

 Warning: 

Do not use or charge the Physilog® if the sensor or its case is 

damaged. Contact Gait Up about what further action is needed when the case 
is damaged. In case the battery is damaged immediately disconnect the 
sensor from any plug and move the device away from flammable materials 
and contact Gait Up

’s customer support. 

 

 

To benefit from the maximal battery life, fully charge Physilog

®

 sensors before 

doing measurements (LED shines green when connected to a computer or 
charger, you can also check the battery level with the Physilog

®

5 companion 

Android Application available on www.gaitup.com/support). 

 

 

Copy trial recordings to your computer: Physilog

®

 is not the best place to keep and 

save important data. GaitUp cannot be held responsible if data stored on the 
sensor are lost. 

 

 

Physilog

®

5 can be used outside. Do not use Physilog

®

5 to measure water-sport 

activities without additional waterproofing. 

 

 

Storage: Store in a cool (0 °C

—30 °C) and dry place. The provided box is the 

perfect place for your sensor. 

 

 

Cleaning: Sensors can be cleaned gently using alcohol wipes. Before you clean the 
device makes sure it is turned off and disconnect all cables. Alcohol used in excess 
may damage the sensor’s waterproofing. Make sure to use wipes and not put liquid 
alcohol directly on the sensor. Do not clean the USB entry with alcohol. 

 

 

Each sensor is individually calibrated by Gait Up and they do not need yearly 
recalibration for gait measurements. 

 

 

Do not throw Physilog

®

 sensors in the normal trash, dispose of it properly 

considering local laws and rules about electronic waste and batteries. 

 

 

Make sure to know which Physilog

®

 file corresponds to which trial for your 

analysis

—First digits of file name are automatically increased (exceeding 999 and 

new date resets the increment back to 1). Note that date of the file generation 
(recorded between one start and stop of the Physilog

®

) is visible in the XLS or CSV 

file and the Matlab structure “header”. If you wish to adapt internal date and time of 
the Physilog

®

, read the corresponding section of the user manual of Physilog

®

5. 

 

 

Do not remove the “conf.bin” file from the Physilog, otherwise it will  return to a 
default configuration based on original configuration. 

 

 

Do not introduce pointy objects into the small hole besides the USB entry, this 
irreversibly damages the waterproofing of the sensor. 

 

 

 

Содержание PhysiRun Lab

Страница 1: ...EPFL Innovation Park Bat C CH 1015 Lausanne www gaitup com TVA CHE 145 136 760 User guide PhysiRun Lab v1 1 1 Date 06 05 2019 Copyright 2018 2019 All Rights Reserved Gait Up SA CH 1015 Switzerland 1...

Страница 2: ...iRun Lab for running analysis The software allows to highlight a multitude of biomechanical parameters to observe the running technique with the GaitUp sensors Physilog 5 The Physilog 5 started from a...

Страница 3: ...guration 4 2 3 Time setting 5 2 4 Fix the Sensors 5 2 5 Protocol 6 3 How to do a measurement with Physilog 5 6 3 1 Start measure from the watch 6 3 2 Start a measure from the Physilog 5 7 4 PhysiRun L...

Страница 4: ...rement or by 3 rapid red blinks when you want to start a new measurement If the Physilog 5 is completely discharged charge at least 30 minutes before using Battery life estimation is available when yo...

Страница 5: ...ctly set your files are saved in a folder with the measurement date If you start the sensors from the watch or the application the time is automatically updated and you thus don t need to follow the i...

Страница 6: ...s BIN are saved in the sensor memory 8Go A file contains the raw data Below you can check the name description 3 1 Start measure from the watch A measurement from the watch allows to start and stop th...

Страница 7: ...e started directly from the Physilog 5 Material required 2 Physilog 5 To start the measure hold the button for 3 seconds on each sensor Check that there is a green blinking on your two sensors after r...

Страница 8: ...t on your computer Save your file in the same folder on your computer 1LF091 BIN for example If the time is correctly configured you file is saved in a folder at the date of measurement for example 18...

Страница 9: ...different result files you want to obtain pdf excel file csv file In the PDF format there are three pages Excel and csv format give the cycle by cycle calculation for each parameter To generate these...

Страница 10: ...llowing instructions carefully For sensors calibration move the gray window to have a part of static period and some running cycles see picture below To move the window click on the top bar You can al...

Страница 11: ...s no space or special character allowed PDF excel and csv will have the same name The language for the PDF report can be selected from available translations in the list To obtain the third page of th...

Страница 12: ...5 to measure water sport activities without additional waterproofing Storage Store in a cool 0 C 30 C and dry place The provided box is the perfect place for your sensor Cleaning Sensors can be cleane...

Страница 13: ...log when the temperature is outside of the range of 0 to 45 C Physilog 5 should be charged through USB connection with a computer An external charger may be used note that no charger is provided by Ga...

Страница 14: ...ks Error 2 3 rapid red blinks Error 3 4 rapid red blinks Error 4 5 rapid red blinks Error 5 6 rapid red blinks Error 6 Error type Error type How to solve the issue Error 1 Disconnect and replug the Ph...

Страница 15: ...apply to any product or component thereof which has been repaired or altered by anyone other than Gait Up in any manner so as in Gait Up s judgement to affect its service ability or any product been s...

Страница 16: ...ice may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operations a 5 When the device is so small or for such us...

Страница 17: ...t us at EPFL Innov Park C CH 1015 Lausanne tel 41 21 633 7527 mail contact gaitup com web www gaitup com Document Version Changes Responsible Date 1 0 0 First public version Jules Gellaerts 29 11 18 1...

Отзывы: