IAC200 Quick Start Guide
Adjustment
Observe lowest recorded output from connected sensors
Set Channel Select DIP switches to first Channel ON and all other
Channels OFF
Apply known load to the first sensor
Adjust SPAN potentiometer to adjust sensor mV output down,
under load, while observing the sensor output till output is equal
to lowest recorded output Span potentiometer can only adjust
outputs down
Continue process of Channel Select DIP Switches, applied known
load, and adjustment of mV output down on each sensor until
each sensor has the same mV output under the sample applied
load To ensure best performance recheck the output of each
sensor while under load and re-adjust if necessary
A
B
C
D
E
C
E
OUT – (WHITE)
OUT + (GREEN)
IN – (BLACK)
IN + (RED)
4
3
2
1
OUTPUT
+E
–E
+S
–S
+E
–E
+S
–S
+E
–E
+S
–S
+E
–E
+S
–S
Load
E
D
Ch 1 Span
Ch 2 Span
Ch 3 Span
Ch 4 Span
C
A
CH SELECT
0
1
CH1
CH2
CH3
CH4
Sensor mV output
Sensor excitation
Serial No.
755523
895325
616263
953512
Output
20mV
18mV
19mV
21mV
C
E
OUT – (WHITE)
OUT + (GREEN)
IN – (BLACK)
IN + (RED)
4
3
2
1
OUTPUT
+E
–E
+S
–S
+E
–E
+S
–S
+E
–E
+S
–S
+E
–E
+S
–S
Load
E
D
Ch 1 Span
Ch 2 Span
Ch 3 Span
Ch 4 Span
C
A
CH SELECT
0
1
CH1
CH2
CH3
CH4
Sensor mV output
Sensor excitation
Serial No.
755523
895325
616263
953512
Output
20mV
18mV
19mV
21mV