Standards 

Electromagnetic 
Compatibility 

IEC 61000-6-2/6-3 

Resistance to Electrical 
Storms 

IEC 61000-6-2; (IEC 61000-4-2 Basic Standard for applicability of tests) 
 

IP Rating  

IEC 60529: IP 54 

Impact Resistance 

5 drops from 1 meter onto concrete floor at room temperature with 
battery in place. Device does not get damaged and there is no loss of 
calibration. 

Vibration 

EN12830:1999 Clause 4.9.3.2 and Test Method 5.6.6 

RoHS 

Compliant (EU directive 2011/65/EU) 

Verification 

In accordance with PQS verification protocol E006/TR05.VP.1 

 

*: The current settings are set according to the requirements of WHO/PQS/E006/TR05.1. Other settings are available 
upon request. Specifications are subject to change without notice.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание LM Pro T

Страница 1: ...on Pvt Ltd 3 mahavir estate karelibaug vadodara 390 018 tel 91 265 2461912 email info gtek india com url www gtek india com OPERATING MANUAL LM Pro Pro LM T User Programmable Temperature Data Logger M...

Страница 2: ...of Key 10 5 USING THE PRODUCT 11 5 1 Configuring the Device 11 5 1 1 Alarm Configuration 11 5 1 2 Batch Configuration 12 5 2 Configure Multiple Devices 13 5 3 Batch Start of the Device 14 5 4 View Mi...

Страница 3: ...ch on Multiple devices 13 Figure 7 Start batch manually 14 Figure 8 View Min Max and current Temperature 14 Figure 9 Insert Manual Tag 14 Figure 10 Manual Stop of Running Batch 17 Figure 11 USB Type A...

Страница 4: ...ersonnel also Do not use the product if there are signs of damage to the housing There are no user serviceable parts inside For any defect please consult the factory or the dealer from where you bough...

Страница 5: ...51 software needs to be installed in PC To download data to generate a report in pdf format and to export data in csv format for further use 2 2 Technical Data Table 1 Technical Specifications Model L...

Страница 6: ...nd Current reading with measurement unit Memory Size 50000 data storage Activation Device activation is by software at preset start delay time If user requires instant activation of the device before...

Страница 7: ...t room temperature with battery in place Device does not get damaged and there is no loss of calibration Vibration EN12830 1999 Clause 4 9 3 2 and Test Method 5 6 6 RoHS Compliant EU directive 2011 65...

Страница 8: ...opened immediately and the device be examined If the device is found damaged it should not be operated and the local representative contacted for instructions Ensure that all accessories and document...

Страница 9: ...anical Dimensions of LMPro T Data Logger Figure 2 Overall Dimensions of LMPro T Data Logger Overall Dimensions Dimension L x W x H mm 95 5 x 45 x 14 8 mm approx Mounting Through hole provided Weight A...

Страница 10: ...tch running condition alarm has been triggered the symbol X NOK will be turned on and will remain even if the alarm is restored b If any time during batch running condition alarm has not triggered the...

Страница 11: ...ical reasons the display intensity of liquid crystal display becomes lower at temperatures below 0 C This has no influence on the measuring accuracy For technical Reasons the battery performance decre...

Страница 12: ...figuration are as follows 1 Alarm No Alarm 1 A1 to Alarm 4 A4 2 Alarm Type Set the alarm type High Low as required 3 Set point Enter the appropriate Temperature value in C as per required alarm condit...

Страница 13: ...ch name of maximum 10 alphanumeric characters 3 Unit of Temperature The unit for temperature is C 4 Store Interval It is the interval between two successive logging of data It can be set between 5 sec...

Страница 14: ...e Batch button to configure the batch for the device 5 2 Configure Multiple Devices If multiple devices are to be configured with the same parameters user should select the multiple batch configuratio...

Страница 15: ...let the batch start at pre configured time only 5 4 View Min Max and Current Temperature Data As described in key functions section 4 3 by pressing Min Max key for 1 sec in normal running condition th...

Страница 16: ...and unit will be seen on display 2 Reading is outside the cumulative alarm setpoint high low for time duration less than alarm high low delay OK sign battery reading with UP Down arrow and unit will...

Страница 17: ...riggered tag event is logged as per alarm Tag Id Alarm restored tag event is logged as per alarm number when Temperature reading comes within the alarm high low limits User entered Tag is logged as Ma...

Страница 18: ...the LMViewE051 application using Stop option 4 User can manually Stop the running batch if required by pressing Min Max key for about 10 seconds as shown in figure 10 Here a Manual Tag event will be...

Страница 19: ...The current temperature reading Battery level and UTC time of the device will be seen as shown in figure 12 Figure 12 Current Temperature reading of the LMPro T Data Logger Select the download option...

Страница 20: ...e trend of measured data 4 Generate csv file of the measured readings 5 Generate PDF report User can download multiple device data in the same file if the configuration is same for the given devices D...

Страница 21: ...s up to 10 in one file to sustain readability of the data in the graph and PDF report 5 8 3 Generating PDF Report To generate the PDF report of data click on icon A pop up window will appear with the...

Страница 22: ...ally select the Generate Report button to save the generated PDF report at appropriate file location as shown in figure 18 Figure 18 Selecting file location for saving PDF report Note Refer to the hel...

Страница 23: ...hen USB port is not in use cover the USB port properly 6 3 Changing the Battery The LMPro T data logger contains a Lithium Battery The end of the battery life is indicated by a low battery symbol the...

Страница 24: ...hen new battery is inserted the min max value for the current data will be calculated from that time onwards and batch recording is resumed Note It is advisable to reconfigure the data logger after ch...

Страница 25: ...ct in USB and LMViewE051 software USB port might be faulty try with other port of PC For how much time the display remains ON after device activation Display turns off after 1 minute of activity if LC...