FCO432 Users Guide Issue 1 

Calibration 

Furness Controls recommend that all pressure transmitters and flow devices are recalibrated 
annually. For critical applications, it is advisable to check the calibration more frequently. 
 
All calibration adjustments of 400 series pressure transmitters are performed through a PC using 
the FCS541 Utility software/cable or using the optional keypad. 
 
Separate functions allow for independent adjustment of the minimum and maximum output, 
instrument gain and digital zero offset adjustment. 
The instruments can be recalibrated provided suitable equipment is available. The basic 
requirements are: 

 Pressure calibrator, such as the Furness Controls FCO560, FCO550 or FRS4 

 Accurate voltage or current meter 

 Suitable power supply 

 
Furness Controls can provide instruments to fulfil all these functions. 
Furness Controls offer a full recalibration and UKAS certification service. 
 

Recalibration using Keypad 

Check the minimum and maximum signal output levels by: 

 Using the 

 or 

 buttons to select 

TEST

, to enter menu press 

 

 Select 

MIN

 for the minimum output 

 Select 

MAX

 for the maximum output 

 

NOTE:  Pressing the 

 button between tests will return the display to 

TEST resulting in a false output reading. 

 

If the output levels are incorrect 

Enter Calibration Mode by holding down the 

 and 

 keys and reset the supply. 

 
Set the minimum output level: 

 Using the 

 or 

 buttons select 

DAC LO

, to enter menu press 

 

 Adjust the value using the 

 or 

 buttons 

 Press when 

 set 

 Press the 

 to exit the menu 

 

Set the maximum output level: 

 Using the 

 or 

 buttons select 

DAC HI

, to enter menu press 

 

 Adjust the value using the 

 or 

 buttons 

 Press 

 when set 

 Press the 

 to exit the menu 

 

Gain Adjustment 

If not already in Calibration Mode, hold down the 

 and 

 keys, and reset the supply. 

 

 Using the 

 or 

 buttons select 

GAIN

, to enter menu press 

 

 Adjust the value using the 

 or 

 buttons to adjust any error 

 Press 

 when set 

 Press the 

 to exit the menu 

 

Linearity and Temperature Compensation Adjustment 

Linearity and temperature compensation adjustments should only be carried out at Furness 
Controls, return to your supplier. 

 

Содержание FCO432

Страница 1: ...factured by Furness Controls Limited Beeching Road Bexhill East Sussex TN39 3LG England Telephone 44 0 1424 819980 Email Sales furness controls com Web Site http www furness controls com FCO432 Differ...

Страница 2: ...Page 2 FCO432 Users Guide Issue 1...

Страница 3: ...ture 14 Menu Navigation 15 Troubleshooting 22 Error Messages 23 Calibration 24 Recalibration using Keypad 24 Recalibration using the FCS541 Utility 25 Calculations 26 Pitot Tube Calculations 26 Lamina...

Страница 4: ...Page 4 FCO432 Users Guide Issue 1 Revision Record Issue Date Summary of changes 1 26 11 18 First release...

Страница 5: ...fuse for safety 2 Always unplug the equipment from the supply before carrying out any servicing 3 The air supply must be turned off before carrying out any work on the pneumatic system 4 The air suppl...

Страница 6: ...Optional Configurations 2 Wire 4 to 20mA Current output 4 Wire Isolated 4 to 20mA Current output 4 Wire Isolated Voltage output 0 to 1V 0 to 2V 0 to 5V 0 to 10V 1V 2V 5V or 10V Remote Zeroing Function...

Страница 7: ...should not be mounted in areas where they will be subjected to high vibration as this could cause spurious readings The enclosure has four mounting holes for M4 screws accessible with the cover remov...

Страница 8: ...e Code A shielded straight field installable connector with screw terminals and a PG9 cable gland suitable for cable 4 to 8mmOD The instrument can be supplied as one of three configurations See the ra...

Страница 9: ...FCO432 Users Guide Issue 1 Page 9 5 Connector wiring Assembly Note Housing alignment tabs must be fitted into the plastic locating slots and not the electrical terminals...

Страница 10: ...FCO432 can incorporate a pair of relays for triggering auxiliary equipment such as alarms Note the relays are supplied from the 24V on the power connector Pin functions Relay states are shown for the...

Страница 11: ...thin specified limits and inspect for poor connections After fixing the instrument in position allow at least one hour before commencing the final commissioning With no pressure applied check the outp...

Страница 12: ...s download site 2 Double click the Setup exe file 3 In the first instance the setup program will determine whether your system files need to be updated 4 When the first stage has finished Setup will n...

Страница 13: ...ure is applied this value should be capable of showing between 5500000 and 5500000 Please make sure the transmitter has stabilised at 20 when doing these tests The Internal Temperature should show a v...

Страница 14: ...reen Pin Set Pin Filter Display Units CST Unit DP Low DP High DEC PNT DSP Low DSP High Relays RL1 FUNC RL1 ON RL1 OFF RL1 DLY DISP RL1 RL2 FUNC RL2 ON RL2 OFF RL2 DLY DISP RL2 AZ Delay Z Lock Purge Te...

Страница 15: ...he data in memory and return to the main menu If no button is pressed for 20 seconds the display will revert back to the main menu and back to the normal display another 20 seconds after that To retur...

Страница 16: ...r 4 20mA The display limits should be set using the Custom Units DP Low DP High Decimal Point Display Low and Display High menus ROOT Activates the root function Both the display and output will be af...

Страница 17: ...next digit press the button to enter the number If a number is entered by mistake exit the menu and try again do not return to the display DP LOW Differential Pressure Low Output Purpose Alters the d...

Страница 18: ...age displayed for Non Failsafe Operation Note xx 1 for Relay 1 only 2 for Relay 2 only 12 for Relays 1 and 2 Relay Trigger Level Rules Relay state changes are suspended when in the menus or setting pa...

Страница 19: ...ss than the relay reset level the relay state will change to off for any signal between the reset and set level The relay will be on for signals above the reset level or below the set level This provi...

Страница 20: ...nnections to allow the master device some additional time to disable its transmitter and prepare for the reply Options 0 to 50 ms FLASH Flash Update Mode Purpose Puts the unit into its Flash Updating...

Страница 21: ...ory defined Unit ID number This is the number that identifies the internal transducer and main board circuitry as a matched unit Output Levels If a pressure is applied to the transmitter that is outsi...

Страница 22: ...lay reads This indicates that a differential pressure is being measured that is greater than the instruments current scaling Unable to change menu settings quickly Pressing either and on the keypad wi...

Страница 23: ...lure The auto zero valve sensor detected a fault Check the 24V supply is connected Error 5 EEPROM extended calibration data error Extended calibration data stored in the EEPROM may have become corrupt...

Страница 24: ...k the minimum and maximum signal output levels by Using the or buttons to select TEST to enter menu press Select MIN for the minimum output Select MAX for the maximum output NOTE Pressing the button b...

Страница 25: ...it count These settings can be adjusted against values measured using calibration equipment Amended settings can be downloaded to the connected transmitter by clicking on the Write to Tx button The Ou...

Страница 26: ...gas at std conditions air 1 Volumetric flow Once the velocity of a gas at a point in a pipe is known the volume flow rate can be calculated To calculate volumetric flow rate the pressure temperature...

Страница 27: ...he area and do not make the corrections for pressure and temperature Mass Flow To calculate the mass flow the volume flow at standard conditions must be known as well as the relative density of the ga...

Страница 28: ...ent in pascals std P Standard pressure in pascals normally 101325 pascals std T Standard temperature in kelvin 273 14 temperature in C lfe T Temperature of gas in laminar flow element in kelvin 273 14...

Страница 29: ...ed which has a maximum differential pressure range of 25mb The desired scaling is 1 to 16mb 4 to 20mA As the FCO432 has differential pressure ports the suction can be applied to the ve pressure port T...

Страница 30: ...e is 30 C The FCO432 is used with the FCO96 20L LFE Laminar Flow Element The LFE is calibrated to give a Differential Pressure of 104Pa for 20 l min of air at standard conditions Using the LFE calcula...

Страница 31: ...lculation the velocity for maximum flow is determined std std std T T P P A Q V 14 293 14 296 103500 101325 4 1 8 V V 5 651 m s Using the Pitot Tube velocity calculation the P for maximum velocity is...

Страница 32: ...tor as standard DIN Electrical Connection The diagrams below show typical wiring configurations Pin 4 to 20mA Output 1 ve 2 ve 3 nc 4 nc See rating label for power supply requirements and configuratio...

Страница 33: ...sure that the external zero signal is detected If the instrument is fitted with the automatic zero valve when the Zero is triggered the display and output are frozen for 1 5 seconds while the pneumati...

Страница 34: ...al Micro USB for instrument configuuration free utility software Optional external RS232 RS485 or USB port Performance Enhanced Accuracy 20 C 10 to 100 range 0 to 10 range 0 25 reading 1 digit 0 025 r...

Страница 35: ...FCO432 Users Guide Issue 1 Page 35 Dimension Drawings Basic Instrument Pipe Mounting Kit...

Страница 36: ...Page 36 FCO432 Users Guide Issue 1 Wall Mounting Kit...

Отзывы: