Reading is outside both the
Single and cumulative
alarm setpoint high/ low for time
duration greater than alarm high/low delay.
-
Alarm sign, bell sign, battery, reading with Up/ Down arrow, single event indication
and unit will be seen on display.
5.7 Measurement of the data
5.7.1 Start Recording data
➢
The Data recording can be started by one of the criteria explained in section 5.3.
➢
The device will store the data as per configurated store interval, which can be
downloaded later on for analysis.
5.7.2 Tag Events
➢
When alarm high/low trigger condition arises, alarm triggered tag event is logged as per
alarm Tag Id.
➢
Alarm restored tag event is logged as per alarm number when Temperature reading
comes within the alarm high/low limits.
➢
User entered Tag is logged as Manual Tag Event
“
Tag
”
with a time stamp.
Tag Id
Tag Event
Description
A1
Alarm1 Triggered
Alarm1 high/low condition has occurred.
B1
Alarm1 Reset
Alarm1 high/low condition is restored.
A2
Alarm2 Triggered
Alarm2 high/low condition has occurred.
B2
Alarm2 Reset
Alarm2 high/low condition is restored.
A3
Alarm3 Triggered
Alarm3 high/low condition has occurred.
B3
Alarm3 Reset
Alarm3 high/low condition is restored.
A4
Alarm4 Triggered
Alarm4 high/low condition has occurred.
B4
Alarm4 Reset
Alarm4 high/low condition is restored.
Tag
Manual Tag
User has entered a tag from device.