IAC200 Quick Start Guide

Adjustment

Observe lowest recorded output from connected sensors 

Set Channel Select DIP switches to first Channel ON and all other 
Channels OFF 

Apply known load to the first sensor 

Adjust SPAN potentiometer to adjust sensor mV output down, 
under load, while observing the sensor output till output is equal 
to lowest recorded output  Span potentiometer can only adjust 
outputs down 

Continue process of Channel Select DIP Switches, applied known 
load, and adjustment of mV output down on each sensor until 
each sensor has the same mV output under the sample applied 
load  To ensure best performance recheck the output of each 
sensor while under load and re-adjust if necessary 

A

B

C

D

E

C

E

OUT – (WHITE)
OUT + (GREEN)
IN – (BLACK)
IN + (RED)

4
3
2
1

OUTPUT

+E

–E

+S

–S

+E

–E

+S

–S

+E

–E

+S

–S

+E

–E

+S

–S

Load

E

D

Ch 1 Span

Ch 2 Span

Ch 3 Span

Ch 4 Span

C

A

CH SELECT

0

1

CH1

CH2

CH3

CH4

Sensor mV output

Sensor excitation

Serial No.

755523

895325

616263

953512

Output

20mV

18mV

19mV

21mV

C

E

OUT – (WHITE)
OUT + (GREEN)
IN – (BLACK)
IN + (RED)

4
3
2
1

OUTPUT

+E

–E

+S

–S

+E

–E

+S

–S

+E

–E

+S

–S

+E

–E

+S

–S

Load

E

D

Ch 1 Span

Ch 2 Span

Ch 3 Span

Ch 4 Span

C

A

CH SELECT

0

1

CH1

CH2

CH3

CH4

Sensor mV output

Sensor excitation

Serial No.

755523

895325

616263

953512

Output

20mV

18mV

19mV

21mV

Содержание IAC200

Страница 1: ...Sensor Solutions Source Load Torque Pressure Multi Axis Calibration Instruments Software www futek com IAC200 Four Channel Summing Junction Box Quick Start Guide...

Страница 2: ...EM1006 A FUTEK reserves the right to modify its design and specifications without notice Please visit https www futek com TermsAndConditions for complete terms and conditions 10 Thomas Irvine CA 9261...

Страница 3: ...e with certification Standard NIST and A2LA calibration requests are started online at https www futek com recalibrationterms Solutions are available at FUTEK to connect your IAC200 to the computer or...

Страница 4: ...nt for amplified IAC200 disconnect the ribbon cable from the center amplifier to the IAC200 board on connection J6 Applies to FSH04736 Voltage output and FSH04737 Current output versions Enable one se...

Страница 5: ...00 Refer to the sensor spec sheet online for the sensor s wiring Connect power supply to IAC200 for sensor excitation Refer to the sensor spec sheet online for max excitation Connect Voltage meter to...

Страница 6: ...apacity mV output level is a product of the sensor mV V rating and excitation Adjust Channel Select for next channel ON and switch OFF prior channel Continue to process until all sensor individual out...

Страница 7: ...inue process of Channel Select DIP Switches applied known load and adjustment of mV output down on each sensor until each sensor has the same mV output under the sample applied load To ensure best per...

Страница 8: ...l connected sensors into mV V by dividing the mV output by the supply voltage used during the setting of the one mV output This will be the one mV V combined signal going into the amplifier Refer to t...

Отзывы: