CST UNIT  (Custom Units) 

Purpose: 

Sets up a user defined unit display. Can only be used when the Display Menu is set 
for CUST or ROOT. A variety of engineering units can be set.    

Note: These are just display labels and use the menu functions DSP LO, DSP 

HIGH and DEC PNT to control the displayed value. 

Options: 

PA, KPA, MB … CCM, CCH (33 options as listed below) 

 

 

Unit  Description 

 

 

Unit 

Description 

PA 

Pascal 

 

18 

CFH 

Cubic Feet per hour 

KPA  Kilo Pascal 

 

19 

CMS  Cubic Metres per second 

MB 

Millibar 

 

20 

CMM  Cubic Metres per minute 

MM 

Millimetre H2O gauge 

 

21 

CMH  Cubic Metres per hour 

IN 

Inches H2O gauge 

 

22 

LBS 

Pounds per second 

HG 

Millimetre Hg 

 

23 

LBM 

Pounds per minute 

IHG 

Inches Hg 

 

24 

LBH 

Pounds per hour 

PSI 

Pounds per square inch 

 

25 

KGS  Kilograms per second 

M/S 

Meters per second 

 

26 

KGM  Kilograms per minute 

MPH  Miles per hour 

 

27 

KGH  Kilograms per hour 

10 

F/S 

Feet per second 

 

28 

MA 

Milliamps 

11 

F/M 

Feet per minute 

 

29 

PC 

Percent 

12 

KMH  Kilometres per hour 

 

30 

SPACE 

Units not displayed 

13 

L/S 

Litres per second 

 

31 

CCS  Cubic Centimetres per second* 

14 

L/M 

Litres per minute 

 

32 

CCM  Cubic Centimetres per minute* 

15 

L/H 

Litres per hour 

 

33 

CCH  Cubic Centimetres per hour* 

16 

CFS  Cubic Feet per second 

 

 

 

 

17 

CFM  Cubic Feet per minute 

 

 

 

 

 
Press the 

 and 

 buttons to select a number, if no more buttons are pressed the cursor will 

move to the next digit, press the 

 button to enter the number. If a number is entered by mistake 

exit the menu and try again (do not return to the display). 

DP LOW (Differential Pressure Low Output) 

Purpose: 

Alters the differential pressure corresponding to the minimum output level.  

 

The value can be set anywhere within the range of the instrument. The display will 
be in the current engineering units selected. 

Options: 

Any differential pressure value within the display or instrument range.  

DP HIGH (Differential Pressure High Output) 

Purpose: 

Alters the differential pressure corresponding to the maximum output level.  

 

The value can be set anywhere within the range of the instrument. The display will 
be in the current engineering units selected. 

Options: 

Any differential pressure value within the display or instrument range.  

DEC PNT (Decimal Point Position) 

Purpose: 

Alters the position of the display decimal point if the display option is set to CUST or 
ROOT. 

Options: 

NNNN, NNN.N, NN.NN, N.NNN or .NNNN

 

DSP LOW (Display Low Reading) 

Purpose: 

Alters the display value corresponding to the minimum output level. Can be set to 
any numeric value within the range of the display. For instance, to display mA set the 
value to 4.00. For %, set this value to 0. 

Options: 

Any numeric value shown on the display 

DSP HIGH (Display High Reading) 

Purpose: 

Alters the display value that corresponds to the maximum output level. Can be set to 
any numeric value within the range of the display. For instance, to display in mA set 
the value to 20.00. For %, set this value to 100. 

Options: 

Any numeric value shown on the display 

Содержание FCO432

Страница 1: ...factured by Furness Controls Limited Beeching Road Bexhill East Sussex TN39 3LG England Telephone 44 0 1424 819980 Email Sales furness controls com Web Site http www furness controls com FCO432 Differ...

Страница 2: ...Page 2 FCO432 Users Guide Issue 1...

Страница 3: ...ture 14 Menu Navigation 15 Troubleshooting 22 Error Messages 23 Calibration 24 Recalibration using Keypad 24 Recalibration using the FCS541 Utility 25 Calculations 26 Pitot Tube Calculations 26 Lamina...

Страница 4: ...Page 4 FCO432 Users Guide Issue 1 Revision Record Issue Date Summary of changes 1 26 11 18 First release...

Страница 5: ...fuse for safety 2 Always unplug the equipment from the supply before carrying out any servicing 3 The air supply must be turned off before carrying out any work on the pneumatic system 4 The air suppl...

Страница 6: ...Optional Configurations 2 Wire 4 to 20mA Current output 4 Wire Isolated 4 to 20mA Current output 4 Wire Isolated Voltage output 0 to 1V 0 to 2V 0 to 5V 0 to 10V 1V 2V 5V or 10V Remote Zeroing Function...

Страница 7: ...should not be mounted in areas where they will be subjected to high vibration as this could cause spurious readings The enclosure has four mounting holes for M4 screws accessible with the cover remov...

Страница 8: ...e Code A shielded straight field installable connector with screw terminals and a PG9 cable gland suitable for cable 4 to 8mmOD The instrument can be supplied as one of three configurations See the ra...

Страница 9: ...FCO432 Users Guide Issue 1 Page 9 5 Connector wiring Assembly Note Housing alignment tabs must be fitted into the plastic locating slots and not the electrical terminals...

Страница 10: ...FCO432 can incorporate a pair of relays for triggering auxiliary equipment such as alarms Note the relays are supplied from the 24V on the power connector Pin functions Relay states are shown for the...

Страница 11: ...thin specified limits and inspect for poor connections After fixing the instrument in position allow at least one hour before commencing the final commissioning With no pressure applied check the outp...

Страница 12: ...s download site 2 Double click the Setup exe file 3 In the first instance the setup program will determine whether your system files need to be updated 4 When the first stage has finished Setup will n...

Страница 13: ...ure is applied this value should be capable of showing between 5500000 and 5500000 Please make sure the transmitter has stabilised at 20 when doing these tests The Internal Temperature should show a v...

Страница 14: ...reen Pin Set Pin Filter Display Units CST Unit DP Low DP High DEC PNT DSP Low DSP High Relays RL1 FUNC RL1 ON RL1 OFF RL1 DLY DISP RL1 RL2 FUNC RL2 ON RL2 OFF RL2 DLY DISP RL2 AZ Delay Z Lock Purge Te...

Страница 15: ...he data in memory and return to the main menu If no button is pressed for 20 seconds the display will revert back to the main menu and back to the normal display another 20 seconds after that To retur...

Страница 16: ...r 4 20mA The display limits should be set using the Custom Units DP Low DP High Decimal Point Display Low and Display High menus ROOT Activates the root function Both the display and output will be af...

Страница 17: ...next digit press the button to enter the number If a number is entered by mistake exit the menu and try again do not return to the display DP LOW Differential Pressure Low Output Purpose Alters the d...

Страница 18: ...age displayed for Non Failsafe Operation Note xx 1 for Relay 1 only 2 for Relay 2 only 12 for Relays 1 and 2 Relay Trigger Level Rules Relay state changes are suspended when in the menus or setting pa...

Страница 19: ...ss than the relay reset level the relay state will change to off for any signal between the reset and set level The relay will be on for signals above the reset level or below the set level This provi...

Страница 20: ...nnections to allow the master device some additional time to disable its transmitter and prepare for the reply Options 0 to 50 ms FLASH Flash Update Mode Purpose Puts the unit into its Flash Updating...

Страница 21: ...ory defined Unit ID number This is the number that identifies the internal transducer and main board circuitry as a matched unit Output Levels If a pressure is applied to the transmitter that is outsi...

Страница 22: ...lay reads This indicates that a differential pressure is being measured that is greater than the instruments current scaling Unable to change menu settings quickly Pressing either and on the keypad wi...

Страница 23: ...lure The auto zero valve sensor detected a fault Check the 24V supply is connected Error 5 EEPROM extended calibration data error Extended calibration data stored in the EEPROM may have become corrupt...

Страница 24: ...k the minimum and maximum signal output levels by Using the or buttons to select TEST to enter menu press Select MIN for the minimum output Select MAX for the maximum output NOTE Pressing the button b...

Страница 25: ...it count These settings can be adjusted against values measured using calibration equipment Amended settings can be downloaded to the connected transmitter by clicking on the Write to Tx button The Ou...

Страница 26: ...gas at std conditions air 1 Volumetric flow Once the velocity of a gas at a point in a pipe is known the volume flow rate can be calculated To calculate volumetric flow rate the pressure temperature...

Страница 27: ...he area and do not make the corrections for pressure and temperature Mass Flow To calculate the mass flow the volume flow at standard conditions must be known as well as the relative density of the ga...

Страница 28: ...ent in pascals std P Standard pressure in pascals normally 101325 pascals std T Standard temperature in kelvin 273 14 temperature in C lfe T Temperature of gas in laminar flow element in kelvin 273 14...

Страница 29: ...ed which has a maximum differential pressure range of 25mb The desired scaling is 1 to 16mb 4 to 20mA As the FCO432 has differential pressure ports the suction can be applied to the ve pressure port T...

Страница 30: ...e is 30 C The FCO432 is used with the FCO96 20L LFE Laminar Flow Element The LFE is calibrated to give a Differential Pressure of 104Pa for 20 l min of air at standard conditions Using the LFE calcula...

Страница 31: ...lculation the velocity for maximum flow is determined std std std T T P P A Q V 14 293 14 296 103500 101325 4 1 8 V V 5 651 m s Using the Pitot Tube velocity calculation the P for maximum velocity is...

Страница 32: ...tor as standard DIN Electrical Connection The diagrams below show typical wiring configurations Pin 4 to 20mA Output 1 ve 2 ve 3 nc 4 nc See rating label for power supply requirements and configuratio...

Страница 33: ...sure that the external zero signal is detected If the instrument is fitted with the automatic zero valve when the Zero is triggered the display and output are frozen for 1 5 seconds while the pneumati...

Страница 34: ...al Micro USB for instrument configuuration free utility software Optional external RS232 RS485 or USB port Performance Enhanced Accuracy 20 C 10 to 100 range 0 to 10 range 0 25 reading 1 digit 0 025 r...

Страница 35: ...FCO432 Users Guide Issue 1 Page 35 Dimension Drawings Basic Instrument Pipe Mounting Kit...

Страница 36: ...Page 36 FCO432 Users Guide Issue 1 Wall Mounting Kit...

Отзывы: