Pitch Diameter Gage Operation Manual
Models MRP 1500/2500/3500
Copyright © 2018 Gagemaker. All rights reserved
10
Setting Up the MRP Gage for External Measurements (continued)
3. Remove the upper block and rotate it 180
degrees.
4. Align the set screws in the upper block with the
holes located at the left end of the MRP frame.
5. Tighten the set screws to secure the upper
block.
6. Using a 3/16” hex wrench, loosen the set
screws in the lower block and remove the lower
wear pad.
7. Rotate the lower block and wear pad 180
degrees and re-attach the wear pad to the
lower block.
8. Partially tighten the set screws in the lower
block to secure the wear pad, but to allow the
lower block to slide back and forth.
Note:
The wear pad is not interchangeable
between MRP 2500 bodies. Each wear pad
is specifically manufactured to the gage
body with which it is supplied.