- 11 -
Do not remove the test lead immediately. Wait until the discharge
is complete before touching the circuit under test.
In order to ensure the measurement accuracy, do not twist the
test lines together.
Guaranteed temperature and humidity of the insulation resistance accuracy
Insulation
resistance
range
Guaranteed the humidity
value of the insulation
resistance accuracy
Guaranteed temperature of
the insulation resistance
accuracy
0
Ω
-100M
Ω
<85% RH(No condensation)
23
℃±
5
℃
101M
Ω
-20G
Ω
<75% RH(No condensation)
21G
Ω
-200G
Ω
<65% RH(No condensation)
Insulation resistance test can only be carried out on an uncharged
circuit. Before testing, check that the test leads are good and confirm that
the circuit under test is uncharged.
Turn the rotary switch to
gear, then press
VSEL
to select the voltage
value to be tested.
One end of the ground wire (black) is connected to the instrument
connection
EARTH
and the other end is connected to the ground end of the
circuit under test. One end of the high-voltage rod test line (red) is
connected to the other end of the instrument
LINE
and the head is in contact
with the circuit under test.If the test has a green branch line as a shielded
line, the accuracy of connecting the GUARD port during testing is better.
If the matching test line does not carry this line, it does not need to be
connected. As shown in the figure, press the
TFRT
test button. The LCD shows
Содержание ES3035