CompuScope 1220

 

 
Absolute Max. Amplitude: 
 

 

1 M

 impedance:  ±15 Volts (continuous) 

 50 

 impedance:  ±5 Volts (continuous),  

±15 Volts (for 1ms duration) 

DC Accuracy 

Input Range 

Accuracy 

± 10 V 

± 0.5% 

± 5 V 

± 0.5% 

± 2 V 

± 0.5% 

± 1 V 

± 0.5% 

± 500 mV 

± 0.5% 

± 200 mV 

± 0.5% 

± 100 mV 

± 1% 

± 50 mV 

± 2% 

Sampling Rates   

20 MS/s, 10 MS/s, 5 MS/s, 2 MS/s, 1 MS/s, 500 kS/s, 200 kS/s,  
100 kS/s, 50 kS/s,20 kS/s, 10 kS/s, 5 kS/s, 2 kS/s, 1 kS/s 

Protection 
 

 

1 M

 impedance:  Diode Clamped 

  50 

 impedance:  No protection 

DYNAMIC PARAMETERS 

Measured using 5 MHz sine wave input at 20 MS/s with amplitude of 95% of full scale on the ± 1 V range.  
Typical values listed below. 

SNR:  

55 

dB 

SFDR:  

57 

dB 

ENOB:  

8.9 bits 

ACQUISITION MEMORY 

Data Storage: 

 

In on-board memory 

Memory sizes: 

 

1 Msamples, 
4 Msamples, 
8 Msamples, 
64 Msamples, 
256 Msamples 
512 Msamples, 
1 Gsamples 

Maximum memory depth:  Up to half on-board memory per channel 

TRIGGERING 

Number of Trigger Inputs:  2 per card 

Trigger Source:   

CH A, CH B, EXT or Software  

Input combination: 

Wired-OR 

Type:  

 

Analog 

triggering 

Sensitivity: 

 

± 20% of full scale  

Level Accuracy:   

± 10% of full scale 

Slope:   

 

Positive or Negative, software selectable 

Post Trigger Data: 

64 points minimum in single record acquisition 
128 points minimum in multiple record acquisition 
Can be defined with a 64 point resolution 

Содержание CompuScope 1220

Страница 1: ...GAGE APPLIED TECHNOLOGIES CS1220 Hardware Manual Reorder MKT HWM PCI01 CS1220 0408...

Страница 2: ...The installation program used to install the GageScope Software InstallShield is licensed software provided by InstallShield Software Corp 900 National Parkway Ste 125 Schaumburg IL InstallShield is C...

Страница 3: ...20 4 CompuScope 1220 compliance statement 7 CompuScope 1220 product introduction 8 CompuScope 1220 specifications 9 CompuScope 1220 ordering information 13 CompuScope 1220 simplified block diagram 14...

Страница 4: ...ompuScope 1220 card If you purchased a CompuScope 1220 4M or CompuScope 1220 8M card you should see a piggyback memory board attached to the back of the card as shown below If you purchased a CompuSco...

Страница 5: ...e Note that some packages will only be available if you have purchased the software and have a key provided by Gage CompuScope Certificate of NIST Traceable Calibration Each CompuScope card is shipped...

Страница 6: ...er Slave Timing Module is used for 4 Card Master Slave system of CS1220 1M 2 Card Master Slave system of CS1220 4M or CS1220 8M 8 Slot Master Slave Timing Module 8 Slot Master Slave Timing Module is u...

Страница 7: ...l Fast Transient Burst Immunity Performance criterion B IEC61000 4 5 Power Line Surge Immunity Performance criterion B IEC61000 4 6 Conducted RF Immunity Performance criterion A IEC61000 4 11 Voltage...

Страница 8: ...es from CompuScope memory to PC memory run as high as 50 MB s for the CS1220 as compared to a few hundred KB s for GPIB CompuScope cards are easier to program as Software Development Kits are availabl...

Страница 9: ...ts Acquisition Memory Worst Case Typical 1M 25 0 W 17 5 W 4M 28 0 W 20 5 W 8M 28 0 W 20 5 W 64M 32 5 W 23 5 W 256M 32 5 W 23 5 W 512M 32 5 W 23 5 W 1G 32 5 W 23 5 W Note Power connector on the deep me...

Страница 10: ...of 95 of full scale on the 1 V range Typical values listed below SNR 55 dB SFDR 57 dB ENOB 8 9 bits ACQUISITION MEMORY Data Storage In on board memory Memory sizes 1 Msamples 4 Msamples 8 Msamples 64...

Страница 11: ...a 64 point resolution Maximum Number of Triggers 4 194 304 MULTI CARD SYSTEMS Operating modes Master Slave or Multiple Independent Maximum Number of Cards Master Slave 2 to 8 cards for 1M memory 2 3 o...

Страница 12: ...nced analysis tools except for Extended Math Professional Edition Provides full functionality of all advanced analysis tools SOFTWARE DEVELOPMENT KITS SDK CompuScope SDK for C C For Windows 98 ME NT 2...

Страница 13: ...1 003 GageScope Software Product Order No GageScope Lite Edition Included GageScope Standard Edition purchased with CompuScope hardware 300 100 351 GageScope Standard Edition purchased independently 3...

Страница 14: ...14 CompuScope 1220 CompuScope 1220 simplified block diagram Figure 1 CS1220 simplified block diagram...

Страница 15: ...ad of the CompuScope card s own oscillator External Trigger BNC Connector used to input an analog or digital signal which may be used as an External Trigger External Trigger is defined exactly as in a...

Страница 16: ...note that much higher PRFs will be achieved using CompuScope Multiple Record mode The following test results were obtained using a computer configured as follows Pentium III 1 GHz processor 512 MB RAM...

Страница 17: ...CompuScope 1220 17 Notes...

Отзывы: