VI.Measuring Principle
Insulation resistance measurement uses a voltage generator to
generate a voltage,
V
, applied across the resistor, measuring the current
I
flowing across the resistor, and calculating the ground resistance
value
R
according to the formula
R=V/I.
VII.Operation Method
1
.
Power On/Off
Press the POWER button to turn the switch on and off. After the power
is turned on,
“
APO
”
is displayed in the lower corner. When it is not in
operation, it will automatically shut down after 15 minutes.
2
.
Battery Voltage Check
After powering on, if the LCD shows a low battery voltage sign
“
”
,
it indicates that the battery is running low. Please charge in time.
Sufficient battery power to ensure measurement accuracy
3
.
DC Voltage Test
Input instrument DC voltage cannot exceed 1000V
Press the button
to switch to the DC voltage test mode, connect the
green test lead to the
V
terminal, the black test lead to the
COM
, and the
LCD displays the real-time DC voltage value.