Pitch Diameter Gage Operation Manual
Models MRP 1500/2500/3500
Copyright © 2018 Gagemaker. All rights reserved
44
Zeroing the MRP Gage for Internal Measurements Using the MIC TRAC
Materials Needed:
MRP gage
MIC TRAC MIC TRAC, CPU, and taper blocks
5/32” hex wrench (supplied with gage)
Cloth
To ensure consistent and accurate readings, the MRP gage should be zeroed on the MIC TRAC
once during each shift, at a minimum.
1. Turn on the MIC TRAC CPU.
2. Clean both of the receiver pads and the
mounting surfaces of the taper blocks using a
cloth.
3. Place the taper block against the receiver pad
shoulder with the raised side of the block
facing outward.
4. While holding the block against the receiver
pad shoulder, insert the two cap screws with
washers into the holes on either side of the
block.
5. While applying pressure toward the receiver
pad shoulder, use a 5/32" hex wrench to
tighten the screws.
6. Repeat this process with other taper block.
!
Raised Side of
Block