VI.Measuring Principle

 

Insulation  resistance  measurement  uses  a  voltage  generator  to 

generate a voltage, 

V

, applied across the resistor, measuring the current 

I

 flowing across the resistor, and calculating the ground resistance 

value 

R

 according to the formula 

R=V/I.

 

 

 
 
 
 
 

 
 
 

VII.Operation Method 

1

Power On/Off 

Press the POWER button to turn the switch on and off. After the power 

is turned on, 

APO

 is displayed in the lower corner. When it is not in 

operation, it will automatically shut down after 15 minutes. 

2

Battery Voltage Check 

After powering on, if the LCD shows a low battery voltage sign 

   

it  indicates  that  the  battery  is  running  low.  Please  charge  in  time. 

Sufficient battery power to ensure measurement accuracy 

3

 DC Voltage Test 

 

Input instrument DC voltage cannot exceed 1000V 

Press the button 

 

to switch to the DC voltage test mode, connect the 

green test lead to the 

V

 terminal, the black test lead to the 

COM

, and the 

LCD displays the real-time DC voltage value. 

 

 

 
 
 
 
 
 
 

Содержание ES3035

Страница 1: ......

Страница 2: ...1...

Страница 3: ...I Operation Method 9 1 Power On Off 9 2 Battery Voltage Check 9 3 DC Voltage Test 9 4 AC Voltage Test 10 5 Insulation Resistance Test 10 6 GUARD Use of Protective Wires 12 7 Polarization Index PI and...

Страница 4: ...e on the tester u Please don t measure in an inflammable place The flame sparkle maybe cause explosion u During usage of tester please stop using it when exposed metal is caused by broken enclosure or...

Страница 5: ...e absorption ratio and polarization index of DC voltage The machine is beautiful and upscale has a wide range high resolution convenient operation easy to carry accurate reliable stable performance st...

Страница 6: ...0 1M 100 1000M 1M 1G 10G 5 rdg 5dgt 0 01G 500V 10 0 10M 3 rdg 5dgt 0 01M 10 100M 0 1M 100 1000M 1M 1G 10G 5 rdg 5dgt 0 01G 1000V 10 0 20M 3 rdg 5dgt 0 01M 20 200M 0 1M 200 2000M 1M 2G 20G 5 rdg 5dgt...

Страница 7: ...itions 23 C 5 C below 75 rh Rated voltage V 100 250 500 1000 2500V 5000V Test Voltage V rated voltage 1 10 Insulation Resistance Range G 0 01M 1000G Insulation Resistance Insulation 0 01M DC Voltage R...

Страница 8: ...Data Storage 500 groups FULL symbol indicates that storage is full Data Review Data review function MR symbol display Overflow Display Exceed measurement range overflow function OL symbol display Alar...

Страница 9: ...lay screen 2 LINE interface 3 V interface 4 GUARD interface 5 EARTH interface 6 Function button area 7 Power on off button 8 Test button 9 changer interface 10 USB interface 11 Security Alligator Clip...

Страница 10: ...ed on APO is displayed in the lower corner When it is not in operation it will automatically shut down after 15 minutes 2 Battery Voltage Check After powering on if the LCD shows a low battery voltage...

Страница 11: ...test circuit wiring is in good condition and whether the circuit under test is energized If the circuit is live it may damage the instrument and affect the measurement accuracy Must wear high voltage...

Страница 12: ...charged circuit Before testing check that the test leads are good and confirm that the circuit under test is uncharged Turn the rotary switch to gear then press VSEL to select the voltage value to be...

Страница 13: ...resistance value In order to avoid this phenomenon as shown in the figure below use the protection wire any conductive bare wire to flow the leakage current through the part After connecting to the pr...

Страница 14: ...on ratio see the following formula for the specific test time but for the large capacity and long term absorption process such as transformers generators cables capacitors and other electrical equipme...

Страница 15: ...s pressed The LCD displays the measured value After the measured value is fixed the absorbance or polarization index can be read After the test is completed press to view the divisor of the absorption...

Страница 16: ...very time you turn it on 9 Alarm Settings After power on long press to turn on and off the alarm function Long press SET key to set the resistance alarm value Press or key to change the current digit...

Страница 17: ...as the third set of data 11 Data Review Deletion After booting or measuring press the MEM button more than 3 seconds to enter the data lookup and store the data read interface MR symbol display Press...

Страница 18: ...figure is displayed VIII Battery Description The instrument is powered by a 12V battery when the battery power is reduced when the voltage drops to 10V the battery symbol is displayed Please replace t...

Страница 19: ...18 Instrument box 1PC...

Страница 20: ...ZhengNeng Electronics Technology Co Address 2F No 15 Baoshu Road Taihe Baiyun District Guangzhou Guangdong China Toll free call 4000 1515 38 Tel 86 20 36544172 Fax 86 20 37319075 Post 510540 WebSite...

Страница 21: ...20...

Отзывы: